Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222686

    Validate with FDA (Live)

    Date Cleared
    2022-10-06

    (30 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Perfit FS Dental Zirconia Fully Sintered Block is used for dental restorations using different CAD / CAM or manual milling machines. All blocks are processed through dental laboratories or by dental professionals.

    Device Description

    The Perfit FS Dental Zirconia Fully Sintered Block consists of zirconium oxide which compresses ceramic for milling, grinding, and further firing. It is normally pre shaded, fully sintered and milled by dental technician with CAD/CAM machine for shaping artificial teeth such as crowns, bridges, veneers, inlay etc.

    The Perfit FS Dental Zirconia Fully Sintered Block are zirconia oxide block made available in different versions and chemical compositions of various colors, shades and dimensions and will further process the discs by milling them to make final dental restorations such as crowns, bridges, veneers, inlays and on lays based on the anatomical rendering of the patient's teeth using CAD/CAM equipment. At the completion of the machining steps, the dental restoration is fired (i.e., sintered) in the oven to harden the Zr02 so that its final properties can be achieved.

    They are available in different models that differ in various specification and color. There are white zirconia products(White) and colored zirconia products, and colored zirconia is also divided into Monolayer(A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4, White) color products , Shade multi-layer(ML A1/A2/A3, A1M, A2M, A3M, A3.5M, A4M, B1M, B2M, B3M, B4M, C1M, C2M, C3M, C4M, D2M, D3M, D4M) color products.

    The white zirconia is composed of Zr02+Y203+Nb205 and an additional inorganic pigment: Al203 and Other oxide. The color zirconia products are derived from the same zirconia powder as the regular white zirconia powder, with the addition of very small amount of inorganic pigments: Fe203 and other oxide. The inorganic pigments generate the color on the prosthetic dental device, after sintering at dental labs that matches natural color of patient's teeth.

    The composition of the Dental Zirconia Block including the white zirconia and the color zirconia conforms to ISO 13356:2015. Implants for surgery -Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP) and the performance conforms to ISO 6872:2015/Amd1:2018, Dentistry: Ceramic Materials.

    AI/ML Overview

    Here's an analysis of the provided FDA 510(k) summary regarding the Perfit FS Dental Zirconia Fully Sintered Block, focusing on acceptance criteria and the study proving it meets those criteria:

    Device: Perfit FS Dental Zirconia Fully Sintered Block
    K Number: K222686
    Predicate Device: K203590, "Perfit FS Dental Zirconia Blank, Perfit ZR UT Dental Zirconia Blank" manufactured by Vatech acucera, Inc.

    This submission is for a dental material (zirconia blocks) used for creating dental restorations, not an AI-powered diagnostic device. Therefore, many of the typical acceptance criteria and study designs relevant to AI (like MRMC studies, human-in-the-loop performance, expert consensus for ground truth, training set information) are not applicable to this type of medical device.

    The acceptance criteria here pertain to the physical, chemical, and biological properties of the material to ensure it is safe and performs as intended.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from ISO Standards / Predicate Device)Reported Device Performance (Subject Device)
    Material Composition Conformity
    ZrO2 + Y2O3 + Nb2O5 + Al2O3 ≥ 99.0% (Regular White zirconia)Composed of ZrO2+Y2O3+Nb2O5 and additional inorganic pigment: Al2O3 and Other oxide (white zirconia)
    ZrO2 + Y2O3 + Nb2O5 + Al2O3 ≥ 99.0% (Pre-Shaded zirconia)Derived from same zirconia powder as regular white zirconia with addition of Fe2O3 and other oxide (colored zirconia)
    Conformity to ISO 13356:2015 (Material)Conforms to ISO 13356:2015
    Inorganic pigments (Fe2O3, MnO2, and Er2O3 < 1.0%)Inorganic pigments specified (Fe2O3 and other oxide for colored zirconia) < 1.0% (as per predicate)
    Mechanical/Physical Properties Conformity
    Conformity to ISO 6872:2015/Amd1:2018 (Performance)Conforms to ISO 6872:2015/Amd1:2018
    Flexural strength ≥ 500 MPa≥ 500 MPa
    Solubility < 2000 µg/cm²< 2000 µg/cm²
    Radioactive: uranium-238 active concentration ≤ 1.0 Bq/guranium-238 active concentration ≤ 1.0 Bq/g
    Biocompatibility Conformity
    Tested to ISO 10993-1:2018 (Biological evaluation)Tested to ISO 10993-1:2018
    Tested to ISO 10993-3:2014 (Genotoxicity, carcinogenicity, reproductive toxicity)Tested to ISO 10993-3:2014
    Tested to ISO 10993-5:2009 (In Vitro cytotoxicity)Tested to ISO 10993-5:2009
    Tested to ISO 10993-6 (Local Effects after Implantation)Tested to ISO 10993-6
    Tested to ISO 10993-10:2010 (Irritation and skin sensitization)Tested to ISO 10993-10:2010
    Tested to ISO 10993-11:2017 (Systemic toxicity)Tested to ISO 10993-11:2017
    Functional Equivalence
    Identical Indications for UseIdentical (for dental restorations using CAD/CAM or manual milling machines)
    Block formBlock form
    Various dimensionsVarious
    Applied for (Milling Machine): CEREC inLab (Sirona)CEREC inLab (Sirona)
    Conditions of Use: Professional use for fabrication of artificial teeth in fixed or removable dentures, crowns, facings, and veneersIdentical
    Single Use: YesYes
    Supplied Sterile: NoNo

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for the "test set" in the traditional sense of a clinical or image-based study. Instead, it refers to bench testing of the material itself according to various ISO standards. These standards typically specify the number of samples required for each particular test (e.g., flexural strength, solubility, biocompatibility assays). The provenance of the data is the manufacturer, Vatech acucera, Inc., as they performed the comparative study and conducted the bench tests. It is retrospective in the sense that the results are presented after the tests were completed. There is no information regarding the country of origin of "data" in the context of patient data, as this is a material science evaluation.


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This question is not applicable for this type of device. The "ground truth" for material properties is established by adherence to recognized international standards (ISO standards) and by direct measurement of physical, chemical, and biological characteristics, not through expert human interpretation of images or clinical outcomes in the same way an AI diagnostic device would be evaluated. The "experts" involved would be material scientists, chemists, and toxicologists who conducted and interpreted these standardized tests.


    4. Adjudication Method for the Test Set

    This question is not applicable for this type of device. Adjudication methods like "2+1" or "3+1" are used to establish consensus or resolve discrepancies among human readers or experts when determining ground truth for clinical or image-based assessments. For material testing, the "ground truth" is derived directly from standardized measurement procedures and acceptance limits defined by the ISO standards.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No. An MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the performance of diagnostic tools (often AI-assisted) and how they impact human reader performance. The Perfit FS Dental Zirconia Fully Sintered Block is a dental restoration material, not a diagnostic device.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a material, not an algorithm. Therefore, "standalone performance" in the context of an algorithm does not apply.


    7. The Type of Ground Truth Used

    The ground truth used for this device is based on:

    • International Standards: Adherence to the specifications and performance criteria defined in recognized ISO standards (e.g., ISO 6872:2015, ISO 13356:2015, ISO 10993 series for biocompatibility).
    • Direct Measurement & Material Analysis: Physical, chemical, and biological properties (e.g., flexural strength, solubility, elemental composition, cytotoxicity) are directly measured and compared against the defined limits within these standards and against the predicate device.
    • Predicate Device Equivalence: The subject device's properties (material, performance, indications for use) are directly compared to those of the legally marketed predicate device (K203590) to demonstrate substantial equivalence.

    8. The Sample Size for the Training Set

    Not applicable. This device is a material, not a machine learning model. Therefore, there is no "training set."


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for a material, this question does not apply.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1