510(k) Data Aggregation

    K Number
    K223616

    Validate with FDA (Live)

    Date Cleared
    2022-12-30

    (25 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Acquire S Fine Needle Biopsy Device is designed to sample targeted submucosal and extramural gastrointestinal lesions through the accessory channel of a curvilinear echoendoscope.

    Device Description

    The Acquire™ S Endoscopic Ultrasound Fine Needle Biopsy (FNB) Device is an endoscopic ultrasound biopsy needle that can be coupled to the biopsy channel of a Curvilinear Array (CLA) Echoendoscope with a standard luer connection and delivered into the digestive tract. The device sheath length can be adjusted to accommodate different model echoendoscopes. The need to acquire samples from lesions within and adjacent to the digestive system's major lumens that can be identified and targeted using the echoendoscope. Both the sheath and needle length can be adjusted based on distance to the target lesion. The sheath and needle length adjustments are set and locked by the physician by using the locking knob mechanisms on the handle of the device. A sample is obtained by penetrating the lesion with the needle while applying suction and manipulating the needle in a back and forth motion to acquire a sample. The sample can be prepared per normal institutional protocol. The Acquire™ Needle has echogenic (visible under ultrasound) features at the distal end to facilitate real time visualization of the device under ultrasound.

    Syringe and stopcock are accessories to provide and control the vacuum suction to aspirate the sample.

    AI/ML Overview

    This document describes the Acquire™ S Endoscopic Ultrasound Fine Needle Biopsy (FNB) Device (K223616) and its substantial equivalence to a predicate device. This is a medical device submission, not an AI/ML device submission, and therefore, the acceptance criteria and study detailed below relate to the physical device's performance, not an algorithm's.

    Here's the information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Since this is a physical medical device and the submission focuses on demonstrating substantial equivalence to a predicate, the "acceptance criteria" are generally that the device performs as intended and is safe and effective for its indicated use, similar to the predicate. The performance reported here refers to the results of non-clinical testing.

    Acceptance Criteria CategoryReported Device Performance
    Non-clinical performance bench testingBench Testing included:- Device Durability: All testing was passing.- Stylet Removal Force: All testing was passing.- Needle Extension Length: All testing was passing.- Adjustable Catheter Length: All testing was passing.- Stylet Cap Tensile: All testing was passing.Overall: ""All testing was passing and demonstrates the device's ability to fulfill non-clinical performance bench testing...requirements."
    Simulated use testingPerformed. All testing was passing. (Implied by the summary statement: "All testing was passing...")
    BiocompatibilityPerformed per the requirements of ISO 10993. All testing was passing. (Implied by the summary statement: "All testing was passing...")
    Sterilization validationPerformed per the requirements of ISO 11135. All testing was passing. (Implied by the summary statement: "All testing was passing...")
    Substantial EquivalenceDemonstrated to the currently cleared Acquire™ Endoscopic Ultrasound Fine Needle Biopsy (FNB) Device (K160845).

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of clinical data or patient samples. The testing described is non-clinical performance bench testing, simulated use testing, biocompatibility, and sterilization validation. Therefore, there are no details about sample size for a patient test set or data provenance (country of origin, retrospective/prospective). The closest equivalent of "samples" would be the number of devices or components tested in the bench/simulated use tests, which is not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable and not provided. The "ground truth" for a physical device like this is its mechanical and material performance against established engineering and medical standards, not diagnostic accuracy requiring expert interpretation.

    4. Adjudication method for the test set

    This information is not applicable and not provided. The testing involves objective mechanical and material properties, not subjective assessments requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study is not applicable as this is a physical medical device, not an AI/ML diagnostic tool used for image interpretation. No such study was mentioned or performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to AI/ML device performance. This device is a physical biopsy needle, not an algorithm. Therefore, "standalone algorithm performance" is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established by engineering and quality standards, and predefined performance specifications for mechanical properties (durability, force, length adjustments) and biological safety (biocompatibility, sterility). The goal is to ensure the device performs reliably and safely, analogous to the predicate.

    8. The sample size for the training set

    This is not applicable as this is a physical medical device, not an AI/ML system requiring a training set.

    9. How the ground truth for the training set was established

    This is not applicable as this is a physical medical device, not an AI/ML system requiring a training set.