Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K231238

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-10-31

    (186 days)

    Product Code
    Regulation Number
    878.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultravision2™ System is indicated for the clearance of smoke and other particulate matter that is created during laparoscopic surgery.

    The Ultravision2™ 5mm Trocar component establishes a path of entry for instruments used in laparoscopic surgery.

    The Ultravision2™ Generator interfaces directly with the electrosurgical generator and serves as a pass-through for RF energy to RF electrosurgical instruments.

    Device Description

    The Ultravision2™ System is a multifunctional system that synchronizes visual field clearing with the activation of smoke-producing electrosurgical devices. The system interfaces with commercially available electrosurgical instruments. The Ultravision2™ Generator connects directly to a commercially available electrosurgical generator (ESU) and passes the RF energy through to the desired electrosurgical instrument connected to the Utravision2™ Generator. The Ultravison2™ System is able to automate the activation of the Ionwand for visual field clearing to the activation of the electrosurgical device to synchronize visual field clearing with the generation of smoke. The Ionwand™ pack comprises a dedicated percutaneous 3mm trocar/catheter which accommodates the Ionwand™ cable that delivers low energy from the generator to the patient. The Ultravision™ 5mm Trocar includes a dedicated Ionwand™ cable. This device is for prescription use only.

    AI/ML Overview

    The information provided does not contain a study that proves the device meets the acceptance criteria. It presents a summary of non-clinical testing performed, including acceptance criteria and results (all "Pass"), but does not detail the methodology, sample sizes, or ground truth establishment for these tests in a way that aligns with a typical performance study.

    Based on the provided text, here's what can be extracted:

    1. Table of acceptance criteria and the reported device performance:

    Test PerformedAcceptance CriteriaReported Device Performance
    Shelf lifeProduct and package must demonstrate stability for the claimed shelf life of 5 years.Pass
    Software verification and validationDevice functions controlled by software must perform as intendedPass
    Electrical safety and electromagnetic compatibilityDevice must meet the requirements of the applicable clauses in the standards (IEC 60601-1, IEC 60601-2-2, EN 60601-1-2)Pass
    Dimensional and physical verification of unitDevice must meet dimensional specification and physical specifications as per internal standards.Pass
    High voltage outputDevice must deliver specified output, into required loads, at given temperatures and humidities, and for the specified product lifetime as per internal standards.Pass
    Generator basic function including:Device must meet specifications for connections and functionality as per internal standards.Pass
    Generator safety measures including:Must meet specifications for device function independent of software, extreme misuse, or single fault conditions as per internal standards.Pass
    Generator user interface hardware controlMust meet specifications for device function independent of software, extreme misuse, or single fault conditions as per internal standards. (Note: Same criterion as "Generator safety measures including:")Pass
    Generator high voltage power managementMust meet specifications for DC output under proximity and HVDC limit protection measure as per internal standards.Pass
    Generator surgical energy detectionDevice must demonstrate energy detection for external ultrasonic energy as per internal standards.Pass
    Mechanical connections and controlsDevice must demonstrate acceptable durability of link cables, fascia connections, and cable retention as per internal standards.Pass
    Surgical generator compatibilityDevice must demonstrate compatibility with applicable generators in terms of load curve characterization, CQM performance, HF leakage, and RF detection. as per internal standards.Pass
    Design validation under simulated use conditionsMust demonstrate that the device can achieve its intended use when used by end users as per internal standards.Pass

    2. Sample sized used for the test set and the data provenance:

    The document summarizes "non-clinical testing" and lists various tests, but it does not specify sample sizes for any of these tests. It also does not provide information on data provenance (e.g., country of origin, retrospective or prospective nature) as these are non-clinical engineering and performance characterization tests, not studies involving patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided as the disclosed tests appear to be engineering and design validation tests rather than clinical studies requiring expert ground truth for interpretation (e.g., image analysis by radiologists).

    4. Adjudication method for the test set:

    This information is not provided for the same reasons as above. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving multiple expert readers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted or described in the provided text. The device is a "Surgical Smoke Precipitator" and an "Ultravision2™ System" which focuses on clearing smoke during laparoscopic surgery, not an AI diagnostic or assistive device for human readers. Therefore, the concept of "how much human readers improve with AI vs without AI assistance" is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The Ultravision2™ System is a medical device for surgical smoke clearance, not an algorithm. Therefore, "standalone (algorithm only)" performance is not applicable. The device's function is physical and mechanical/electrical in nature for smoke precipitation synchronization.

    7. The type of ground truth used:

    The "ground truth" for the reported tests appears to be engineering specifications and established standards (e.g., ASTM, IEC, internal standards). For example, for "Dimensional and physical verification," the ground truth is "dimensional specification and physical specifications as per internal standards." For "Electrical safety," it's "the requirements of the applicable clauses in the standards."

    8. The sample size for the training set:

    This information is not applicable as the device is not an AI/machine learning system that requires a training set in the typical sense.

    9. How the ground truth for the training set was established:

    This information is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1