    Intended Use

    The Intimate Rose Vaginal Dilators are tools intended to dilate the vagina in controlled stages to help relieve symptoms of vaginismus.

    Device Description

    The Intimate Rose Vaginal Dilators are intended to treat women suffering from vaginismus (including related dyspareunia). Vaginismus is the involuntary spasm of the muscles in the vaginal wall which then inhibits sexual intercourse by making it painful or impossible. Dyspareunia is the pain experienced during sexual intercourse caused by physical and/or emotional problems. The device is used as a tool to dilate the vagina in controlled stages. It is a reusable device that comes into contact with the vaginal mucosal membrane for a prolonged (> 24 hours to ≤ 30 days) duration. The Intimate Rose Vaginal Dilators are comprised of 8 progressively larger and color-coded medical-grade silicone dilators.

    AI/ML Overview

    The provided document is a 510(k) summary for the Intimate Rose Vaginal Dilators. It describes the device, its intended use, and compares it to a predicate device to demonstrate substantial equivalence. However, it does not contain information about acceptance criteria or a study that typically involves human readers or an AI algorithm’s performance.

    This document describes a device approval process based on non-clinical performance testing for a physical medical device (vaginal dilators), not an AI/ML-driven device. Therefore, the requested information regarding acceptance criteria, human reader studies, AI assistance, ground truth, and training set details for AI/ML performance is not present in the provided text.

    Based on the document, I can only provide information related to the device's physical and biological properties.

    Here's what can be extracted from the document regarding the device's evaluation, focusing on the non-clinical tests performed:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" for each performance test in a quantitative manner. Instead, it states that the tests demonstrate the device is "acceptable" or "non-" (e.g., non-cytotoxic). The reported performance is the "test results" and the conclusion.

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility:
    Non-cytotoxic (per ISO10993-5:2009)The subject devices were found to be non-cytotoxic.
    Non-sensitizing (per ISO10993-10:2010)The subject devices were found to be non-sensitizing.
    Non-irritating (Vaginal irritation testing per ISO10993-10:2010)The subject devices were found to be non-irritating.
    Not causing acute systemic toxicity (per ISO10993-11:2017)The subject devices were found not to be causing acute systemic toxicity.
    Reprocessing:
    Adequate reprocessing information providedDevice labeling included reprocessing information per FDA's 2015 guidance document.
    Physical Properties:
    Acceptable Appearance (No specific threshold given)Evaluated (Implied satisfactory outcome as no issues raised for substantial equivalence)
    Acceptable Diameter (No specific threshold given)Evaluated (Implied satisfactory outcome as no issues raised for substantial equivalence)
    Acceptable Hardness (No specific threshold given)Evaluated (Implied satisfactory outcome as no issues raised for substantial equivalence)
    Acceptable Compression (No specific threshold given)Evaluated (Implied satisfactory outcome as no issues raised for substantial equivalence)
    Acceptable Density (No specific threshold given)Evaluated (Implied satisfactory outcome as no issues raised for substantial equivalence)

    2. Sample size used for the test set and the data provenance:

    • Sample size: The document states that "all 8 colors of the subject device" were tested for biocompatibility and physical properties. It does not provide a specific number of individual dilator units tested for each characteristic.
    • Data provenance: Not explicitly stated as "country of origin" or "retrospective/prospective." The testing was non-clinical laboratory testing performed to international and FDA guidance standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This question is not applicable. The device is a physical medical device, not an AI/ML diagnostic or interpretive tool that requires expert-established ground truth on clinical data for performance evaluation in the context of this 510(k) submission. The "ground truth" for these tests would be the established biological or physical standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This question is not applicable. Adjudication methods are typically for subjective clinical assessments or image interpretations, which are not part of this device's non-clinical performance testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This question is not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

    This question is not applicable. This is not an AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the non-clinical tests described:

    • Biocompatibility: The "ground truth" is defined by the international standards (ISO 10993 series) for toxicity, sensitization, irritation, and systemic effects.
    • Physical Properties: The "ground truth" for physical characteristics like diameter, hardness, compression, and density would be the device's design specifications and material properties, compared against measured values.

    8. The sample size for the training set:

    This question is not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established:

    This question is not applicable. This is not an AI/ML device.