510(k) Data Aggregation

    K Number
    K230358

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-16

    (95 days)

    Product Code
    Regulation Number
    882.5808
    Age Range
    12 - 21
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SAVI Dual™ Migraine Therapy device is indicated for the acute and prophylactic treatment of migraine headache in adolescents (age 12 and older) and adults.

    Device Description

    The SAVI Dual Migraine Therapy device is a portable, hand-held device that delivers a brief single pulse of magnetic energy at 0.9 Tesla to the back of the head to induce an electrical current in a portion of the brain, called the occipital cortex, to stop or lessen the effects of migraine headaches. Since a single pulse of magnetic stimulation is emitted, this method of stimulation is called single pulse transcranial magnetic stimulation or sTMS. The SAVI Dual is indicated for the acute and prophylactic treatment of migraine headache in adolescents (age 12 and older) and adults. The device is designed for patient use where treatments are self-administered and can be delivered in a variety of settings including the home or office. The device is intended for prescription use only. To use the device, the user must either insert a SIM chip to use the device for a programmed duration or the user may connect wirelessly via a cellular connection for the programmed duration. The programmed duration corresponds to the prescribed months of use.

    AI/ML Overview

    The provided text describes the regulatory clearance for the SAVI Dual™ Migraine Therapy device and establishes its substantial equivalence to a predicate device (SpringTMS) and a reference device (sTMS mini). The focus is on non-clinical performance testing rather than clinical study results comparing device performance against specific disease metrics.

    Therefore, the acceptance criteria and study information will be presented based on the available non-clinical testing data provided in the document.


    Description of Acceptance Criteria and Study Proving Device Meets Acceptance Criteria

    The SAVI Dual™ Migraine Therapy device underwent non-clinical performance testing to demonstrate its substantial equivalence to the predicate device, SpringTMS (K182976), and a reference device, sTMS mini (K161663). The acceptance criteria were primarily focused on ensuring that the SAVI Dual maintained similar performance characteristics to the predicate device regarding its magnetic pulse output and safety standards, despite hardware and software modifications.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    Magnetic Pulse Characteristics vs. TimeThe SAVI Dual device was expected to have the same magnetic pulse specification as the predicate device (SpringTMS) and perform within that specification. The measured rate of change of the magnetic field should be substantially equivalent to the predicate.Both SAVI Dual and the predicate device have the same specification for magnetic pulse shape, and both devices tested within specification. The measured rate of change of the magnetic field is substantially equivalent. No new issues of safety or efficacy have been raised.
    Magnetic Pulse Field MapThe Magnetic Pulse Field Map for the SAVI Dual was expected to be substantially equivalent to that of the predicate device.The Magnetic Pulse Field Maps for the subject device and the predicate device are substantially equivalent. No new issues of safety or efficacy have been raised.
    Location of 5 Gauss LineThe location of the 5 Gauss line for the SAVI Dual was expected to be substantially equivalent to that of the predicate device.The location of the 5 Gauss line for the subject device and the predicate device are substantially equivalent. No new issues of safety or efficacy have been raised.
    Software Verification & ValidationThe SAVI Dual software was expected to be tested against requirements of the Software Requirements Specification (SRS) and ensure device operations result in a magnetic pulse substantially equivalent to the predicate device. The testing should raise no new safety or efficacy issues.The SAVI Dual software was tested against requirements of the Software Requirements Specification (SRS), and no new issues of safety or efficacy have been raised. The software requirements specify device operations that result in the delivery of a magnetic pulse that is substantially equivalent to the magnetic pulse of the predicate device.
    Electromagnetic Compatibility & Electrical SafetyThe SAVI Dual device was expected to meet all acceptance criteria in accordance with the following standards: IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-1-11. The testing should raise no new safety or efficacy issues.The subject device met all acceptance criteria for Electromagnetic Compatibility and Electrical Safety testing in accordance with IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-1-11. No new issues of safety or efficacy have been raised. Therefore, the SAVI Dual is substantially equivalent to the predicate.

    2. Sample size used for the test set and the data provenance

    The document details non-clinical performance testing rather than testing on human subjects or clinical data sets for efficacy. Therefore, there is no stated sample size for a "test set" in the context of patient data. The "test set" in this context refers to the individual SAVI Dual™ devices or components that underwent the various non-clinical engineering and safety tests (e.g., magnetic pulse characteristics, software performance, EMC/electrical safety). The provenance of this data is from eNeura, Inc.'s internal testing facilities. The testing described is prospective, as it was performed on the proposed SAVI Dual device to support its regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable in the context of the provided document. The "ground truth" for non-clinical performance testing is established by engineering specifications, international consensus standards (e.g., IEC 60601 series), and direct comparison to the predicate device's measured performance. It does not involve human expert adjudication in the way clinical diagnostic studies do. The experts involved would be the engineers and technicians performing the tests and comparing the results to established physical and regulatory requirements.

    4. Adjudication method for the test set

    Not applicable. As described above, the "test set" refers to physical device characteristics and engineering performance, not interpretative medical decisions requiring adjudication by experts. Results are compared against specific quantitative thresholds and established standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The SAVI Dual™ Migraine Therapy device is a hardware device (transcranial magnetic stimulator), not an AI-powered diagnostic or interpretive tool. The provided document does not describe any MRMC studies or AI assistance in the context of its use.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The SAVI Dual is a therapeutic device that stimulates the brain. Its performance is evaluated based on its physical output (magnetic field characteristics) and safety, not as a standalone algorithm delivering a diagnosis or prediction without human interaction. While it has software/firmware, its "performance" in this context is its ability to produce the intended physical magnetic pulse, which was tested.

    7. The type of ground truth used

    The "ground truth" for the non-clinical performance testing (magnetic pulse characteristics, software function, electrical safety) was based on:

    • Established engineering specifications and design requirements for the device.
    • Measured performance of the legally marketed predicate device (SpringTMS) for comparative equivalence in magnetic field characteristics.
    • International consensus standards for medical device safety and electromagnetic compatibility (IEC 60601 series).

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm that requires a "training set" of data. Its development involved engineering design and testing, not machine learning.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" for this type of device.