510(k) Data Aggregation

    K Number
    K233590

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-02-23

    (107 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Overiet Charting Assist is a Medical Image Management and Processing System (MIMPS) intended to detect natural dental structures including detection of tooth anatomy (enamel, pulp), and tooth numbering, as well as dental structures added through past restorative treatments: implants, crowns, endodontic treatment (previous root canal treatment), fillings. The device is intended to assist dental professionals in producing dental charts based on image analysis.

    The Overiet Charting Assist detects these finding (BW) and periapical (PA) images for patients with primary and/or permanent teeth (Ages 5 and above), and panoramic (Pano) radiographs for patients with only permanent teeth. The device is not intended as a replacement for a complete clinical judgment that considers other relevant information from the image or patient history.

    Device Description

    Overjet Charting Assist is a Medical Image Management and Processing System (MIMPS) device aimed to assist dental professionals in detecting dental structures and producing dental charting data based on interpretation of 2D dental radiographs. The device is intended to detect natural dental structures including detection of tooth anatomy (enamel, pulp), and tooth numbering, as well as dental structures added through past restorative treatments: implants, crowns, endodontic treatment (previous root canal treatment), fillings. The device is intended to assist dental professionals in producing dental charts based on image analysis.

    The Overjet Charting Assist detects these findings on bitewing (BW) and periapical (PA) images for patients with primary and/or permanent teeth (Ages 5 and above), and panoramic (Pano) radiographs for patients with only permanent teeth. The device is not intended as a replacement for a complete clinician's review or clinical judgment that considers other relevant information from the image or patient history.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    ParameterAcceptance Criteria (Explicit or Implied)Reported Device Performance
    Past Restorative Structures (Overall)"All conducted tests produced results that exceeded predefined acceptance criteria." (Implied: High sensitivity and specificity)Tooth Level Sensitivity: 88.3% (95% CI: 86.6%, 90.1%) Specificity (overall across specific structures): Ranged from 91.5%-100% across clinical sites; Specificity for individual structures are high (e.g., Fillings: 0.986, RCT: 0.999, Crown: 0.994, Implant: 0.998) Dice: 0.918 (0.078) (95% CI: 0.914, 0.923) PPV: 0.958 (95% CI: 0.948, 0.967) NPV: 0.945 (95% CI: 0.935, 0.954)
    Dental Tooth Anatomy (Overall)"All conducted tests produced results that exceeded predefined acceptance criteria." (Implied: High sensitivity and specificity)Tooth Level Sensitivity: 95.9% (95% CI: 95.1%, 96.5%) Specificity (overall across specific anatomy): Ranged from 86.3% (Enamel) to 95.1% (Pulp). Dice: 0.836 (0.098) (95% CI: 0.832, 0.842) PPV: 0.967 (95% CI: 0.960, 0.973) NPV: 0.710 (95% CI: 0.672, 0.745)
    Tooth Numbering (Overall Classification Accuracy)"All conducted tests produced results that exceeded predefined acceptance criteria." (Implied: High accuracy)Overall Classification Accuracy: 98.8% (95% CI: 98.4%, 99.1%)
    Manual Charting Reduction Rate"The predefined Acceptance Criteria established for the standalone study are based on the current state of dental practice and are appropriate to demonstrate that the Overjet Charting Assist device performs in accordance with specifications and will meet user needs and intended uses." (Implied: Significant reduction in manual operations)Mean % Reduction per image in Manual Charting Operations: 80.5% (95% CI: 79.0%, 81.9%)

    Note: The document explicitly states, "All conducted tests produced results that exceeded predefined acceptance criteria," but does not explicitly list numerical thresholds for each criterion. The reported performance metrics are presented as the results that met or exceeded these unstated criteria.

    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: 634 images (259 Bitewing, 239 Periapical, 136 Panoramic).
      • Data Provenance: From various clinics across the U.S. (44 distinct clinics). The data includes male and female patients of primary and permanent dentition. The study explicitly states it uses data from all U.S. regions (Southeast, Southwest, Northeast, West). The data collection method (retrospective or prospective) is not explicitly stated, but the mention of "obtained from male and female patients" and "collected from 44 distinct clinics" suggests retrospective collection of existing images.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Not explicitly stated as a numerical count for each image, but the ground truth was established by "trained dentists."
      • Qualifications of Experts: Described as "trained dentists." No specific years of experience or board certifications are provided.
    3. Adjudication method for the test set:

      • The ground truth was established by "majority pixel voting" among trained dentists. This implies a consensus-based adjudication method where multiple experts reviewed the images, and their agreement (or majority vote) determined the ground truth. The exact number of experts involved in each "vote" is not specified (e.g., 2+1, 3+1).
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • A formal MRMC comparative effectiveness study demonstrating human reader improvement with AI assistance (vs. without AI) was not explicitly detailed.
      • The study primarily focused on the standalone performance of the device and its ability to reduce manual charting operations. The metric "Mean % Reduction per image in Manual Charting Operations with (95% Cl) when using the OChA device is 80.5% (79.0%, 81.9%)" quantifies the efficiency gain, which is an indirect measure of how much human effort is reduced, rather than a direct measure of improved diagnostic accuracy by human readers with vs without AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance evaluation was conducted. The document explicitly states: "Standalone performance of the Overjet Charting Assist device was evaluated..." and "Tooth Level Standalone Analysis for subgroups of past restorative structures...", "Tooth Level Standalone Analysis for subgroups of dental tooth anatomy...", etc.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The ground truth used was expert consensus based on "robust consensus reference standard established by trained dentists via majority pixel voting."
    7. The sample size for the training set:

      • The training set sample size is not explicitly mentioned in the provided text. The document describes the test set and its performance evaluation but does not detail the training data or its size.
    8. How the ground truth for the training set was established:

      • Since the training set details are not provided, the method for establishing its ground truth is also not stated in this document.