510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Profisil® Fluoride Varnish is a 5 % sodium fluoride varnish which produces mechanical occlusion of the dentinal tubules in the treatment of tooth hypersensitivity.

    Device Description

    Profisil® Fluoride Varnish is designed to securely adhere to the tooth surface for several hours while releasing fluoride ions. The varnish contains 5 % sodium fluoride suspended in a mucosa-friendly, pleasantly flavored dimethicone gel. The varnish is available in following versions: mint, berry and unflavored.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a dental device, "Profisil Fluoride Varnish." This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than conducting new clinical trials for effectiveness. Therefore, the information typically requested about acceptance criteria and studies proving a device meets them for new effectiveness claims, especially those involving AI or comparative effectiveness with human readers, is not directly applicable to this specific submission.

    However, the document does detail how the new device demonstrates equivalence to its predicate, focusing on non-clinical performance and biocompatibility. Here's a breakdown of the relevant information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    For "Profisil® Fluoride Varnish," the acceptance criteria for non-clinical performance are based on the international standard EN ISO 17730:2020 "Dentistry - fluoride varnishes."

    Performance CharacteristicAcceptance Criteria (EN ISO 17730:2020)Reported Device Performance (Profisil® Fluoride Varnish)Conclusion
    pH valueNot explicitly stated, but within standard requirementsAssessed and verified against EN ISO 17730:2020Fulfilled
    Dynamic viscosityNot explicitly stated, but within standard requirementsAssessed and verified against EN ISO 17730:2020Fulfilled
    Fluoride releaseNot explicitly stated, but within standard requirementsAssessed and verified against EN ISO 17730:2020Fulfilled; equivalent rate to predicate
    Electron microscopy (SEM) studyNot explicitly stated, but within standard requirementsAssessed and verified against EN ISO 17730:2020Fulfilled
    Total FluorideRelevant for functioning as intended (EN ISO 17730:2020)Contains 5% sodium fluoride (identical to predicate)Equivalent

    Biocompatibility Acceptance Criteria:

    TestStandardAcceptance CriteriaReported Device Performance (Profisil® Fluoride Varnish)
    CytotoxicityISO 10993-5Meets ISO 10993-5 requirementsMet requirements
    SensitizationISO 10993-10Meets ISO 10993-10 requirementsMet requirements
    IrritationISO 10993-23Meets ISO 10993-23 requirementsMet requirements
    PyrogenicityISO 10993-11Meets ISO 10993-11 requirementsMet requirements
    Systemic toxicityISO 10993-11Meets ISO 10993-11 requirementsMet requirements

    2. Sample Size Used for the Test Set and Data Provenance

    • Test set sample size: Not explicitly stated as this is a non-clinical device evaluation. The "test set" would refer to the samples of the Profisil® Fluoride Varnish subjected to the various physical and chemical tests (e.g., specific batches for pH, viscosity, fluoride release tests) and biocompatibility studies (e.g., cell cultures for cytotoxicity, animal models for sensitization/irritation). The exact number of samples tested for each property is not detailed in the summary.
    • Data provenance: The standard EN ISO 17730:2020 is an international standard. Biocompatibility standards like ISO 10993 series are also international. The summary indicates these tests were performed to demonstrate compliance with these standards, suggesting laboratory testing specifically for the Profisil® product. There is no information on country of origin of data beyond the manufacturer being based in Germany. The data is prospective as it was generated to support the current submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The "ground truth" for this type of device comparison is based on established scientific and engineering principles codified in international standards (EN ISO 17730:2020 for dental fluoride varnishes) and biocompatibility standards (ISO 10993 series). The results are objectively measured physical and chemical properties and biological responses, not subjective interpretations by human experts in the sense of medical image analysis.

    4. Adjudication Method for the Test Set

    Not applicable, as the "test set" involves objective laboratory measurements against predefined standard requirements rather than subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable. This device is a dental fluoride varnish, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies are irrelevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    Not applicable for the same reason as above; this is a physical dental material, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the current device's performance evaluation is based on International Standards:

    • EN ISO 17730:2020 "Dentistry - fluoride varnishes" for physical and chemical properties (pH, dynamic viscosity, fluoride release, SEM).
    • ISO 10993 series for biocompatibility studies (cytotoxicity, sensitization, irritation, pyrogenicity, systemic toxicity).

    These standards define the acceptable range or nature of results for these properties, which serve as the objective "ground truth" for evaluation.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm and therefore does not have a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.