Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K230520

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-03-21

    (22 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The display is intended to be used for displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners.
    The display is intended to be used for displaying and viewing digital images (excluding digital mammography) for review and analysis by trained medical practitioners. The display may be used in dental applications.

    Device Description

    Both the Nio Color 2MP (MDNC-2521) and the Nio Color 3MP (MDNC-3521) are medical computer displays designed for general radiology imaging applications.
    The MDNC-3521 model can also be used in dental applications.
    The devices can also be used for home reading in radiology.

    The MDNC-2521 is a derivative of the MDNC-2221.
    The MDNC-3521 is a derivative of the MDNC-3421.

    The modified displays are effectively identical to the respective predicate devices except for the following changes:
    Updated LCD panel with same resolution and dimensions compared to the respective predicate devices New housing, display stand and internal mechanics, with similar functionality and design principle compared to the respective predicate devices Updated internal electronics boards, with similar functionality and design principle compared to the respective predicate devices Updated firmware, with similar functionality and design principle compared to the respective predicate devices New packaging, with similar functionality and design principle compared to the respective predicate devices or any other Barco diagnostic display Rephrasing of the intended usage environment, because of the evolution towards more home reading in radiology Small update in the intended user description The modified device has the following similarities compared to the unmodified device: The same intended use The same operating principle The same fundamental technology

    The displays can be used optionally with QAWeb Enterprise software, listed under D332294 as a class 1 device with product code LHO. QAWeb Enterprise is a calibration software that is intended as a quality assurance software for the displays. QAWeb Enterprise software helps to keep the display DICOM compliant.

    The display can be used optionally with Intuitive Workflow Tools, cleared in K191845 as a class 2 device with product code PGY. The Intuitive Workflow Tools are accessories for image enhancement on diagnostic displays: SpotView: The Barco SpotView display feature allows focusing on a region of interest in an image by boosting the display's backlight such that the maximum luminance in provided inside the region of interest. SpotView also enables focused observation during reading by dimming images outside the region of interest and increasing the contrast in this region. Also magnification and inversion of pixels are possible with SpotView. AAM - Application Appearance Manager: This workflow tool allows you to set the luminance as well as the color space for each application that is on the workstation. There are often multiple windows open on a screen, but not all of them need the high brightness of the diagnostic applications.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study demonstrating that the Barco Nio Color 2MP (MDNC-2521) and Nio Color 3MP (MDNC-3521) displays meet these criteria for substantial equivalence to their predicate devices. The study focuses on physical performance bench tests rather than clinical efficacy with AI or human readers.

    Here's the breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the comparison to predicate devices and the performance bench tests. The Predicate Device columns represent the established performance, which the new devices aim to be substantially equivalent to. The Device for which listing is sought columns show the reported performance of the new devices.

    Acceptance Criteria (Implied by Predicate Performance) and Reported Device Performance

    Feature/TestMDNC-2521 (New Device) PerformanceMDNC-2221 (Predicate) PerformanceMDNC-3521 (New Device) PerformanceMDNC-3421 (Predicate) Performance
    Common Device Parameters
    Screen technologyIPS-SFT Color LCDIPS-ProIPS-SFT Color LCDIPS-TFT Color LCD
    Active screen size (diagonal)541 mm (21.3")540 mm (21.3")541 mm (21.3")540 mm (21.3")
    Active screen size (HxV)433 x 325 mm (17.1 x 12.8")432 x 324 mm (17.0 x 12.8")433 x 325 mm (17.1 x 12.8")432 x 324 mm (17.0 x 12.8")
    Aspect ratio (H:V)4:34:34:34:3
    Resolution2MP (1600 x 1200 pixels)2MP (1600 x 1200 pixels)3MP (2048 x 1536 pixels)3MP (2048 x 1536 pixels)
    Color imagingYesYesYesYes
    Gray imagingYesYesYesYes
    Bit depth30 bit30 bit30 bit30 bit
    Viewing angle (H, V)178°178°178°178°
    Uniformity CorrectionULTULTULTULT
    SteadyColor CalibrationYes (in MXRT Display Controller)NoYes (in MXRT Display Controller)No
    Ambient Light Compensation (ALC)Yes, reading room selectionYes, reading room selectionYes, reading room selectionYes, reading room selection
    Ambient Light SensorYesYesYesYes
    Front sensorYesYesYesYes
    Maximum luminance (panel typical)1000 cd/m²800 cd/m²1050 cd/m²900 cd/m²
    DICOM calibrated luminance600 Cd/m²500 cd/m²600 Cd/m²500 cd/m²
    Contrast ratio (panel typical)2000:11400:12000:11400:1
    Response time ((Tr+Tf)/2) (typical)12 ms[*]10 ms12 ms[*]20 ms
    Housing colorBlack (RAL 9004) / White (RAL 9003)RAL 9003 / RAL 9004Black (RAL 9004) / White (RAL 9003)RAL 9003 / RAL 9004
    Video input signals2x DisplayPort 1.41x DVI, 1x DisplayPort2x DisplayPort 1.4DVI-D Dual Link, DisplayPort
    USB ports2x USB-B 2.0 upstream, 5x USB-A 2.0 downstream1x USB 2.0 upstream, 2x USB 2.0 downstream2x USB-B 2.0 upstream, 5x USB-A 2.0 downstream1x USB 2.0 upstream, 3x USB 2.0 downstream
    Power rating24 VDC, 4 A24 VDC, 3.75 A24 VDC, 4 A24 VDC, 4 A
    Power consumption37 W (nominal)50 W (nominal)45 W (nominal)50 W (nominal)
    Operating temperature0 °C to 35 °C (20 °C to 30 °C within specs)0 °C to 35 °C (15 °C to 30 °C within specs)0 °C to 35 °C (20 °C to 30 °C within specs)0 °C to 40 °C (15 °C to 35 °C within specs)
    Storage temperature-20 °C to 60 °C-20 °C to 60 °C-20 °C to 60 °C-20 °C to 60 °C
    Operating humidity8% to 80% (non-condensing)8 % to 80 % (non-condensing)8% to 80% (non-condensing)8 % to 80 % (non-condensing)
    Storage humidity5% to 85% (non-condensing)5% to 85% (non-condensing)5% to 85% (non-condensing)5% to 85% (non-condensing)
    Minimum operating pressure70 kPa minimum70 kPa minimum70 kPa minimum70 kPa minimum
    Storage pressure50 to 106 kPa50 to 106 kPa50 to 106 kPa50 to 106 kPa
    Bench Tests
    Spatial resolution – MTFSimilar characteristicsN/A (implied by predicate)Similar characteristicsN/A (implied by predicate)
    Pixel defects, ArtifactsSimilar characteristicsN/A (implied by predicate)Similar characteristicsN/A (implied by predicate)
    Temporal ResponseSimilar characteristicsN/A (implied by predicate)Similar characteristicsN/A (implied by predicate)
    Maximum and Minimum LuminanceSimilar characteristicsN/A (implied by predicate)Similar characteristicsN/A (implied by predicate)
    Luminance response, Conformance to DICOM GSDFSimilar characteristicsN/A (implied by predicate)Similar characteristicsN/A (implied by predicate)
    Angular Dependency of LuminanceSimilar characteristicsN/A (implied by predicate)Similar characteristicsN/A (implied by predicate)
    Luminance uniformitySimilar characteristicsN/A (implied by predicate)Similar characteristicsN/A (implied by predicate)
    Reflection coefficientSimilar characteristicsN/A (implied by predicate)Similar characteristicsN/A (implied by predicate)
    Veiling glare or small-spot contrastSimilar characteristicsN/A (implied by predicate)Similar characteristicsN/A (implied by predicate)
    Color trackingSimilar characteristicsN/A (implied by predicate)Similar characteristicsN/A (implied by predicate)
    EMC and Safety standardsCompliantN/A (implied by predicate)CompliantN/A (implied by predicate)

    Note: "Similar characteristics" is the conclusion of the bench tests as stated in the document.
    [] The intrinsic response time of the new LCD panel on the modified device is further improved with Barco's RapidFrame technology, a proprietary medical overdrive algorithm which improves the temporal response of the system.*

    2. Sample Size Used for the Test Set and Data Provenance

    The study described is a series of physical performance bench tests conducted on the device itself and in comparison to predicate devices, rather than a study involving medical images or patient data.

    • Sample Size for Test Set: Not applicable in the context of clinical images/data. The tests were performed on the modified devices (MDNC-2521 and MDNC-3521).
    • Data Provenance: Not applicable. The "data" here refers to the physical performance measurements of the display hardware.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not applicable. The ground truth for display performance parameters (e.g., luminance, resolution) is established by metrology standards and measurement equipment, not by human expert opinion in this context.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Physical performance measurements do not inherently require an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. The submission focuses on device characteristics and substantial equivalence to existing predicate devices based on physical performance.
    • Effect Size: Not applicable as no MRMC study was conducted.

    6. If a Standalone (Algorithm Only Without Human-in-the Loop Performance) Was Done

    • Standalone Performance: Not applicable. The device is a medical computer display, and its performance is assessed based on its physical properties and adherence to display standards, not as an AI algorithm. While it can be optionally used with software like QAWeb Enterprise for calibration and Intuitive Workflow Tools for image enhancement, the submission focuses on the display hardware itself.

    7. The Type of Ground Truth Used

    The ground truth for this device's evaluation is based on physical measurement standards and the performance specifications of the predicate devices. The "Guidance for Industry and FDA Staff: Display Devices for Diagnostic Radiology" (issued in 2022) serves as the reference for the physical laboratory testing instructions, which define the expected performance characteristics.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This is a hardware device (display) and not an AI algorithm that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable. As this is not an AI algorithm, there is no training set or ground truth established for training purposes.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1