    K Number
    K233064

    Validate with FDA (Live)

    Device Name
    Copra Supreme
    Date Cleared
    2024-09-18

    (358 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Copra Supreme zirconia blanks are indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.

    Device Description

    Pre-sintered zirconia blanks for the fabrication of individual dental restorations. Copra Supreme disks come in both white and pre-shaded 0M1-D4 in various sizes.

    AI/ML Overview

    The provided document is a 510(k) summary for a dental device (zirconia blanks) and does not describe an AI/ML-enabled device or a study involving human readers or ground truth established by experts. Therefore, I am unable to provide the detailed information requested regarding acceptance criteria and performance evaluation of an AI device.

    The document primarily focuses on establishing substantial equivalence for a physical dental material (zirconia blanks) to legally marketed predicate devices. The "Testing Summary" section on page 5 states: "The physical properties of Copra Supreme were tested according to ISO 6872:2015 and all parameters meet the standard. Flexural strength, coefficient of thermal expansion, solubility and radioactivity were tested." This refers to material property testing, not performance criteria for a diagnostic AI algorithm.

    The table presented on pages 6-7 compares the characteristics of the subject device (Copra Supreme zirconia blanks) to predicate and reference devices, including:

    • Acceptance Criteria/Performance Metrics (as reported for a physical material):
      • Flexural Strength (MPa)
      • Fracture Toughness (MPa·m1/2)
      • Solubility (µg/cm²)
      • Radioactivity (Bq·g-1 238U)

    Since the request is about an AI/ML device and the provided document is for a physical dental material, most of the requested fields are not applicable.