510(k) Data Aggregation

    K Number
    K222967

    Validate with FDA (Live)

    Date Cleared
    2022-11-23

    (57 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Avanti Orthopaedics 2.2mm Cannulated Compression Screws are intended for fixation of fractures, fusions, osteotomies, non-unions of small bones in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.

    The Avanti Orthopaedics 3.2mm, and 4.0mm Cannulated Compression Screws are intended for fixatures, fusions, osteotomies, non-unions of small bones and small bone fragments in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.

    Device Description

    The Avanti Orthopaedics Cannulated Compression Screw System are metallic bone screws manufactured from Stainless Steel (ASTM F138). The screws are available in multiple lengths and diameters, and are intended to be used as stand-alone bone screws for internal bone fixation of fractures, fusions, osteotomies, non-unions, and malunions in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by the screw fixation. The Avanti Orthopaedics Cannulated screws are cannulated for use with guidewires for precise placement in bone.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for Avanti Orthopaedics 2.2mm, 2.7mm, 3.2mm, and 4.0mm Cannulated Screws. This document focuses on demonstrating substantial equivalence to predicate devices, primarily through non-clinical performance testing of the device hardware itself, rather than the performance of an AI/ML powered medical device.

    Therefore, the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/ML medical device (referring to aspects like sample size, ground truth, expert consensus, MRMC studies, etc.) is not present in the provided document.

    The document entirely pertains to the physical characteristics and performance of orthopedic screws.

    Here's what can be extracted about the non-clinical performance testing of these orthopedic screws:


    Summary of Acceptance Criteria and Device Performance for Orthopedic Screws

    The provided document describes the acceptance criteria and performance for physical orthopedic devices, specifically Avanti Orthopaedics 2.2mm, 2.7mm, 3.2mm, and 4.0mm Cannulated Screws. The testing is focused on the mechanical properties and safety of the screws, not on an AI/ML algorithm's performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Defined by)Reported Device Performance
    Predetermined acceptance criteria as defined in the FDA guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers -Performance Criteria for Safety and Performance Based Pathway".The device met all predetermined acceptance criteria as defined in the referenced FDA guidance.

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document states that "Testing was performed per ASTM F543-17."
    • Sample Size: Not explicitly stated in the document. ASTM F543-17 (Standard Specification and Test Method for Metallic Bone Screws) would specify the required number of samples for each mechanical test (e.g., torsion, bending, dynamic fatigue).
    • Data Provenance: Not applicable in the context of clinical data/images. This testing pertains to the physical device. The testing was non-clinical (laboratory testing).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not applicable. Ground truth in this context refers to the physical properties and performance of the screw under specific test conditions, as defined by engineering standards (ASTM F543-17) and FDA guidance. No human expert interpretation of data is involved in establishing "ground truth" for mechanical testing.

    4. Adjudication Method for the Test Set:

    • Not applicable. This is mechanical testing against defined specifications, not a subjective assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. This is a physical medical device (orthopedic screw), not an AI/ML software device. MRMC studies are used for evaluating diagnostic or treatment planning software performance with human readers.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device.

    7. The Type of Ground Truth Used:

    • Engineering Standards and Specifications: The "ground truth" for the device's performance is established by the specified mechanical properties and performance requirements outlined in ASTM F543-17 and the FDA guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers - Performance Criteria for Safety and Performance Based Pathway". This is objective, measurable data.

    8. The Sample Size for the Training Set:

    • Not applicable. This is a physical device, not an AI/ML algorithm that requires training data.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable.

    In summary, the provided FDA document is a 510(k) submission for conventional orthopedic screws, not an AI/ML device. Therefore, the detailed questions about AI/ML device performance (like training sets, ground truth establishment by experts, MRMC studies, etc.) are outside the scope of this document.