    Device Name
    Surgical Gown
    Date Cleared
    2023-01-26

    (170 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. In addition, this surgical gown meets the requirements of AAMI Level 4 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). The Surgical Gowns are single use, disposable medical devices, provided sterile.

    Device Description

    The Surgical Gown is composed of collar, body, sleeve and tie. The back is full opening, the neck and waist are laced, the sleeve are made of polyester elastic closure by sewing, and the rest are made of SMS nonwoven material, in a blue color. It has been tested according to AAMI PB70:2012 and meet AAMI Level 4 barrier level protection for a surgical gown. The proposed devices are single use, disposable medical devices and EO that sterilized. They are available in six sizes: S, M, L, XL, XXL and XXXL

    AI/ML Overview

    This document describes the safety and performance of a Surgical Gown, not an AI-powered medical device. Therefore, many of the requested details about acceptance criteria for AI model performance and ground truth establishment are not applicable.

    However, I can extract the relevant acceptance criteria and the reported device performance for the physical surgical gown based on the provided text.


    Device: Surgical Gown (K222403)
    Intended Use: To protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids, and particulate material during surgical procedures, meeting AAMI Level 4 barrier protection.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance is assessed through non-clinical laboratory tests.

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance (Results)
    AATCC 42Assess resistance to water impact penetrationLevel 4, ≤1.0gPass
    AATCC 127Assess hydrostatic resistanceLevel 4, ≥50cmPass
    ASTM D5034Assess adequate tensile strengthWarp/Length ≥ 30N; Weft/Width ≥ 30NPass
    ASTM D5587Assess adequate tear resistanceWarp/Length ≥ 10N; Weft/Width ≥ 10NPass
    ASTM D1683Assess adequate seam strengthSleeve Seam: ≥ 30N; Armhole Seam: ≥ 30N; Shoulder Seam: ≥ 30NPass
    ASTM F1868Assess evaporative resistance of fabrics(Ref)≥0.06(kPa·m²/W)Pass
    ASTM F1671Resistance to Penetration by Blood-Borne Pathogens using Phi-X174 BacteriophageNo penetration (Assay Titer < 1 PFU/ml)Pass
    ISO 9073-10Assess acceptable lint and other particles generation in the dry stateLog10 < 4Pass
    16 CFR Part 1610Flammability testingClass 1Pass
    ISO 11737-2Sterility assurance10-6Pass
    ISO 10993-5Biocompatibility - cytotoxicityNon-cytotoxicPass
    ISO 10993-10Biocompatibility - irritationNon-irritatingPass
    ISO 10993-10Biocompatibility - sensitizationNon-sensitizingPass
    ISO 10993-7Verify acceptable sterilant residualsEO residual ≤ 4 mg/device; ECH residual ≤ 0 mg/devicePass

    Missing/Not Applicable Information (for AI/Software Device)

    The following points are not applicable to this submission as it concerns a physical medical device (surgical gown) rather than a software or AI-powered device. The document explicitly states "No clinical study is included in this submission."

    1. Sample size used for the test set and the data provenance: Not applicable. Tests are performed on physical samples of the surgical gown materials and finished products, not a "test set" of data in the AI sense.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth is established by standardized material testing methods, not expert human interpretation of data.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For this device, the "ground truth" is defined by the performance standards of the physical and chemical tests listed (e.g., specific thresholds for water penetration, tensile strength, sterility, biocompatibility).
    7. The sample size for the training set: Not applicable (no AI training involved).
    8. How the ground truth for the training set was established: Not applicable.