Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222601

    Validate with FDA (Live)

    Date Cleared
    2022-10-27

    (59 days)

    Product Code
    Regulation Number
    866.4001
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BioFire® FilmArray® Pneumonia Panel plus (BioFire Pneumonia Panel plus) is a multiplexed nucleic acid test intended for use with BioFire® FilmArray® 2.0 (BioFire® FilmArray® Torch (BioFire Torch) systems for the simultaneous detection and identification of nucleic acids from Middle East Respiratory Syndrome Coronavirus (MERS-CoV) and multiple respiratory viral and bacterial nucleic acids, as well as select antimicrobial resistance genes, in sputum-like specimens (induced or expectorated sputum, or endotracheal aspirates) or bronchoalveolar lavage (BAL)-like specimens (BAL or mini-BAL) obtained from individuals meeting MERS-CoV clinical and/or epidemiological criteria.

    Testing with BioFire Pneumonia Panel plus should not be performed unless the patient meets clinical and/or epidemiologic criteria for testing suspected MERS-CoV specimens. This includes: clinical signs and symptoms associated with MERS-CoV infection, contact with a probable or confirmed MERS-CoV case, history of travel to geographic locations where MERS-CoV cases were detected, or other epidemiological links for which MERS-CoV testing may be indicated.

    The following bacteria are reported semi-quantitatively with bins representing approximately 10^4, 10^5, or ≥10^7 genomic copies of bacterial nucleic acid per milliliter (copies/mL) of specimen, to aid in estimating relative abundance of nucleic acid from these common bacteria within a specimen:

    Bacteria reported with bins of 10^4, 10^5, 10^6, or ≥10^7 copies/mL

    • · Acinetobacter calcoaceticus-baumannii complex
    • · Enterobacter cloacae complex
    • · Escherichia coli
    • · Haemophilus influenza
    • · Klebsiella aerogenes
    • · Klebsiella oxytoca
    • · Klebsiella pneumoniae group
    • · Moraxella catarrhalis
    • · Proteus spp.
    • · Pseudomonas aeruginosa
    • · Serratia marcescens
    • · Staphylococcus aureus
    • · Streptococcus agalactiae
    • · Streptococcus pneumoniae
    • · Streptococcus pyogenes

    The following atypical bacteria, viruses, and antimicrobial resistance genes are reported qualitatively:

    Atypical Bacteria

    • · Chlamydia pneumoniae
    • · Legionella pneumophila
    • · Mycoplasma pneumoniae

    Viruses
    · Middle East respiratory syndrome coronavirus (MERS-CoV)

    • · Adenovirus
    • · Coronavirus
    • · Human metapneumovirus
    • · Human rhinovirus/enterovirus
    • · Influenza A virus
    • · Influenza B virus
    • · Parainfluenza virus
    • · Respiratory syncytial virus

    Antimicrobial Resistance Genes

    • · CTX-M
    • · IMP
    • · КРС
    • · NDM
    • · OXA-48-like
    • · VIM
    • · mecA/C and MREJ (MRSA)

    The detection and identification of specific viral and bacterial nucleic acids from MERS-CoV and other respiratory pathogens, as well as the estimation of relative abundance of nucleic acid from common bacterial analytes, within specimens collected from individuals meeting MERS-CoV clinical and/or epidemiological criteria aids in the differential diagnosis of MERS-CoV infection, if used in conjunction with other clinical and epidemiological information in accordance with the guidelines provided by the appropriate public health authorities.

    BioFire Pneumonia Panel plus MERS-CoV positive results are for the identification of MERS-CoV. The definitive identification of MERS-CoV requires additional testing and confirmation procedures in consultation with the appropriate public health authorities (e.g., local or state public health departments, etc.) for whom reporting is necessary. The diagnosis of MERS-CoV infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence in addition to the identification of MERS-CoV.

    BioFire Pneumonia Panel plus MERS-CoV negative results, even in the context of a BioFire Pneumonia Panel plus positive result for one or more of the common respiratory pathogens, do not preclude MERS-CoV infection and should not be used as the sole basis for patient management decisions. The levels of MERS-CoV that would be present in sputum-like or BAL-like specimens from individuals with early infection and from asymptomatic MERS-CoV cariers are not well understood. A negative BioFire Pneumonia Panel plus MERS-CoV result in an asymptomatic individual does not rule out the possibility of future illness and does not demonstrate that the individual is not infectious.

    Viral culture should not be attempted on specimens with positive BioFire Pneumonia Panel plus results for MERS-CoV unless a BSL 3 facility is available to receive and culture specimens.

    Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test, pathogens below the limit of detection, or in the case of bacterial analytes, present at levels below the lowest reported 10^4 copies/mL bin. Detection of analytes does not rule out co-infection with other organisms; the agent(s) detected by the BioFire Pneumonia Panel plus may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible lower respiratory tract infection.

    Detection of bacterial nucleic acid may be indicative of colonizing or normal respiratory flora and may not indicate the causative agent of pneumonia. Semi-quantitative Bin (copies/mL) results generated by the BioFire Pneumonia Panel plus are not equivalent to CFU/mL and do not consistently correlate with the quantity of bacterial analytes compared to CFU/ mL. For specimens with multiple bacteria detected, the relative abundance of nucleic acids (copies/mL) may not correlate with the relative abundance of bacteria as determined by culture (CFU/mL). Clinical correlation is advised to determine significance of semi-quantitative Bin (copies/mL) for clinical management.

    The antimicrobial resistance gene detected may or may not be associated with the agent(s) responsible for disease. Negative results for these antimicrobial resistance gene assays do not indicate susceptibility to corresponding classes of antimicrobials, as multiple mechanisms of antimicrobial resistance exist.

    Antimicrobial resistance can occur via multiple mechanisms. A "Not Detected" result for a genetic marker of antimicrobial resistance does not indicate susceptibility to associated antimicrobial drugs or drug classes. A "Detected" result for a genetic marker of antimicrobial resistance cannot be definitively linked to the microorganism(s) detected. Culture is required to obtain isolates for antimicrobial susceptibility testing, and BioFire Pneumonia Panel plus results should be used in conjunction with culture results for determination of bacterial susceptibility or resistance.

    Due to the genetic similarity between human rhinovirus and enterovirus, the test cannot reliably differentiate them. A positive Rhinovirus/Enterovirus result should be followed up using an alternate method (e.g., cell culture or sequence analysis) if differentiation is required.

    Culture is required to identify pathogens not detected by the BioFire Pneumonia Panel plus, to further speciate analytes in genus, complex, or group results if desired, to identify bacterial pathogens present below the 10°4 copies/mL bin if desired, and for antimicrobial susceptibility testing.

    Device Description

    The FilmArray Pneumonia Panel plus is designed to simultaneously identify MERS-CoV and 26 potential pathogens of lower respiratory tract infection (LRTI) and associated antimicrobial resistance (AMR) genes from a sputum-like (induced and expectorated sputum as well as endotracheal aspirate, ETA) or bronchoalveolar lavage (BAL)-like (BAL and mini-BAL) specimens obtained from individuals meeting MERS-CoV clinical and/or epidemiological criteria in a time (~1 hour). The FilmArray Pneumonia Panel plus is compatible with BioFire Diagnostics' (BioFire) PCR-based in vitro diagnostic BioFire FilmArray 2.0 (K143178) and BioFire FilmArray Torch (K160068) systems for infectious disease testing. A specific software module (i.e., FilmArray Pneumonia Panel plus pouch module) is used to perform FilmArray Pneumonia Panel plus testing on these systems.

    A test is initiated by loading Hydration Solution into one port of the FilmArrav pouch and a soutumlike or BAL-like sample mixed with the provided Sample Buffer into the port of the FilmArray Pneumonia Panel plus pouch and placing it in a FilmArray instrument. The pouch contains all the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArray Software quides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.

    The FilmArray instrument contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically-controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the PCR reactions and the melt curve analysis.

    Nucleic acid extraction occurs within the FilmArray pouch using mechanical and chemical lvsis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the FilmArray performs a nested multiplex PCR that is executed in two stages. During the first stage, the FilmArray performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the end of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.

    The FilmArray Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

    AI/ML Overview

    The provided text is a summary of a Special 510(k) submission for the BioFire FilmArray Pneumonia Panel plus. The purpose of this submission is solely to modify the labeling of the device to include new limitations regarding the Adenovirus2 assay's performance closer to its expiration date.

    Therefore, the document does not describe a new study to prove device performance against acceptance criteria. Instead, it refers to a problem identified through stability studies and a voluntary recall. The premise of this 510(k) submission is that because there are no changes to the actual device, its fundamental scientific technology, performance, and risk are unchanged from the legally marketed predicate device (K181324).

    The acceptance criteria table below is inferred from the identified issue and the proposed label modifications, not from a performance study demonstrating device capabilities.


    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (from stability study and recall)
    Adenovirus C Detection Sensitivity (within 6 months of expiration)Impaired: For Adenovirus species C, when pouches are within 6 months of expiration, there is an increased risk of false negative results due to a 10-100x loss in sensitivity (i.e., impairment leading to an increase in the Limit of Detection - LoD). This led to a voluntary recall (Recall Event 88117/ Z-2039-2021, Z-2040-2021).
    Adenovirus C Detection Sensitivity (more than 6 months from expiration)Unimpacted: The test performance for Adenovirus species C for pouches more than 6 months from the expiration date is not impacted.
    Other Adenovirus Species Detection PerformanceUnimpacted: Performance for other adenovirus species is not impacted regardless of the pouch's proximity to the expiration date.
    General Device Performance (excluding identified Adenovirus C issue)Unchanged/Equivalent to Predicate: "The fundamental scientific technology, performance, and risk of the FilmArray Pneumonia Panel plus is unchanged from the legally marketed FilmArray Pneumonia Panel plus." This implies that for all other analytes and under normal conditions (i.e., pouches not within 6 months of expiration for Adenovirus C detection), the device performance meets the established criteria of the predicate device (K181324).

    Study Details (Based on the information provided)

    1. Sample size used for the test set and the data provenance:

      • The document does not explicitly state the sample size of the test set used in the stability study that identified the Adenovirus C issue.
      • The data provenance is from stability studies conducted by BioFire Diagnostics, LLC. It is retrospective in the sense that the issue was discovered after initial marketing and through ongoing stability monitoring of manufactured product.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided as the document describes a stability study related to device shelf-life rather than a diagnostic performance study against a clinical ground truth. The "ground truth" here would have been the presence/absence and concentration of adenovirus species C in the stability samples, likely confirmed by a reference method in a laboratory setting. Details on experts or their qualifications are not given.
    3. Adjudication method for the test set:

      • This information is not provided. Given it's a stability study on a diagnostic assay, adjudication in the context of expert consensus on clinical cases is not directly applicable. The determination of "false negative" would have been based on comparison to expected results or a reference method.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is a multiplexed nucleic acid test (an in vitro diagnostic device), not an imaging AI device that involves human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • The FilmArray Pneumonia Panel plus is an automated diagnostic assay. Its performance (detection of nucleic acids) is inherently "standalone" in the sense that the instrument and software interpret the results without human interpretation of raw data. The stability study assessed this standalone performance.
    6. The type of ground truth used:

      • For the stability studies that identified the issue, the ground truth was likely analytical truth (known presence, absence, and concentration of specific analytes like Adenovirus species C) in manufactured control samples or spiked specimens. This is standard for stability testing of IVD assays.
    7. The sample size for the training set:

      • This document describes a modification to labeling based on post-market stability findings, not a de novo development or training of an algorithm. Therefore, information about a "training set" is not applicable and not provided.
    8. How the ground truth for the training set was established:

      • As there is no mention of a training set for an algorithm, this information is not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1