510(k) Data Aggregation

    K Number
    K232674

    Validate with FDA (Live)

    Device Name
    PainFreeze II
    Date Cleared
    2024-02-12

    (164 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PainFreeze II Medium Stream and Mist Spray are vapocoolants (skin refrigerants) intended for topical op skin. intact mucous membranes (oral cavity, nasal passageways and the lips) and minor open wounds. PainFreeze II's skin refrigerant controls pain associated with injections (venipuncture, IV starts, cosmetic procedures), minor surgical (lancing boils, incisions, drainage of small abscesses, and sutures), and the temporary relief of minor sports injuries (sprains, bruising, cuts and abrasions).

    Device Description

    The Subject Device consists of a chemical blend of 60% HFO-1233zd (Trans-1-chloro-3.3.3- trifluoropropene) and 40% HFO-1234ze (trans- 1.3.3.3-tetrafluoroprop-1- ene) that produces a cooling effect upon contact with the skin, intact mucous membranes, and minor open wounds. The product is delivered in the form of an aerosol in either a mist or stream spray. The device's delivery system controls the amount of chemical blend that is dispensed, the mist spray configuration produces very fine droplets that cools at the points of contact. The stream spray configuration produces a pinpoint stream that contacts the skin surface at a specific single location. Upon contact with the skin, the product begins to evaporate immediately as it contacts the skin. The low evaporation rate of the product's chemical blend is what creates the coldness. The Subject Device is non- flammable and has a lower global warming potential than the predicate and provides equivalent performance.

    The Subject Device is topically sprayed onto the skin. It creates a cooling effect on the surface of the application site by the immediate evaporation of the product from the skin surface.

    With both product configurations, the evaporation creates the cold sensation. As the distance from the target surface is increased, the dispersion of the droplets in both the mist and stream is increased. Increasing the surface area of contact and decreasing the size of the droplets increases the evaporation rate. The increase in evaporation rate correlates to an increase in the cooling effect.

    The chemical blend is incorporated in pressurized canisters, which will be offered in two (2) volumes 35 and 115 ml. The canister consists of a high-pressure canister, valve, and actuator.

    AI/ML Overview

    The provided document is an FDA 510(k) Premarket Notification clearance letter for a medical device called "PainFreeze II," which is a vapocoolant (skin refrigerant). This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving enhanced clinical effectiveness or accuracy via a traditional clinical trial or performance study typical for diagnostic/AI devices.

    Therefore, the document does not contain the kind of detailed information requested about acceptance criteria, study design for proving device performance, sample sizes for test/training sets, expert qualifications, or multi-reader multi-case studies that would be present for a diagnostic device or an AI/ML algorithm. The "performance" being assessed here is primarily related to safety, physical characteristics, and equivalence to a predicate, not diagnostic accuracy or clinical outcome improvement.

    Based on the provided text, here's what can be extracted and what is not available:

    1. A table of acceptance criteria and the reported device performance:

    The document describes non-clinical testing performed to demonstrate that the device is similar to its predicate and meets safety/performance requirements. The "criteria" are implied by the successful meeting of test requirements for each evaluation.

    Acceptance Criteria (Implied)Reported Device Performance
    Temperature exiting canister similar to PredicateEvaluation demonstrated the temperature exiting from the canister is similar to the Predicate.
    Biocompatibility per ISO 10993-1 (for compromised skin, <24h): - Cytotoxicity - Sensitization - Irritation - Acute Systemic Toxicity - Material Mediated PyrogenicityAll test criteria were successfully met for each evaluation.
    Stability for stated shelf-life limits (24 months): - Visual Inspection - Total Delivery - Pressure - Spray Rate - Leakage/Weight Loss - MicrobiologicalTest criteria were met at each assessment interval, which demonstrated the Subject Device's stability.
    Non-flammableNon-flammable (listed as a similarity to predicate)
    Mechanism has positive shut-off releaseMechanism has positive shut-off release (listed as a similarity to predicate)

    2. Sample sizes used for the test set and the data provenance:

    • Sample Size for Test Set: Not specified. The document refers to "evaluations" and "assessments" without quantification of the number of units or tests performed.
    • Data Provenance: Not specified. These are laboratory-based non-clinical tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable and not provided. The ground truth for this device's "performance" is based on physical and chemical testing standards and comparison to a predicate, not expert consensus on diagnostic interpretations.

    4. Adjudication method for the test set:

    Not applicable and not provided. Similarly to point 3, there's no diagnostic component requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a vapocoolant, not a diagnostic AI system or a device intended to assist human readers. Therefore, no MRMC study was conducted. The document explicitly states "Clinical Testing: Not Applicable."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used:

    The ground truth used for demonstrating the device's characteristics relies on:

    • Physical and chemical measurements (e.g., temperature exiting canister, pressure, spray rate, leakage, boiling point).
    • Standardized biocompatibility testing (ISO 10993-1).
    • Comparison to the characteristics of the legally marketed predicate device.

    8. The sample size for the training set:

    Not applicable. This is a physical device, not an AI/ML algorithm that requires a "training set."

    9. How the ground truth for the training set was established:

    Not applicable. See point 8.