Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K223209

    Validate with FDA (Live)

    Device Name
    Elekta Unity
    Date Cleared
    2023-02-23

    (129 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.

    Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems. The Elekta Unity 1.5T MR scanner is a magnetic resonance imaging system that produces cross-sectional images in any orientation of the internal structure of the whole body before, during, and after the radiotherapy treatment.

    When interpreted by a trained physician magnetic resonance images acquired before, during, and after the radiotherapy treatment yield information that can be useful in diagnosis and may assist therapy planning, patient positioning and treatment delivery related to radiation oncology.

    Device Description

    Elekta Unity is a multifunctional digital linear accelerator designed to assist licensed medical practitioners in the delivery of ionizing radiation to defined volumes (e.g. malignant and benign tumors). Elekta Unity is capable of both intensity modulated radiation therapy (IMRT) and image guided radiation therapy (IGRT).

    The Elekta Unity 1.5T MRI scanner is a magnetic resonance imaging sub-system that produces cross-sectional images in any orientation of the internal structure of the whole body before, during, and after the radiotherapy treatment.

    When interpreted by a trained physician the images acquired before, during, and after the radiotherapy treatment vield information that can be useful in diagnosis and may assist therapy planning, patient positioning, and treatment delivery related to radiation oncology.

    In addition to the motion monitoring with manual interrupt previously cleared with predicate Unity devices, the subject device Elekta Unity has added the following motion management strategies:

    • · Anatomic Position Monitoring (APM) with Manual interrupt (also referred to as True Tracking as it 'tracks' in 3D and enables the system to gate using the 3D position of the anatomy's motion)
    • · APM with Anatomic Tolerance Check (ATC)
    • Adaptive Therapy with optional Baseline Shift (BLS) Recovery.

    These modifications are not contained solely within the Elekta Unity Device, as the full clinical benefit is achieved with interoperability of Unity, Monaco RTP, and MOSAIQ OIS.

    AI/ML Overview

    The provided text is a 510(k) summary for the Elekta Unity MR Linac and describes updates to an existing device rather than a new standalone AI/ML device with a clinical performance study. Therefore, a direct answer to all points regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/ML algorithm's clinical performance cannot be fully provided from this document.

    Specifically, the document states: "No animal or clinical tests were performed to establish substantial equivalence with the predicate device." This indicates that a human-in-the-loop (MRMC) or standalone (algorithm only) clinical performance study as typically understood for AI/ML diagnostic or treatment planning algorithms was not conducted for this 510(k) submission. The clearance is based on substantial equivalence by demonstrating that the new motion management strategies do not change the fundamental scientific technology or raise different questions of safety or effectiveness.

    However, I can extract the information relevant to the device's technical specifications and the non-clinical testing performed to establish substantial equivalence.

    Here's a breakdown of the available information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present "acceptance criteria" in the traditional sense of a clinical performance study against a ground truth. Instead, it provides a comparison of "Technological Characteristics" between the subject device (Elekta Unity with new features) and its predicate device (Elekta Unity K212114). The "acceptance" here relates to demonstrating that the new features do not negatively impact the established performance and safety characteristics, or that they meet expected performance defined by engineering specifications.

    Technological CharacteristicAcceptance Criteria (Predicate Performance - Implied)Reported Device Performance (Subject Device)
    Radiation Source / Beam7MV Bremsstrahlung X-Rays✓ (Same)
    Method of IMRTMLC based cone-beam delivery✓ (Same)
    CollimationField shaping, Multi Leaf Collimator (MLC)✓ (Same)
    MLC materialTungsten Alloy✓ (Same)
    Number of leaves80 leaf pairs✓ (Same)
    Range of MLC collimated beam size0.5cm x 0.5cm to 57.4cm x 22cm✓ (Same)
    GantryRing Gantry, collision with patient not possible✓ (Same)
    Radiation Head ShieldingLead, Tungsten Alloy, and Steel shielding✓ (Same)
    Source control mechanismDual channel dose monitoring system✓ (Same)
    Radiation Transmission through head0.2% of the primary beam✓ (Same)
    Isocenter distance143.5 cm✓ (Same)
    Isocenter accuracy (Radius)0.5 mm✓ (Same)
    Max Dose RateClinical use: 450 cG/min at isocentre✓ (Same)
    Static Dose Accuracy>95% points passing 3%/3mm (high dose); >95% passing 5mm/5% (low dose); 1% agreement for output factors✓ (Same)
    Motion Management StrategiesMotion monitoring with manual interrupt✓ (Same + APM with manual interrupt, APM with ATC, Adaptive Therapy with optional BLS Recovery)
    End-to-End Gating LatencyNot applicable to previous predicate; new feature[-200 to +200 ms] (radiation off) and [-200 to +280 ms] (radiation on) for ATC
    Patient table degrees of freedom2 (vertical & longitudinal)✓ (Same)
    Integrated imagingMagnetic resonance imaging system✓ (Same)
    MR Physical Characteristics (Bore Diameter)700 mm✓ (Same)
    MRI Frequency64 MHz✓ (Same)
    Field Strength1.5T✓ (Same)
    Field of ViewUp to 500 mm Sequence dependent✓ (Same)
    Field Homogeneity< 2 ppm over 50 cm x 50 cm x 45 cm volume✓ (Same)
    Field Stability≤ 0.1 ppm/hour✓ (Same)
    3D Imaging Volumes (cm)0.5 x 0.5 x 0.8 (min) / 56 x 56 x 40 (max)✓ (Same)
    3D Imaging Resolution (cm)0.01 x 0.01 x 0.1 (min) / 0.875 x 0.875 x 1 (max)✓ (Same)
    2D Imaging Planes (cm)AP x HF: 0.5 x 0.5 (min) / 56 x 56 (max)✓ (Same)
    2D Imaging Resolution (mm)Selectable: 0.011 x 0.011 mm (min) / 8.75 x 8.75 mm (max); Slice thickness: 0.01 mm (min) / 705 mm (max); ≤1 mm over 20 cm FOV✓ (Same)
    Geometric Accuracy≤ 2 mm over 34 cm FOV (Guaranteed); ≤ 2 mm over 42 cm FOV (Typical)✓ (Same)
    Signal to Noise120✓ (Same)
    Diffusion Weighted Imaging (DWI)DWI can be imported for Off-line review✓ (Same)
    Treatment Planning and Delivery SystemGPU-based Monte Carlo dose calculation algorithm (GPUMCD) using Elekta Monaco✓ (Same)
    Dose Output ModellingDose output modelled with monitor units✓ (Same)
    Dose DisplayDisplay of Linac delivery parameters✓ (Same)

    The specific performance of the new motion management strategies (APM with manual interrupt, APM with ATC, Adaptive Therapy with optional BLS Recovery) is demonstrated through consistency with existing features and the quantitative specification for End-to-End Gating Latency.

    2. Sample size used for the test set and data provenance

    Not applicable, as no clinical performance study was conducted. The assessment was based on non-clinical (engineering and software) verification and validation testing, and comparison of technological characteristics to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts

    Not applicable, as no clinical performance study requiring expert adjudication for ground truth was conducted. Ground truth for the MR images themselves is implicitly the physical reality captured by the scanner, with interpretation by "trained physician" for diagnosis, therapy planning, patient positioning, and treatment delivery.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable, as no clinical performance study was conducted.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done, stated as "No animal or clinical tests were performed to establish substantial equivalence with the predicate device."

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No standalone clinical performance study was done for the purpose of this 510(k) clearance, as explicitly stated that no animal or clinical tests were performed. The "algorithm" here refers to the system's control and motion tracking, which is part of the integrated device functionality, not a separate AI diagnostic or treatment planning algorithm being evaluated for standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical testing, the "ground truth" would be engineering specifications, physical measurements, and adherence to recognized standards (e.g., IEC 60601 series for safety and essential performance). For the imaging component, the imaging system produces cross-sectional images for interpretation by a "trained physician," where the intended ground truth for diagnostic or guidance purposes is the patient's anatomy and pathology as determined by clinical practice.

    8. The sample size for the training set

    Not applicable. This document describes a medical device with updated motion management software, not a de novo AI/ML algorithm requiring a training set in the typical sense for machine learning model development. The software capabilities are validated through engineering and functional testing.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.