Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K233609

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-03-28

    (136 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CORE 500 Digital Stethoscope is intended to be used by clinicians or lay users to electronically amplify, filter, and transfer body sounds and three lead electrocardiogram (ECG) waveforms. The CORE 500 Digital Stethoscope also displays ECG waveforms and heart rate on the display and accompanying mobile application (when prescribed or used under the care of a clinician or by lay users).

    A lay user is not intended to interpret or take clinical action based on the device output without consulting with a qualified healthcare professional.

    Device Description

    CORE 500 Digital Stethoscope (CORE 500) is an electronic stethoscope with integrated electrodes for electrocardiogram (ECG). The device consists of a chestpiece, detachable earpiece (Eko Earpiece) and a mobile application (Eko App) and is intended as a digital auscultation tool on patients requiring physical assessment by the clinicians or lay users. CORE 500 provides the ability to amplify, filter, and transfer body sounds with the chestpiece diaphragm, and three lead ECG through electrodes integrated around the chestpiece. The device can be used in a professional healthcare facility and for home use.

    CORE 500 features three auscultation modes for a better auscultation experience by filtering acoustic data and enhancing the primary frequency range of particular body sounds: Cardiac Mode for heart sounds, Pulmonary Mode for lung sounds, and Wide Band Mode for general auscultation. CORE 500 also detects and computes the heart rate in real time based on the phonocardiogram (PCG) data.

    AI/ML Overview

    This FDA 510(k) summary for the Eko Health, Inc. CORE 500 Digital Stethoscope (K233609) describes the device's technical specifications and how it compares to a predicate device. Regarding acceptance criteria and detailed study results, the document provides a general overview rather than specific performance metrics.

    Here's an analysis of the provided information concerning acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria with corresponding reported device performance values for the CORE 500 Digital Stethoscope in the way one might expect for a clinical performance study. Instead, it lists the types of nonclinical testing performed and asserts that the device complies with standards or demonstrates performance.

    Here's a summary of the reported performance without specific numerical acceptance criteria from the document:

    Acceptance Criteria (Inferred from testing type)Reported Device Performance
    Biocompatibility (ISO 10993-1:2018)Concluded that the CORE 500 Digital Stethoscope is biocompatible.
    Electrical safety (IEC 60601-1-11, IEC 60601-2-47)Demonstrated compliance with standards for safety.
    Electromagnetic Compatibility (EMC) (IEC 60601-1-2)Demonstrated compliance with standards for EMC.
    Software Verification and Validation (FDA guidance for Content of Premarket Submissions for Device Software Functions)Software is verified and validated.
    Usability Testing (IEC 62366-1)Intended users are able to achieve intended use with Instructions for Use.
    Audio performanceRigorous bench testing demonstrated product performance.
    Electrical and mechanical function verificationRigorous bench testing demonstrated product performance.
    Heart rate measurementRigorous bench testing demonstrated product performance.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not provide specific sample sizes for test sets, data provenance, or whether studies were retrospective or prospective. The performance data section focuses on nonclinical testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not provided in the document. The performance data is described as "nonclinical testing" and does not appear to involve expert-adjudicated ground truth as typically found in clinical studies assessing diagnostic accuracy.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not provided. As the document focuses on nonclinical performance, an adjudication method on a clinical test set is not described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. The device, the CORE 500 Digital Stethoscope, is primarily an electronic stethoscope for amplifying, filtering, and transferring body sounds and ECG waveforms, and displaying ECG and heart rate. It is not described as having an AI diagnostic interpretation component that would typically be evaluated in an MRMC study with human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not explicitly state that a standalone (algorithm only) performance study was done for any specific AI functionality. The device displays ECG waveforms and heart rate, but the document does not describe it as having an autonomous diagnostic algorithm for complex conditions.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Given that the performance data described is "nonclinical testing" (bench testing, biocompatibility, electrical safety, software V&V, usability), the concept of "ground truth" as it applies to clinical diagnostic accuracy (e.g., expert consensus, pathology) is not applicable or described in this section. The testing would have focused on meeting technical specifications and regulatory standards.

    8. The sample size for the training set

    The document does not mention a training set or its sample size. This type of information would typically be provided for devices involving machine learning or AI algorithms with extensive training phases, which is not the primary focus of the performance data in this submission.

    9. How the ground truth for the training set was established

    Since no training set is mentioned (see point 8), there is no information on how ground truth for a training set was established.


    Summary of Device and Performance Context:

    The K233609 submission for the CORE 500 Digital Stethoscope primarily focuses on demonstrating substantial equivalence to its predicate device (K230111) and a reference device (K200776), particularly for its expanded "Over-The-Counter Use" and inclusion of "lay users." The performance data provided are centered on foundational nonclinical tests to ensure safety, efficacy, and compliance with general device regulations and standards. It's not a submission for a novel diagnostic AI algorithm requiring extensive clinical performance studies with ground truth establishment by experts. The "nonclinical testing" confirms the device's technical functionality, biocompatibility, electrical safety, software validation, and usability for its intended purpose of amplifying, filtering, and transferring body sounds and ECG waveforms, and displaying basic heart rate and ECG.