    Intended Use

    The product is indicated for use on moderate to highly exuding acute and chronic wounds such as:

    • · Diabetic ulcers
    • · Leg ulcers (arterial ulcers, venous ulcers and leg ulcers of mixed etiology)
    • · Pressure injuries (stage 2-4)
    • Traumatic wounds
    • · Partial thickness (second degree) burns
    • · Donor sites
    • · Post-operative surgical wounds
    • · Exudate absorption in oncology wounds, such as fungoides-cutaneous tumors, fungating carcinoma, cutaneous metastasis, Kaposi's sarcoma, and angiosarcoma
    Device Description

    Biatain Fiber Ag is a soft, sterile, primary wound dressing made mainly from non-woven gelling carboxymethyl cellulose (CMC) fibers. In addition, non-gelling bicomponent (BiCo) fibers are incorporated to obtain strength and integrity of the products both in dry and gelled condition. Ionic silver (Ag+) is added for antimicrobial efficacy within the dressing, and ethyl lauroyl arginate (LAE) is added as combined surfactant and chelator within the dressing.

    AI/ML Overview

    The provided text is a 510(k) summary for the Biatain Fiber Ag wound dressing. This document describes the device, its indications for use, comparison with a predicate device, and performance data to support its substantial equivalence. However, it does not contain information about an AI/ML-enabled medical device study, nor does it describe acceptance criteria and performance in the context of AI/ML algorithms.

    Therefore, I cannot provide details on the following, as they are not present in the document:

    • A table of acceptance criteria and reported device performance related to AI/ML.
    • Sample size used for the test set and data provenance.
    • Number of experts and their qualifications for ground truth.
    • Adjudication method for the test set.
    • MRMC comparative effectiveness study or human reader improvement with AI.
    • Standalone algorithm performance.
    • Type of ground truth used for AI/ML.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    Instead, the document details the performance data for a traditional wound dressing:

    1. Performance Data Provided (Related to a traditional wound dressing):

    Test CategoryAcceptance Criteria/MethodReported Device Performance
    BiocompatibilityConformance with ISO 10993-1:2018Concluded that the dressings have a safe toxicological profile for their intended use. Battery of tests included: Cytotoxicity, Irritation, Sensitization, Acute systemic toxicity, Sub-acute systemic toxicity, Pyrogenicity, Implantation.
    Physical/Mechanical PropertiesNot explicitly stated "acceptance criteria" values, but testing was conducted on key performance parameters.Absorption capacity, Fluid retention capacity, Wet strength (gelled condition), Lateral shrinkage upon absorption, Volume expansion upon absorption, Lateral spreading of fluid. (No specific numerical results provided in the summary).
    Antimicrobial Effectiveness within the Dressing≥4 log reduction (AATCC 100) antimicrobial effectiveness against specified organisms for up to 7 days.Confirmed by in vitro tests. Achieved ≥4 log reduction (AATCC 100) against 9 organisms (3 Gram positive bacteria, 4 Gram negative bacteria, 1 yeast, and 1 mold) for up to 7 days within the dressing.
    Animal StudyWound healing performance not hindered.A full thickness wound healing model in porcine study demonstrated that Biatain Fiber Ag and the predicate device did not hinder the wound healing process.

    Summary of Device and Study Type:

    The document describes performance data for the Biatain Fiber Ag, which is a wound dressing, not an AI/ML device. The studies listed are standard performance tests for such a product to demonstrate its safety and effectiveness in its intended use, particularly for absorbency, wet strength, and antimicrobial properties.