510(k) Data Aggregation

    K Number
    K231917

    Validate with FDA (Live)

    Device Name
    VEA Align
    Manufacturer
    Date Cleared
    2024-01-05

    (190 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This cloud-based software is intended for orthopedic applications in both pediatric and adult populations. 2D X-ray images acquired in EOS imaging's imaging systems is the foundation and resource to display the interactive landmarks overlayed on the frontal and lateral images. These landmarks are available for users to assess patient-specific global alignment. For additional assessment, alignment parameters compared to published normative values may be available. This product serves as a tool to aid in the analysis of spinal deformities, degenerative diseases, lower limb alignment disorders, and deformities through precise angle and length measurements. It is suitable for use with adult and pediatric patients aged 7 years and older. Clinical judgment and experience are required to properly use the software.

    Device Description

    VEA Align is a software indicated for assisting healthcare professionals with global alignment assessment through clinical parameters computation. The product uses biplanar 2D X-ray images, exclusively generated by EOS imaging's EOS (K152788) and EOSedge (K202394) systems and generates an initial placement of the patient anatomic landmarks on the images using a machine learning-based algorithm. The user may adjust the landmarks to align with the patient's anatomy. Landmark locations require user validation. The clinical parameters communicated to the user are inferred from the landmarks and are recalculated as the user adjusts the landmarks. The product is hosted on a cloud infrastructure and relies on VEA Portal for support capabilities. such as user access control and data access. 2D X-ray image transmissions from healthcare institutions to the cloud are managed by VEA Portal is a Class I 510(k)-exempt device (LMD).

    AI/ML Overview

    The provided text describes the VEA Align device and its performance testing to support its substantial equivalence to a predicate device. However, it does not contain a detailed table of acceptance criteria with reported device performance metrics that would typically be found in a comprehensive study report. It states that "Direct comparison between skeletal landmark locations between the subject VEA Align device and predicate sterEOS Workstation (K172346) met acceptance criteria for algorithm performance," but it does not quantify these criteria or the specific performance results.

    Therefore, some of the requested information cannot be directly extracted from the provided text. I will provide what is available and note what is missing.

    Here's the breakdown of the information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states: "Direct comparison between skeletal landmark locations between the subject VEA Align device and predicate sterEOS Workstation (K172346) met acceptance criteria for algorithm performance." However, the specific quantitative acceptance criteria (e.g., maximum allowable error for landmark placement) and the actual numerical performance results (e.g., mean absolute error) are not provided in this text.

    Acceptance CriteriaReported Device Performance
    Not specified quantified acceptance criteria for landmark location comparison.Met acceptance criteria for algorithm performance for direct comparison between skeletal landmark locations and the predicate device. Specific metrics (e.g., mean error, standard deviation) are not provided.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 555 patients.
    • Data Provenance: The images were acquired from "EOS (K152788) and EOSedge (K202394) systems." The country of origin and whether the data was retrospective or prospective are not explicitly stated.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the text. The document refers to the predicate device manually deforming a 3D model through control points to match X-ray contours, which implies expert interaction in the past, but it does not describe how ground truth was established for the 555-patient test set for the VEA Align device.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not provided in the text.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs. without AI assistance.

    A MRMC comparative effectiveness study involving human readers with and without AI assistance is not mentioned in the provided text. The performance testing focuses on the standalone algorithm's comparison to the predicate device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Yes, a standalone performance assessment was done. The text states:
    "Standalone performance assessment of the machine learning algorithm. The testing dataset consisted of 555 patients... Direct comparison between skeletal landmark locations between the subject VEA Align device and predicate sterEOS Workstation (K172346) met acceptance criteria for algorithm performance."

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the standalone performance assessment appears to be based on the "skeletal landmark locations" derived from the predicate sterEOS Workstation (K172346). This implies that the predicate's output, which involved manual deformation by users ("The 3D model is deformed manually by the user through control points up to matching accurately the X-ray contours. This deformation is performed by using the common linear least squares estimation algorithm."), served as the reference for the VEA Align's automated landmark placement. It is not explicitly stated that an independent expert consensus or pathology was used directly for the 555-patient test set for the standalone evaluation of VEA Align, but rather conformance to the predicate's output.

    8. The Sample Size for the Training Set

    The sample size for the training set is not explicitly stated in the provided text. It mentions that the machine learning algorithm was "trained from data generated by EOS Imaging's imaging systems", but it doesn't quantify the size of this training dataset.

    9. How the Ground Truth for the Training Set Was Established

    The text states that the machine learning algorithm learns to generate "an initial placement of the patient anatomic landmarks on the images" and that "The user may adjust the landmarks to align with the patient's anatomy." For the predicate device, it mentions "identification of anatomical landmarks" or "a model of bone structures derived from an a priori image data set from 175 patients (91 normal patients, 47 patients with moderate idiopathic scoliosis and 37 patients with severe idiopathic scoliosis), and dry isolated vertebrae data for spine modeling."

    While it implies that human interaction and potentially pre-existing models established the ground truth used for training, the specific methodology and who established the ground truth labels for the VEA Align training set are not detailed. It implies the machine learning was "trained from data generated by EOS Imaging's imaging systems," which suggests leveraging existing data from their systems and prior approaches (potentially like the predicate).