510(k) Data Aggregation

    K Number
    K223684

    Validate with FDA (Live)

    Date Cleared
    2023-09-05

    (271 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    F&P 820 System Humidifier Base:

    The F&P 820 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases when using a continuous or intermittent ventilator system or a continuous gas flow.

    This system is intended for both noninvasive and invasive therapies. The addition of heat and humidity to the supply of cold and dry respiratory gases provided through noninvasive ventilation is beneficial to prevent drying of the patient's airways.

    The F&P 820 System is designed for adult and pediatric patients (excluding neonate), requiring a flow range ≥ 5 L/min. It is designed for use in hospitals, long term care facilities and homes, under the prescription of qualified medical professionals.

    F&P 820A10 Breathing Tubes:

    The F&P 820 series breathing tubes are an accessory to the F&P 820 System and are compatible with F&P 820 series humidifiers.

    The F&P 820 System is intended to provide therapeutic levels of heat and humidity to a patient's inspiratory gases when using a continuous or intermittent ventilator system or a continuous gas flow.

    This system is intended for both noninvasive and invasive therapies. The addition of heat and humidity to the supply of cold and dry respiratory gases provided through noninvasive ventilation is beneficial to prevent drying of the patient's airways.

    This system is designed for adult and pediatric (excluding neonate) use in hospitals, long term care facilities and homes, under the prescription of qualified medical professionals.

    F&P MR325 Humidification Chamber:

    The MR325 humidification chamber is an accessory to the F&P 820 System and is compatible with F&P 820 series humidifiers.

    The F&P 820 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases when using a continuous or intermittent ventilator system or a continuous gas flow.

    This system is intended for both noninvasive and invasive therapies. The addition of heat and humidity to the supply of cold and dry respiratory gases provided through noninvasive ventilation is beneficial to prevent drying of the patient's airways.

    This system is designed for adult and pediatric (excluding neonate) use in hospitals, long term care facilities and homes, under the prescription of qualified medical professionals.

    Device Description

    The Fisher & Paykel 820 Humidification (also referred to as F&P 820 System) is designed to provide respiratory humidification by providing heat, humidity, and delivering respiratory gases to patients.

    The F&P 820 System is a revision of the predicate device platform, the Fisher & Paykel MR810 Respiratory Humidifier (K143646).

    AI/ML Overview

    The provided text is a 510(k) Summary for the F&P 820 Humidification System. It describes the device, its intended use, comparison with predicate devices, and the non-clinical testing performed to demonstrate substantial equivalence.

    However, the information requested in your prompt regarding acceptance criteria and a study that proves the device meets the acceptance criteria with specific details like sample size, expert ground truth establishment, adjudication methods, MRMC studies, and separate training/test sets with their ground truth establishment for an AI/ML powered device is not present in this document.

    This document pertains to a traditional medical device (a respiratory humidifier system), not an AI/ML-powered diagnostic or therapeutic device. Therefore, the detailed acceptance criteria and study design elements typically associated with AI/ML device validation (like sensitivity, specificity, AUC, human reader improvement with AI assistance, sample sizes for training/test sets, and expert consensus for ground truth) are not applicable to this submission and are consequently not described.

    The "Performance Data" section (Section VII) focuses on demonstrating compliance with relevant international standards for medical devices (e.g., electrical safety, EMC, biocompatibility, software verification/validation, human factors, and bench/performance testing for humidification output, thermal overshoot, resistance to flow, etc.). These are the "acceptance criteria" for this type of device, and the summary states that the device "meets the applicable requirements" and "demonstrates an appropriate profile."

    To directly answer your prompt, based only on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't present "acceptance criteria" in the format of a table as you might expect for an AI/ML study (e.g., minimum sensitivity). Instead, it states that the device was tested to "applicable requirements" of various standards, and the performance described is primarily via comparison to the predicate device in Table 1, where the subject device's performance characteristics are presented as meeting or being equivalent to those of the predicate and relevant standards.

    Here's an extraction of relevant performance characteristics from Table 1, framed as criteria and performance. Note: These are not explicit "acceptance criteria" numbers as you would see for predictive models, but rather technical specifications and compliance claims.

    Acceptance Criteria (Derived from Standards/Predicate)Reported Device Performance (F&P 820 Humidification System)
    Humidity Performance
    Setting 1, 2, or 3 (Flow Range 5-70 L/min)≥ 12 mg/L
    Setting 4 (Flow Range 5-40 L/min)≥ 33 mg/L
    Gas temperature at patient end of breathing tube for flow range
    Setting 126-30 °C
    Setting 229-37 °C
    Setting 332-40 °C
    Setting 4 (Flow Range 5-40 L/min)35-43 °C
    Time to reach set temperature<60 mins
    Ambient Temperature Range18 – 26 °C
    AlarmsAlarms exceed 45 dbA @1m; Auditory alarm pause: 120 seconds (Different from predicate, but conforms to IEC 60601-1-8:2006 + A1:2012)
    Transport Conditions-25 to 70 °C (Broader range than predicate, but equivalent as it encompasses predicate's range)
    Service Life7 Years
    IEC 60601-1 ClassificationCLASS II (Different from predicate CLASS I, but conforms to AAMI/IEC 60601-1:2005 + AMD1:2012)
    BiocompatibilityMeets ISO 10993-1:2018 and ISO 18562-1:2017: "demonstrates an appropriate biocompatibility profile for the device."
    Electrical Safety, EMC, AlarmsComplies with ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, IEC 60601-1-2:2014 + AMD1:2020 Ed 4.1, AIM Standard 7351731 Rev. 2.00 2017-02-23. Alarms testing in accordance with IEC 60601-1-8:2006 and A1:2012. "demonstrated the appropriate electrical safety and electromagnetic compatibility profile."
    Software Verification & ValidationConducted and documented as recommended by FDA guidance.
    Human FactorsStudy conducted and documented as recommended by FDA guidance. "demonstrated that the human factors assessment had been sufficiently evaluated and was safe and effective for the intended users, uses, and use environments."
    Resistance to flow, compliance, and gas leak testingIn line with ISO 5367:2014.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified. For a non-AI/ML device, performance testing usually involves a defined number of units tested to ensure design specifications are met, but this is not typically expressed as a "test set sample size" in the context of clinical data used for AI/ML performance evaluation.
    • Data Provenance: Not applicable/specified. The testing performed is bench/laboratory testing of the physical device, not analysis of clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. Ground truth as established by human experts is not relevant for the type of bench and performance testing conducted for this device. Ground truth, in this context, refers to measurements from calibrated equipment and adherence to standard specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are used to resolve discrepancies in human expert labeling of data, which is not part of this device's testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. An MRMC study is relevant for AI/ML diagnostic aids. This is a respiratory humidifier system, not a diagnostic imaging AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm-only device. Its performance is inherent to its physical design and functionality.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" for this device's performance is based on measurements from calibrated test equipment and adherence to established engineering and medical device standards (e.g., ISO, IEC). For instance, humidity output is measured by the device itself and validated against a standard, not derived from expert consensus on patient outcomes.

    8. The sample size for the training set:

    • Not applicable. This device does not employ machine learning that would require a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. No training set is used for this device.