Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K231471

    Validate with FDA (Live)

    Date Cleared
    2023-09-08

    (109 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Air/Water Bottle Tubing is to connect an air/CO2 source, a sterile water source (water bottle), and an endoscope to supply air/CO2 and water during gastrointestinal endoscopic procedures. This device is intended to be used for not more than 24 hours. It is a 24-hour multi-patient use device.

    The CO2 Source Tubing is intended to be used with a carbon dioxide (CO2) source with the purpose of supplying CO2 to the endoscope during endoscopic procedures. This device is intended to be used for not more than 24 hours. It is a 24-hour multi-patient use device.

    Device Description

    The proposed device includes two (2) categories:

    • Air/Water Bottle Tubing
    • CO2 Source Tubing

    The Air/Water Bottle Tubing and CO2 Source Tubing are intended for 24-hour multi-patient use. Air/water bottle tubing is supplied in sterile and CO₂ Source Tubing is supplied in non-sterile.

    6.1 The Air/Water Bottle Tubing
    The Air/Water Bottle Tubing is manufactured for use in conjunction with a sterile water bottle, and together with OLYMPUS®140/240, 160/260, 180/280, 190/290 series endoscopes. The Air/Water Bottle Tubing is individually packed in sealed package, sold as a sterile device. The Air/Water Bottle Tubing is designed to be attached to the air/water port of the endoscopes to provide irrigation through the air/water channel to the distal end of endoscope.

    There are two types of the device Air/Water Bottle Tubing provided by the manufacturer, such as type A and type B. Type B has an additional air extended tubing but type A does not. The clinician selects the appropriate model according to the actual clinical needs.

    6.2 The CO2 Source Tubing
    The CO2 Source Tubing is manufactured for use in conjunction with air/water bottle tubing and CO2 insufflator. The CO2 Source tubing is individually packed in a sealed package, and sold as a non-sterile device. The CO2 Source Tubing is designed to be attached to the air/water bottle tubing and the outlet of the CO2 insufflator to provide irrigation and CO2 insufflation through the air/water channel to the distal end of endoscope.

    AI/ML Overview

    The provided FDA 510(k) summary (K231471) describes the Chuangxiang Medical Technology Co., LTD.'s Air/Water Bottle Tubing and CO2 Source Tubing. This submission focuses on non-clinical performance data to demonstrate substantial equivalence to predicate devices, rather than a clinical study involving human subjects or complex AI algorithms. Therefore, many of the requested categories related to clinical studies, AI performance metrics, and expert adjudication are not applicable to this document.

    Here's a breakdown based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document details various performance tests conducted on the devices. It claims these tests successfully demonstrated essential performance and substantial equivalence to predicate devices, implying that the devices met pre-defined acceptance criteria for these tests. However, the specific quantitative acceptance criteria or detailed numerical results for each test are not provided in this summary. Instead, the summary states that the testing "successfully demonstrated essential performance" and that the manufacturer "believes that no significant differences exist."

    Acceptance Criteria CategoryReported Device Performance (Summary)
    Air/Water Bottle Tubing
    Assembling IntegritySuccessfully demonstrated essential performance
    Two-Way Valve IntegritySuccessfully demonstrated essential performance
    Air Flow TestSuccessfully demonstrated essential performance
    Water Flow TestSuccessfully demonstrated essential performance
    Air Leak TestSuccessfully demonstrated essential performance
    Flow Clamp TestSuccessfully demonstrated essential performance
    Water Backflow TestSuccessfully demonstrated essential performance
    24-Hour Use TestSuccessfully demonstrated essential performance
    Endoscope CompatibilitySuccessfully demonstrated essential performance
    Compatibility With BottleSuccessfully demonstrated essential performance
    CO2 Source Tubing
    Assembling IntegritySuccessfully demonstrated essential performance
    CO2 Delivery TestSuccessfully demonstrated essential performance
    Water Delivery TestSuccessfully demonstrated essential performance
    Air Leak TestSuccessfully demonstrated essential performance
    Compatibility With Air/Water Bottle TubingSuccessfully demonstrated essential performance
    SterilizationEO sterilization cycle achieved SAL of 10^-6; EO residuals below ISO 10993-7:2008 limits.
    Shelf LifeThree (3) years shelf life validated per ASTM F1980-21 and ISO 11607-1/2:2019 standards; essential performance achieved before and after.
    BiocompatibilityAll patient contacting materials meet ISO 10993-1:2009 and FDA guidance, including Acute Systemic Toxicity, In Vitro Cytotoxicity, Skin Sensitization, and Intracutaneous Reactivity. Devices classified as surface device with mucosal membrane contact for limited duration.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document states that performance tests were performed on "samples from initial lots." However, the specific sample size for each test or for the overall test set is not quantified.

    The data provenance is retrospective, as the testing was conducted by the manufacturer (Zhejiang Chuangxiang Medical Technology Co., LTD. in China) on their newly manufactured devices to support the 510(k) submission. The country of origin of the data is China.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not Applicable. This submission relies on non-clinical, bench testing, not expert interpretation of medical images or patient data to establish ground truth. The "ground truth" for these tests would be the measurement against established engineering or performance specifications.

    4. Adjudication Method for the Test Set

    Not Applicable. As this is non-clinical bench testing, there is no need for expert adjudication of results in the way it would be applied to a clinical study involving human interpretation. The results are based on objective measurements and predefined performance parameters.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. This submission is for medical device accessories (tubing) and does not involve an AI algorithm or human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not Applicable. This submission does not involve an AI algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for these non-clinical tests is based on engineering specifications, industry standards (e.g., ISO, ASTM), and predefined performance parameters for air/water flow, leak integrity, compatibility, sterilization, shelf life, and biocompatibility.

    8. The Sample Size for the Training Set

    Not Applicable. There is no AI algorithm involved, so no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. There is no AI algorithm involved, so no need to establish ground truth for a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1