Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222239

    Validate with FDA (Live)

    Date Cleared
    2022-08-19

    (24 days)

    Product Code
    Regulation Number
    862.1358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SmartBolus Calculator is software intended for the management of diabetes in persons aged 2 and older requiring rapid-acting U-100 insulin. The SmartBolus Calculator calculates a suggested bolus dose based on user-entered carbohydrates, most recent sensor glucose value (or blood glucose reading if using fingerstick), rate of change of the sensor glucose (if applicable), insulin on board (IOB), and programmable correction factor, insulin to carbohydrate ratio, and target glucose value. The SmartBolus Calculator is intended for single patient, home use and requires a prescription.

    Device Description

    The SmartBolus Calculator is a software device that is a component of the Omnipod 5 Automated Insulin Delivery System. The SmartBolus Calculator exists on the Omnipod 5 App portion of the Omnipod 5 ACE Pump and relies on the user interface of the App. The SmartBolus Calculator receives input parameters and settings from the other components of the system and calculates a suggested bolus amount of insulin to correct an elevated glucose level (a correction bolus) and/or to cover carbohydrates from a meal (meal bolus). The SmartBolus Calculator allows users to have the option of populating the current estimated glucose value and trend, which is communicated by the connected iCGM. Users may also manually enter the estimated glucose value or a blood glucose (BG) reading from a blood glucose meter. In addition to glucose, the suggested bolus dose is calculated based on the following parameters: user-entered carbohydrates, rate of change of the sensor glucose (if using a CGM), correction factor, insulin to carbohydrate ratio, target qlucose value, and insulin on board (IOB). Once the calculation is complete, the user has the option of delivering the suggested dose of insulin, modifying the amount, or canceling. The SmartBolus Calculator can be used in the Omnipod 5 Automated Insulin Delivery System with both Manual Mode and Automated Mode.

    AI/ML Overview

    The information provided focuses on the substantial equivalence of the SmartBolus Calculator to a predicate device, specifically regarding the expansion of its age range of intended users. The document details the clinical study performed to support this expanded indication, rather than a general acceptance criteria study for the device's core functionality.

    Here's an analysis of the provided information based on your requested points:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal acceptance criteria in a tabular format for the SmartBolus Calculator's overall performance. Instead, the "acceptance criteria" can be inferred from the clinical study's primary objective, which was to evaluate safety based on glucose metrics. The study aimed to show that the use of the SmartBolus Calculator in the expanded age group (2.0-5.9 years) did not significantly worsen these safety metrics compared to manual entry of blood glucose values.

    Acceptance Criteria (Inferred from Study Objective)Reported Device Performance
    Safety Metrics (4-hour post bolus period):
    Mean % time Blood Glucose < 70 mg/dL (Phase 2 vs Phase 1)Mean decrease of 1.13% in Phase 2 (4.03%) vs. Phase 1 (5.16%). (P = 0.6250, not statistically significant)
    Mean % time Blood Glucose > 180 mg/dL (Phase 2 vs Phase 1)Mean decrease of 2.03% in Phase 2 (33.2%) vs. Phase 1 (35.2%). (P = 1.0000, not statistically significant)
    Number of deaths0
    Number of serious adverse events0
    Number of unanticipated adverse device effects (UADE)0
    Number of non-serious adverse events1 (prolonged hyperglycemia)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 5 subjects in the preschool cohort (aged 2.0-5.9 years).
    • Data Provenance: The study was a "single-arm, multi-center, prospective clinical study" conducted across "2 US clinical sites." Therefore, the data is prospective and from the USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document does not provide details on the number or qualifications of experts used to establish ground truth for the test set. The clinical study collected glucose metrics directly from the subjects, which served as the primary data for evaluation.

    4. Adjudication Method for the Test Set

    The document does not mention an explicit adjudication method for the clinical study's data. Clinical study data, particularly objective metrics like blood glucose levels, typically do not require adjudication in the same way imaging or subjective diagnostic interpretations might. Safety events would be reviewed by the study investigators and reported.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This study was a clinical trial evaluating the safety and effectiveness of the device in a specific patient population, not a comparative study of human readers with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This study was a human-in-the-loop study. The SmartBolus Calculator is a software device that recommends a bolus dose, but the user "has the option of delivering the suggested dose of insulin, modifying the amount, or canceling." The clinical study evaluated the device's performance when used by patients/caregivers in Manual Mode.

    7. The Type of Ground Truth Used

    The ground truth for the clinical study was based on objective clinical outcomes data (glucose metrics and adverse events) directly measured from the study participants during the two phases of the study (Manual Mode with manual BG entry vs. Manual Mode with SmartBolus Calculator use).

    8. The Sample Size for the Training Set

    The document does not provide information on the sample size for the training set used to develop the SmartBolus Calculator. The submission focuses on the clinical validation of the device's expanded indication, not its initial development.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1