Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K230730

    Validate with FDA (Live)

    Device Name
    Apyx
    Manufacturer
    Date Cleared
    2023-07-25

    (131 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apyx is intended for attaching sutures to ligaments of the pelvic floor.

    Device Description

    The Apyx device is indicated for anchoring sutures to ligaments of the pelvic floor. The device consists of an implantable nitinol anchor with 4 prongs configured with either non-absorbable or resorbable suture. The anchor-suture assembly is contained within a cartridge, wherein the anchor / suture assembly is deployed to the target site from the cartridge with an applicator. An optional use retriever may be used to remove the anchor in the event of sub-optimal placement of an Anchor during the index procedure. The Apyx also includes an optional securement element which may be used as a suture retention device to distribute suture tension over a larger tissue area and aid in wound healing.

    The Apyx device is a single use device which is supplied sterile.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a study proving the device meets those criteria in the typical format of a clinical or performance study with specific metrics like sensitivity, specificity, or AUC, as these would be associated with diagnostic or AI-powered devices.

    However, based on the information provided, here's a breakdown of what can be extracted regarding performance and testing:

    1. A table of acceptance criteria and the reported device performance

    The document mentions various tests conducted, generally implying that the device met the requirements of these tests to demonstrate safety and effectiveness. Since specific numerical acceptance criteria (e.g., "pull-out strength > X N") and the exact reported performance values are not explicitly stated, I will infer the acceptance as "Met requirements" based on the conclusion of substantial equivalence.

    Acceptance Criterion (Implied)Reported Device Performance
    BiocompatibilityMet requirements (ISO 10993 compliant)
    Chemical Characterization and Toxicological Risk AssessmentMet requirements (ISO 10993-18 compliant)
    Corrosion ResistanceMet requirements
    Packaging ValidationMet requirements
    Shelf-Life ValidationMet requirements
    Securement Element Pull-OutMet requirements
    Cadaver Model Testing (Delivery)All anchors accurately delivered into target ligament; no complications or unanticipated risks observed. Successfully performed.
    Cadaver Model Testing (Retrieval)Successfully performed. Successfully removed.
    Cadaver Model Testing (Securement)All anchors secured using the securement element and properly fixed at the desired position with ease.

    2. Sample size used for the test set and the data provenance

    • Sample size for test set: The "Cadaver Testing" section does not specify the number of cadavers or the number of anchors tested. It uses general phrasing like "users," "all anchor delivery and retrieval procedures," and "All anchors."
    • Data provenance: Not explicitly stated, but cadaver testing implies a controlled laboratory environment. The country of origin of the data is not mentioned. It is prospective testing carried out for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided. The cadaver testing describes the actions of "users" (implying medical professionals) who delivered and retrieved anchors, but doesn't mention independent expert assessment or ground truth establishment in a formalized way for outcomes like accuracy or safety observed during the testing. The "ground truth" here is the observed outcome of the procedure itself (e.g., successful placement, no damage).

    4. Adjudication method for the test set

    • Not applicable/Not described. The cadaver testing describes qualitative observations of successful delivery, retrieval, and proper fixation. There's no mention of an adjudication process for resolving disagreements between multiple observers, as might be seen in diagnostic studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This device is a surgical instrument (fixation device), not an imaging or AI-assisted diagnostic tool. Therefore, an MRMC study and AI-related effectiveness are not relevant or discussed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This device is a manual instrument for surgical use, not an algorithm.

    7. The type of ground truth used

    • For the cadaver testing, the ground truth was the direct observation of the physical outcome and success of the procedure (e.g., "anchors were accurately delivered," "no damage observed," "properly fixed"). This is akin to direct experimental observation and physical verification rather than consensus, pathology, or outcomes data in a clinical trial.

    8. The sample size for the training set

    • Not applicable. This device is a physical medical instrument, not a machine learning algorithm, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for a physical medical instrument.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1