Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K233509

    Validate with FDA (Live)

    Date Cleared
    2024-04-10

    (162 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as a cervical intervertebral body fusion device, the Zavation IBF implants are intended to be used with autograft or allogenic bone graft comprised of cancellous and/or corticancellous bone graff in skeletally mature patients. Cervical IBF implants are intended for use at one level in the cervical spine, from C2 to T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is intended to be used in patients who have had six weeks of non-operative treatment.

    When used as a lumbar intervertebral body fusion device, the Zavation IBF implants are intended to be used with autograft or allogenic bone graft comprised of cancellous bone graft in skeletally mature patients. The lumbar IBF implants are intended for use at either one levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non -operative treatment. For all the above indications the Zavation IBF implants are intended to be used with supplemental internal fixation appropriate for the implanted level, including Zavation Pedicle Screw System and Zavation Cervical Plate System.

    The VariSync Plate is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.

    The VariSync Spacer is an interbody fusion device in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (0) weeks of non-operative treatment. The VariSync Spacer is to be filled with autograft or allogenic bone graft comprised of cancellous and/or corticancellous bone graft in skeletally mature patients. These devices are intended to be used with supplemental fixation such as the Zavation VariSync Plate, Zavation Midline Plate, Zavation EZ Plate, or Zavation Cervical Plate Systems. When used with the VariSync Plate, the assembly takes on the indications of the VariSync Spacer, with the VariSync Plate acting as the supplemental fixation.

    When used as a cervical intervertebral body fusion device, the LABYRINTH implants are intended to be used with autograft or allogenic bone graft comprised of cancellous and/or corticancellous bone graff in skeletally mature patients. Cervical IBF implants are intended for use at one level in the cervical spine, from C2 to T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is intended to be used in patients who have had six weeks of non-operative treatment.

    When used as a lumbar intervertebral body fusion device, the LABYRINTH implants are intended to be used with autograft or allogenic bone graft comprised of cancellous bone graft in skeletally mature patients. The lumbar IBF implants are intended for use at either one levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non -operative treatment. For all the above indications the LABYRINTH implants are intended to be used with supplemental internal fixation appropriate for the implanted level, including Zavation Pedicle Screw System and Zavation Cervical Plate System.

    The Zavation ALIF System is a stand-alone anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). The interior of the spacer component of the Zavation ALIF System is to be filled with autogenous bone graft material.

    The Zavation ALIF System spacer and plate assembly are an integrated for stand-alone use when used with screws. When used with anchors only the recessed plate may be used, and the assembly is intended for use with additional supplemental fixation that has been cleared by the FDA for use in the lumbar spine.

    Hyperlordotic interbody devices (>20 degrees) must be used with supplemental fixation (e.g. posterior fixation) that has been cleared by the FDA for the use in the lumbar spine.

    Device Description

    The Zavation IBF implants offers a variety of heights, widths, and lengths. There are six main configurations: ALIF, LLIF, TLIF, T-PLIF, PLIF and CIF. The different configurations allow for multiple surgical technique options. The implants are manufactured from medical grade PEEK (Polyetheretherketone) Superior Polymers Magnolia PEEK (ASTM F2026).

    The Zavation IBF implants are available in a range of sizes, as well as parallel and lordotic angled implants, to accommodate variations in patients' anatomy. In addition, tantalum beads or pins are embedded in the implants as an option to help allow for radiographic visualization. The ends of the implants have machined teeth which are designed to engage with the vertebral body end plates. This modification seeks clearance for the addition of an ion beam assisted deposition (IBAD) coating made from Grade 2 Commercially Pure Titanium that meets chemical composition requirements of ASTM F67 on previously cleared PEEK devices. The (IBAD) coated PEEK implants will be provided non-sterile.

    The VariSync Plate is an anterior, cervical fixation device available in various heights and widths to fit the anatomical needs of a wide variety of patients. The plates are made from titanium alloy, as specified in ASTM F136. The Screws for use with the VariSync Plates are manufactured from titanium alloy, as specified in ASTM F136.

    VariSync Spacers are anterior cervical interbody fusion devices used to provide structural stability in skeletally mature individuals following discectomy. The spacers are available in various heights and footprints to fit the anatomical needs of a wide variety of patients. These devices are to be filled with autogenous bone graft material. Protrusions on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to resist expulsion. The VariSync Spacers are manufactured from Superior Polymers Magnolia PEEK, with tantalum markers, as specified in ASTM F2026 and F560. This modification seeks clearance for the addition of an ion beam assisted deposition (IBAD) coating made from Grade 2 Commercially Pure Titanium that meets chemical composition requirements of ASTM F67 on previously cleared PEEK devices. The (IBAD) coated PEEK implants will be provided non-sterile.

    The LABYRINTH implants offer a variety of heights, widths and lengths. There are six main configurations: ALIF, LLIF, TLIF, T-PLIF, PLIF and CIF. The different configurations allow for multiple surgical technique options. The implants are manufactured from medical grade PEEK (Polyetheretherketone) Superior Polymers Magnolia PEEK (ASTM F2026). The LABYRINTH implants are available in a range of sizes, as well as parallel and lordotic angled implants, to accommodate variations in patients' anatomy. In addition, tantalum beads or pins are embedded in the implants as an option to help allow for radiographic visualization. The ends of the implants have machined teeth which are designed to engage with the vertebral body end plates. The LABYRINTH device has a porous structure on the end plate and through the device. This modification seeks clearance for the addition of an ion beam assisted deposition (IBAD) coating made from Grade 2 Commercially Pure Titanium that meets chemical composition requirements of ASTM F67 on previously cleared PEEK devices. The (IBAD) coated PEEK implants will be provided sterile.

    The Zavation ALIF System includes a spacer, plate, screws, and anchors. The spacer component is assembled to an interbody plate and implanted anteriorly. The spacer components are available in a variety of materials, depths, widths, and heights. The plate component includes three or four holes for inserting bone screws or anchors. The plate component also includes a lock at each hole. The bone screws are available in a variety of diameters and lengths. The anchors are available in a variety of lengths. The interbody plate components are available in a variety of heights. This modification seeks clearance for the addition of an ion beam assisted deposition (IBAD) coating made from Grade 2 Commercially Pure Titanium that meets chemical composition requirements of ASTM F67 on previously cleared PEEK devices. The (IBAD) coated PEEK implants will be provided non-sterile.

    AI/ML Overview

    The provided text describes several medical devices: Zavation IBF System, Zavation Varisync Plate System and Varisync Spacer System, LABYRINTH System, and Zavation ALIF System. The 510(k) summary indicates that this submission is for the addition of an ion beam assisted deposition (IBAD) Titanium coating to previously cleared PEEK devices. The acceptance criteria and the study to prove the device meets these criteria are focused on the coating properties rather than the overall device performance for fusion, as the underlying PEEK devices were previously cleared.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Standard)Reported Device Performance (Property Tested)
    Not explicitly stated as acceptance criteria, but tested against ASTM standardsTensile Bond Strength
    Not explicitly stated as acceptance criteria, but tested against ASTM standardsStatic Shear Strength
    Not explicitly stated as acceptance criteria, but tested against ASTM standardsShear Fatigue Strength
    Not explicitly stated as acceptance criteria, but tested against ASTM standardsTabor Abrasion Test

    Note: The document states "The specifications of the titanium coating are identical to that of the primary predicate device cleared in (K201367)." This implies that the acceptance criteria for these tests would be met if the results are comparable to the predicate device's coating performance. However, specific numerical thresholds for acceptance are not provided in this document. The conclusion states that "substantial equivalence to the cited legally marketed predicate devices" was demonstrated and "any differences between the subject devices and predicate devices do not impact the safety and effectiveness."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size used for the coating property tests (Tensile Bond Strength, Static Shear Strength, Shear Fatigue Strength, Tabor Abrasion Test). It also does not provide information on the data provenance (e.g., country of origin, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable. The study described focuses on bench performance testing of coating properties, not clinical or image-based evaluations requiring expert interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable. The study described focuses on bench performance testing of coating properties, not clinical or image-based evaluations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The document describes a 510(k) submission for an interbody fusion device with a new coating, not an AI-powered diagnostic or assistive technology.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable. The document describes a medical device (interbody fusion system), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the coating properties tests, the "ground truth" would be the established performance standards or specifications defined by the ASTM standards (ASTM F1147, ASTM F1044, ASTM F1160, and ASTM F1978) and the performance of the predicate device's coating (from K201367).

    8. The sample size for the training set

    This section is not applicable. The document describes bench testing of a medical device coating, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    This section is not applicable, as there is no training set for the described bench tests.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1