Found 1 results

510(k) Data Aggregation

    K Number
    K222529

    Validate with FDA (Live)

    Date Cleared
    2022-11-14

    (84 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

    Device Description

    The Surgical Face Mask is single use, three-layer, flat-pleated style with ear loops/tie-coverall and nose piece. The Surgical Face Mask is manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ties/ear loops are held in place over the users' mouth and nose by two ties/ear loops welded to the facemask. The loops are made of polyester and spandex, the ties are made of spun-bond polypropylene. The nose clamp in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of polypropylene and aluminum. The Surgical Face Mask will be provided in blue. The Surgical Face Mask is sold as non-sterile and are intended to be single use, disposable devices.

    AI/ML Overview

    The provided text describes the regulatory clearance for a Surgical Face Mask (K222529). This is a medical device, not an AI/ML-driven device, therefore, many of the typical acceptance criteria and study designs associated with AI/ML systems (such as expert consensus, MRMC studies, training set details, or an 'algorithm only' performance) are not applicable.

    The acceptance criteria here pertain to the physical and performance characteristics of the surgical face mask, comparing it to established industry standards (ASTM, ISO, EN, CFR) for medical face masks. The study proving the device meets these criteria is a non-clinical performance and biocompatibility testing.

    Here's the information extracted from the provided text, adapted to the requested format where applicable, and noting when a requested AI/ML specific criterion is not relevant to this type of device:


    Device Name: Surgical Face Mask (Xiantao Rayxin Medical Products Co., Ltd.)
    Regulatory Classification: Class II Surgical Apparel (Product Code FXX, 21 CFR 878.4040)
    Predicate Device: Surgical Face Mask (K212368, Anhui Tiankang Medical Technology Co., Ltd.)

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/CriterionPurposeAcceptance CriteriaReported Results
    Synthetic Blood Penetration (ASTM F1862)Demonstrate resistance to liquid penetrationLevel 1: 29 samples out of 32 pass (AQL 4%) at 80mmHg Level 3: 29 samples out of 32 pass (AQL 4%) at 160mmHgLevel 1: Pass - 32 out of 32 pass at 80mmHg (3 lots) Level 3: Pass - 32 out of 32 pass at 160mmHg (3 lots)
    Particulate Filtration Efficiency (PFE) (ASTM F2299)Demonstrate particulate filtrationLevel 1: ≥95% Level 3: ≥98%Level 1: Pass - Average 98.50% Level 3: Pass - Average 98.83%
    Bacterial Filtration Efficiency (BFE) (ASTM F2101)Demonstrate bacterial filtrationLevel 1: ≥95% Level 3: ≥98%Level 1: Pass - Average 99.27% Level 3: Pass - Average 99.79%
    Differential Pressure (Delta P) (Breathability) (EN 14683 Annex C)Demonstrate breathabilityLevel 1: ≤ 5.0 mmH2O/cm² Level 3: ≤ 6.0 mmH2O/cm²Level 1: Pass - Average 3.11 mmH2O/cm² Level 3: Pass - Average 3.97 mmH2O/cm²
    Flammability (16 CFR 1610)Demonstrate flame resistanceClass IPass
    Cytotoxicity (ISO 10993-5)Demonstrate cytotoxic biocompatibilityNon-cytotoxicPass - Under the conditions of the study, the device is non-cytotoxic.
    Skin Irritation (ISO 10993-10)Demonstrate non-irritabilityNon-irritatingPass - Under the conditions of the study, the device is non-irritating.
    Skin Sensitization (ISO 10993-10)Demonstrate non-sensitizationNon-sensitizingPass - Under the conditions of the study, the device is non-sensitizing.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Synthetic Blood Penetration: 32 samples per test condition for each of 3 lots.
    • Other Performance Tests (PFE, BFE, Delta P, Flammability): Specific sample sizes are not explicitly stated for each of these, but performance is reported as averages or 'Pass' implying multiple samples were tested to demonstrate compliance with the standards.
    • Biocompatibility Tests: Sample sizes are not explicitly stated.
    • Data Provenance: The tests appear to be laboratory performance tests, not clinical data from patients. The sponsor, Xiantao Rayxin Medical Products Co., Ltd., is based in China. The specific location where these tests were performed is not detailed, but it would typically be a certified testing laboratory. The data is based on prospective testing of the manufactured devices to the specified standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Not applicable. This is a hardware device where ground truth is established through standardized laboratory measurements and physical properties, not human interpretation or expert consensus. The "ground truth" is defined by the performance standards themselves (e.g., a specific percentage filtration, a specific pressure differential).

    4. Adjudication Method for the Test Set:

    • Not applicable. See point 3. Testing involves objective measurements against established physical and chemical standards, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No. This is not an AI/ML-assisted diagnostic device, so an MRMC study is not relevant. The device's effectiveness is based on its physical barrier properties and breathability, not its ability to assist human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • No. This is a physical product (surgical face mask), not an algorithm or software. The concept of "standalone performance" for an algorithm is not applicable here.

    7. The Type of Ground Truth Used:

    • Standardized Physical and Chemical Measurements & Definitions: Ground truth is defined by the requirements of the specific ASTM, ISO, EN, and CFR standards referenced. For example, "Level 1 pass at ≥95%" for BFE means that any value of 95% or higher, as measured by the ASTM F2101 method, constitutes a 'pass'. This is an objective, quantitative ground truth.

    8. The Sample Size for the Training Set:

    • Not applicable. This is a manufactured product, not an AI/ML model that undergoes a "training" phase.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. See point 8.