510(k) Data Aggregation

    K Number
    K230046

    Validate with FDA (Live)

    Date Cleared
    2023-03-25

    (78 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The vinyl examination glove is intended to be worn on the hands of examination between patient and examiner. This is a single-use, powder-free, non-sterile device.

    Device Description

    The proposed device is a powder free vinyl patient examination glove, provided as nonsterile and disposable device. The proposed device is provided with clear color. There are four sizes: small, medium, large and extra-large as options.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance data for Vinyl Examination Gloves, a Class I medical device. It does not pertain to an AI/ML-based medical device. Therefore, many of the requested fields regarding AI/ML device testing (e.g., sample size for training set, number of experts for ground truth, MRMC study, standalone algorithm performance) are not applicable.

    Here's the information extracted from the document, focusing on the device performance and acceptance criteria for these physical examination gloves:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Performance / Results
    ASTM D5250 (Physical Dimensions Test)To verify physical dimensions- Length: ≥ 230mm - Width (Small): 85±5mm - Width (Medium): 95±5mm - Width (Large): 105±5mm - Width (Extra-Large): 115±5mm - Thickness (Finger): ≥ 0.08mm - Thickness (Palm): ≥ 0.08mmPass
    ASTM D5250 (Physical Properties - Before Aging)To determine tensile strength and ultimate elongation before aging- Tensile Strength: ≥ 11MPa - Ultimate Elongation: ≥ 300%Pass
    ASTM D5250 (Physical Properties - After Aging)To determine tensile strength and ultimate elongation after aging- Tensile Strength: ≥ 11MPa - Ultimate Elongation: ≥ 300%Pass
    ASTM D5151Freedom from pinholesMeets the ASTM D5151 Test for AQL 2.5 requirementsPass
    ASTM D6124Powder ResidueMeets ASTM D6124 requirements: < 2.0mgPass
    ISO 10993-10To determine if the device material is an irritantNon-irritatingUnder the conditions of the study, not an irritant. / Pass
    ISO 10993-10To determine if the device material is a sensitizerNon-sensitizingUnder conditions of the study, not a sensitizer. / Pass
    ISO 10993-11To determine if the device material extracts pose a systemic toxic responseNon-acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivo. / Pass

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each test mentioned (e.g., number of gloves tested for physical properties, pinholes, or biocompatibility). The tests are based on established ASTM and ISO standards, which typically specify sampling plans.

    • Data Provenance: The tests were conducted internally by the manufacturer, Guangdong Kingfa Sci.& Tech.Co., Ltd., in China. The documentation implies these were prospective tests conducted on the manufactured product to demonstrate compliance with standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not an AI/ML device requiring expert interpretation of results. The "ground truth" for the performance of examination gloves is based on objective, quantifiable physical and chemical properties measured according to international standards (ASTM, ISO).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI/ML device requiring human adjudication of qualitative data.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on objective measurements against established engineering and biocompatibility standards (ASTM D5250, ASTM D5151, ASTM D6124, ISO 10993-10, ISO 10993-11). For example, tensile strength and elongation are measured directly, and biocompatibility is assessed through standardized in-vivo and in-vitro tests with specific endpoints.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device; there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.