510(k) Data Aggregation

    K Number
    K231087

    Validate with FDA (Live)

    Date Cleared
    2023-08-16

    (121 days)

    Product Code
    Regulation Number
    872.4120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Guided Surgery Kit is designed to hold various dental surgical drills and instruments to organize, steam sterilize, and transport between uses. The guided surgical drills are intended to cut into maxilla or mandible to create an osteotomy for endosseous dental implant placement.

    The kit is to be enclosed in a FDA cleared steam sterilizable wrap (maximum thickness KC300) and sterilized in a FDA cleared sterilizer for one of the following cycles:

    (1) Prevacuum Steam – At 132°C for 4 minutes with a 20 minute dry time.
    (2) Gravity Steam At 132°C for 15 minutes with a 30 minute dry time.
    · The kit is intended for sterilization of non-porous loads.
    · Do not stack kits during sterilization.
    · Implant Direct Sybron Manufacturing LLC does not make any lumen claims for the Sterilizable Guided kit.

    Device Description

    The Guided Surgery Kit is offered in 2 complete kit variations for 3 implant systems: Legacy, Simply lconic, and InterActive as a reusable perforated tray for purposes of transport, steam sterilization and storage of dental instruments, similar to the reference device Standard Sterilizable Tray (K202524). The kit is sold non-sterile and contains site preparation instruments, dental driving tools, prosthetic driving tools and ratchet tool which can be used for implant placement. The proposed Guided Surgical Drills are reusable surgical instruments designed to prepare an osteotomy for a dental implant procedure. The features remain unchanged from the predicate Surgical Drills (K200265) except for the addition of a Guide Body and Shoulder Stop. The addition of a Guide Body allows the proposed drills to function with guide sleeves which can be integrated into an existing surgical quide template to assist in the drilling sequence and placement of dental implants.

    AI/ML Overview

    The provided document, an FDA 510(k) clearance letter for a "Guided Surgery Kit," does not describe a study involving an AI/Artificial Intelligence device that requires the detailed information requested in the prompt.

    The device in question, a "Guided Surgery Kit," is a collection of dental surgical drills and instruments to organize, steam sterilize, and transport between uses, including drills intended to cut into maxilla or mandible for dental implant placement. The document focuses on demonstrating substantial equivalence to predicate devices through comparisons of indications for use, technological characteristics, and performance data, primarily regarding biocompatibility, cleaning, and sterilization validation.

    Therefore, many of the requested categories related to AI/ML device performance studies, such as "acceptance criteria for device performance," "sample sizes for test set," "number of experts for ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," "type of ground truth," "sample size for training set," and "ground truth for training set," are not applicable to the content of this regulatory submission.

    The document explicitly states: "Clinical performance data is not required to establish substantial equivalence for the subject device." This further indicates that no human-in-the-loop or AI-specific performance study was conducted or reported in this submission for this particular device.