510(k) Data Aggregation

    K Number
    K231617

    Validate with FDA (Live)

    Date Cleared
    2023-11-09

    (160 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MAGNETOM Free.Max system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal, and oblique cross-sectional images that display the internal structure and/or function of the head. body, or extremities. Other physical parameters derived from the images may also be produced. Depending on the region of interest, contrast agents may be used. These images and the physical parameters derived from the images when interpreted by a trained physician vield information that may assist in diagnosis.

    MAGNETOM Free.Max may also be used for imaging during interventional procedures when performed with MR-compatible devices such as MR Safe biopsy needles.

    MAGNETOM Free.Star system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal, and oblique cross-sectional images that display the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images may also be produced. Depending on the region of interest, contrast agents may be used. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist in diagnosis.

    Device Description

    MAGNETOM Free.Max and MAGNETOM Free.Star with syngo MR XA60A include new and modified features compared to the predicate devices MAGNETOM Free.Max and MAGNETOM Free.Star with syngo MR XA50A (K220575, cleared on June 24, 2022).

    Below is a high-level summary of the new and modified hardware and software features compared to the predicate devices MAGNETOM Free.Max and MAGNETOM Free.Star with syngo MR XA50A:

    Hardware
    New hardware features:

    • Contour Knee coil
    • Respiratory Sensor

    Modified hardware features:

    • myExam 3D Camera
    • Host computer
    • MaRS

    Software
    New Features and Applications:

    • Injector coupling
    • Respiratory Sensor Support
    • myExam RT Assist (only for MAGNETOM Free.Max)
    • myExam Autopilot Hip
    • Deep Resolve Boost
    • Complex Averaging
    • HASTE_Interactive (only for MAGNETOM Free.Max)
    • BEAT_Interactive (only for MAGNETOM Free.Max)
    • Needle Intervention AddIn (only for MAGNETOM Free.Max)

    Modified Features and Applications:

    • Deep Resolve Sharp
    • Deep Resolve Gain
    • SMS Averaging

    Other Modifications:

    • Indications for Use (only for MAGNETOM Free.Max)
    • MAGNETOM Free.Max RT Edition marketing bundle (only for MAGNETOM Free.Max)
    AI/ML Overview

    The provided text describes information about the submission of the MAGNETOM Free.Max and MAGNETOM Free.Star MRI systems for FDA 510(k) clearance, and references a specific AI feature called "Deep Resolve Boost." However, it does not contain acceptance criteria or a detailed study proving the device meets specific performance criteria for the AI feature.

    The section titled "Test statistics and test results" for Deep Resolve Boost (Table 1, page 7) mentions that the impact of the network was characterized by quality metrics such as PSNR and SSIM, and visual inspection. It also states: "After successful passing of the quality metrics tests, work-in-progress packages of the network were delivered and evaluated in clinical settings with cooperation partners." This suggests internal testing and evaluation, but does not provide the specific numerical acceptance criteria or the detailed results of these tests.

    Therefore, I cannot fully complete the requested table and answer all questions due to the lack of this specific information in the provided document.

    However, I can extract the available information regarding the AI feature "Deep Resolve Boost" as much as possible:

    1. Table of acceptance criteria and the reported device performance:

    Metric / CriteriaAcceptance Criteria (Stated or Implied)Reported Device Performance (Specifics not provided in document)
    Deep Resolve Boost
    Peak Signal-to-Noise Ratio (PSNR)Must pass initial quality metrics tests.Quantified, but specific numerical values are not reported.
    Structural Similarity Index (SSIM)Must pass initial quality metrics tests.Quantified, but specific numerical values are not reported.
    Visual Inspection for ArtifactsMust ensure potential artifacts are detected that are not well captured by PSNR/SSIM.Images visually inspected.
    Clinical EvaluationMust be evaluated in clinical settings with cooperation partners."work-in-progress packages of the network were delivered and evaluated in clinical settings with cooperation partners." (No specific results or findings reported in this document.)

    2. Sample size used for the test set and the data provenance:

    • Test Set (Validation set for AI feature Deep Resolve Boost):
      • Sample Size: 1,874 2D slices.
      • Data Provenance: "in-house measurements and collaboration partners." The document does not specify the country of origin.
      • Retrospective or Prospective: Retrospectively created from ground truth by data manipulation and augmentation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of experts: Not specified.
    • Qualifications of experts: The document states the "acquired datasets represent the ground truth for the training and validation," but it does not specify how this ground truth was established in terms of expert involvement for the test set. It mentions "clinical settings with cooperation partners" for evaluation, but this is distinct from ground truth establishment.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not specified. The document states "acquired datasets represent the ground truth," suggesting pre-existing data or a different method of ground truth establishment than explicit reader adjudication for this AI feature.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • The document states "No clinical tests were conducted to support substantial equivalence for the subject device" (page 10). It mentions that "work-in-progress packages of the network were delivered and evaluated in clinical settings with cooperation partners," but this is not described as an MRMC comparative effectiveness study, nor are any results on human reader improvement reported.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • The performance of the "Deep Resolve Boost" AI feature was characterized by "quality metrics such as peak signal-to-noise ratio (PSNR) and structural similarity index (SSIM)" and visual inspection, which suggests a standalone evaluation of the algorithm's output against a reference standard. Specific results are not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For Deep Resolve Boost: "The acquired datasets represent the ground truth for the training and validation." Input data for training was "retrospectively created from the ground truth by data manipulation and augmentation." This implies that high-quality, likely clinical-grade, MRI scans acquired without the AI feature were considered the "ground truth" to which the AI-processed images were compared. It's not explicitly stated if this "ground truth" itself was established by expert consensus, but it infers it from high-quality clinical acquisition.

    8. The sample size for the training set:

    • For Deep Resolve Boost: 24,599 2D slices.

    9. How the ground truth for the training set was established:

    • "The acquired datasets represent the ground truth for the training and validation. Input data was retrospectively created from the ground truth by data manipulation and augmentation. This process includes further under-sampling of the data by discarding k-space lines, lowering of the SNR level by addition of noise and mirroring of k-space data."
    • This indicates that "ground truth" was established by using full, high-quality MR images. The "input data" for the AI model (which the AI then "boosts") was intentionally degraded (under-sampled, noised) from this high-quality ground truth. The AI's task is to reconstruct the degraded input data back to resemble the original high-quality "ground truth."