510(k) Data Aggregation

    K Number
    K233816

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-28

    (272 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The REDEMPTION™ Charcot Plating System bone plates and screws are indicated for use in bone fractures, osteotomies, and fixation of bones and bone fragments in the upper and lower extremities, primarily of the hand, wrist, foot, and digits. Specific examples include: medial column fusion (talus, navicular, cuboid, first metatarsal) for patients with osteopenic bone in neuropathic osteoarthropathy (Charcot).

    Device Description

    The REDEMPTION™ Charcot Plating System contains bone plates, screws, and accompanying instruments to aid in the surgical procedures. The system is a multi-indication reconstruction solution providing polyaxial locking technology and low-profile design. The system is optimized for use in osteopenic bone, such as those of neuropathic osteoarthropathy (Charcot) patients.

    The system contains poly-axial locking and non-locking 4.0mm and 5.5mm diameter screws with threads optimized for osteopenic bone purchase.

    All implant components are manufactured from titanium (Ti-6AI-4V, ASTM F136).

    Specific instrumentation including wires, drills, torx drivers, and drill guides are required for use with the system. The REDEMPTION™ instruments are manufactured from stainless steel, with a few Aluminum and Silicone components.

    AI/ML Overview

    I am sorry, but the provided text from the FDA 510(k) K233816 document for the "REDEMPTION™ Charcot Plating System" does not contain any information about acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of an AI/ML medical device.

    The document describes a traditional 510(k) submission for a metallic bone fixation system (plates and screws), not an AI/ML-driven diagnostic or therapeutic device. The "Performance Data" section details mechanical and material testing for the bone fixation system, such as:

    • Static four-point bend testing per ASTM F382-17
    • Screw Torsional Properties per ASTM F543-17 A1
    • Screw Insertion and Removal Torque per ASTM F543-17 A2
    • Screw Pullout Testing per ASTM F543-17 A3
    • Cleaning and sterilization validations

    These tests are standard for orthopedic implants to demonstrate structural integrity, and biocompatibility, and ensure the device can be properly cleaned and sterilized.

    Therefore, I cannot provide the requested information regarding:

    1. A table of acceptance criteria and reported device performance for an AI/ML device.
    2. Sample size, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, or ground truth types, as these are concepts applicable to AI/ML software validation, not the mechanical testing of orthopedic hardware described in this document.