Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222608

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-16

    (109 days)

    Product Code
    Regulation Number
    880.6880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The pre-vacuum steam sterilizer models 3870HSG and 3870HSG-WS are in hospitals and other healthcare facilities (e.g. hospitals, nursing homes, extended-care facilities, freestanding surgical centers, clinics, and medical and dental offices) for the purpose of sterilizing heat stable medical devices: wrapped and unwrapped solids (IUSS), hollow and porous products.

    Device Description

    The autoclave models 3870HSG-WS and the 3870HSG are dynamic-air-removal steam-heated sterilizers that are using steam as the sterilizing agent, which are suitable for large medical centers, dental clinics and operating rooms for the purpose of sterilizing heat stable medical devices: wrapped and unwrapped solids (IUSS), hollow and porous products.

    The model 3870HSG-WS has an automatic door and is designed to be a fast cycle sterilizer that is mobile and can operate independent of building utilities due to its internal water reservoirs. The autoclave includes the option of water recycling, which allows the operator to manually fill the tap water reservoir with water that will be reused for 10 cycles, after which the reservoir is automatically drained and, in case the device is connected to a water inlet connection, be refilled automatically.

    The model 3870HSG has an automatic door and is designed to be a fast cycle sterilizer. Contrary to the 3870HSG-WS, the 3870HSG need to be connected to the building utilities system (the water system). For both autoclave configurations (HSG/HSG-WS) a 9kW built-in steam generator is used to supply the steam for the sterilization process.

    The autoclaves are Class B devices equipped with a vacuum system that allows:

    • . Removal of residual air from packs and porous load and most kinds of tubes (rubber, plastic etc.) by vacuum at the first stage of the cycle.
    • . Better steam penetration into the load; resulting in effective sterilization.
    • . Better temperature uniformity.
    • . Post sterilization drying phase - during the drying stage draws air through a HEPA filter (0.2um) and pushes that air through the heated chamber to remove moisture and facilitate the drying operation. Drying is performed with the door closed.

    The autoclave models 3870HSG-WS and 3870HSG are stand-alone devices that do not need to interact with other devices or to interact with any person(s) or patient(s) besides the current device operator.

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of a study conducted for the Tuttnauer Pre-vacuum Steam Sterilizer models 3870HSG and 3870HSG-WS. This is a medical device (a sterilizer), and the study focuses on its performance in sterilizing medical devices.

    Here's an breakdown of the information requested based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device: Tuttnauer Pre-vacuum Steam Sterilizer models 3870HSG and 3870HSG-WS

    Test NamePurposeAcceptance Criteria (from document)Reported Device Performance (Results)
    Electrical SafetyVerifying that device and its components meet electrical safety requirements.Meeting standard specificationPass
    EMCVerifying that the device meets EMC requirements.Meeting standard specificationPass
    Pressure vessel testingVerifying that the pressure vessel used for the T-Edge meets the requirements for pressure vessel and is safe for use.Meeting standard specificationPass
    Bowie & Dick testVerify air removal performance (for dynamic air removal sterilizers).The Bowie-Dick test indicator sheet shall show a uniform color change.Pass
    Air-leak-rate (vacuum) testVerify air removal performance (for dynamic air removal sterilizers).Average leak rate of 1 millimeter of mercury (mmHg) (0.13 kPa) (0.019 psia) per min or less over the measured time interval.Pass
    Empty chamber tests (121°C/132°C) - on wrapped and unwrapped (IUSS) loadTo ensure that the sterilizer is capable of providing steady-state thermal conditions within the chamber consistent with the desired sterility assurance level (SAL) in the load.The temperature shall not exceed more than 3°C above the sterilization temperature. The temperature shall not be below the sterilization temperature. Actual exposure time.Pass
    Full chamber load test (121°C/132°C) - on wrapped and unwrapped (IUSS) loadTo ensure that the sterilizer is capable of providing steady-state thermal conditions within the chamber consistent with the desired sterility assurance level (SAL) in the load.The temperature shall not exceed more than 3°C above the sterilization temperature. The temperature shall not be below the sterilization temperature. Actual exposure time.Pass
    Moisture retention tests - fabric PCD and wrapped instruments PCDTo ensure that the sterilizer is capable of meeting the moisture retention criteria.Moisture retained by the fabric PCD shall cause no more than a 3% increase in pre-sterilization test pack weight, and the pack shall exhibit no wet spots. When examined immediately after completion of the cycle, the wrapped instrument packs shall have no wet spots on the outer wrappers. Moisture retained by the 100% cotton towel shall cause no more than a 20% increase in the pre-sterilization weight of the towel.Pass
    Biological performance with a textile PCDVerifying biological performance on half sterilization time.Tested cycle has a 10⁻⁶ SAL or an SAL providing a greater assurance of sterility when the textile PCD is used.Pass
    Biological performance with wrapped instrument PCDVerifying biological performance on half sterilization time.Tested cycle has a 10⁻⁶ SAL or an SAL providing a greater assurance of sterility when the wrapped instrument PCD is used.Pass
    Biological performance with unwrapped instrument PCD (IUSS) – minimum and maximum load testsVerifying biological performance on half sterilization time.There shall be no growth observed in the vials containing turbines or in the extraction of any of the turbines, except for the positive controls. No growth shall be observed with the BIs except the positive control BIs. Growth should be observed for the positive control turbine and BI.Pass

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document describes "tests" and "cycles" but does not explicitly state a quantitative sample size (e.g., number of sterilization runs) for each specific test. It implies multiple trials were conducted to meet each acceptance criterion, as performance tests like "Empty chamber tests" or "Biological performance" would typically involve multiple runs to demonstrate consistency and reliability. However, no specific quantitative sample size is provided in the document for the test set.
    • Data Provenance: The document does not specify the country of origin of the data. It is a submission to the U.S. Food & Drug Administration (FDA) by Tuttnauer LTD. based in Jerusalem, Israel, implying the tests were conducted by the manufacturer or a contracted lab. The tests are described as non-clinical tests, which means they are laboratory-based and simulated scenarios, not directly from patient data. The context suggests these were prospective tests performed specifically for this 510(k) submission, rather than retrospective analysis of existing data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not mention human expert involvement in establishing ground truth for the test set. The tests described are objective, quantitative performance measurements of a sterilizer (e.g., temperature, pressure, leak rate, biological indicator growth). Ground truth is established by physical measurements and the absence/presence of microbial growth, not by expert consensus on interpretations.

    4. Adjudication method for the test set

    Not applicable. As noted above, the ground truth is established by objective physical and biological measurements, not through human interpretation requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done.
    • This device is a steam sterilizer, not an AI-powered diagnostic imaging device or an AI assistant for human readers. Therefore, the concepts of human readers, AI assistance, or MRMC studies are not relevant to this type of device and its performance evaluation as described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, in essence, standalone performance was assessed. The device itself is an automated sterilizer. Its "performance" refers to its ability to achieve specific physical parameters (temperature, pressure, dry time) and biological outcomes (sterility assurance level, lack of growth). The tests described (e.g., Bowie & Dick test, air-leak-rate test, empty/full chamber load tests, biological performance tests) directly measure the sterilizer's function without human intervention during the cycle itself. Human operators set up and observe the results, but the "performance" being evaluated is the device's independent operation.

    7. The type of ground truth used

    The ground truth used for these tests is based on:

    • Physical measurements: Temperature (within a specified range), pressure (leak rate), weight change (moisture retention). These are objectively measurable physical properties.
    • Biological outcomes: The absence of microbial growth in biological indicators (BIs) or test turbines after sterilization, demonstrating a sterility assurance level (SAL) of 10⁻⁶. Positive controls are used to confirm test viability. This is an objective biological outcome.
    • Visual inspection: Uniform color change for the Bowie-Dick test, absence of wet spots for moisture retention.

    8. The sample size for the training set

    This document describes the validation of a physical device (a steam sterilizer). It is not an AI/ML device, so there is no "training set" in the context of machine learning model development. The device operates based on preprogrammed cycles and physical principles, not on learned data.

    9. How the ground truth for the training set was established

    Since this is not an AI/ML device, there is no training set and therefore no ground truth established for a training set. The performance is assessed against established engineering and biological standards for sterilization.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1