Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222237

    Validate with FDA (Live)

    Device Name
    Cala kIQ
    Manufacturer
    Date Cleared
    2022-11-22

    (119 days)

    Product Code
    Regulation Number
    882.5897
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cala kIQ is indicated to aid in the temporary relief of hand tremors in the treated hand following stimulation in adults with essential tremor.

    Cala kIQ is indicated to aid in the temporary relief of postural and tremor symptoms that impact some activities of daily living in the treated hand following stimulation in adults with Parkinson's disease.

    Device Description

    Cala kIQ is a small, lightweight, wrist-worn stimulator device designed to aid in the relief of hand tremors by applying a calibrated Transcutaneous Afferent Patterned Stimulation (TAPS) to the median and radial nerves of a patient's wrist.

    The TAPS output pattern is calibrated to the individual patient's tremor frequency. During device setup, the patient performs a postural hold, during which the onboard sensors measure the tremor frequency which is then used to customize the TAPS output pattern to deliver therapy to the median and radial nerves at the appropriate, patient-specific, optimal stimulation frequency.

    Cala kIQ is comprised of the following components:

    • Stimulator: Contains sensors, electronics, and firmware for delivering TAPS therapy . and provides user interface
    • Band: Contains multi-use, conductive electrodes spaced at intervals to align with . targeted nerves and attachment points for Cala kIQ Stimulator
    • . Base Station: Provides charging function and contains an indicator light to alert the user to operating conditions.

    The Cala kIQ Stimulator contains the electronics, sensors, and firmware for calibrating and delivering TAPS therapy. The patient uses the Cala kIQ Stimulator buttons to complete device calibration, adjust the stimulation amplitude, and initiate therapy sessions. The Cala kIQ Stimulator is attached to the Cala kIQ band, which includes embedded electrodes placed at appropriate intervals to properly target the median and radial nerves. To accommodate a broad distribution of wrist sizes, the band is available in three sizes (small, medium, and large).

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Device: Cala kIQTM
    Predicate Device: Cala Trio™

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly define clear quantitative acceptance criteria for the Cala kIQ device across all metrics in the format you requested, especially for the Parkinson's Disease indication. Instead, it relies on demonstrating comparable effectiveness to the predicate device for Essential Tremor and provides descriptive results for Parkinson's Disease.

    However, based on the provided clinical data, we can infer performance and the study's findings directly:

    Acceptance Criterion (Inferred from Study Goals)Reported Device Performance (Cala kIQ for Parkinson's Disease)
    Aid in temporary relief of hand tremors in the treated hand (Essential Tremor)Demonstrated as equivalent to Cala Trio™: "Clinician-rated tremor assessments (TETRAS and MDS-UPDRS), and subject-rated assessments (BF- ADLs) results were similar for Cala kIQ and Cala Trio." (Study 2)
    Aid in temporary relief of postural and tremor symptoms involving some Activities of Daily Living (ADLs) in the treated hand (Parkinson's Disease)MDS-UPDRS Dominant Hand Tasks (average across 6 tasks): - Week 2: -0.44 ± 0.44 change (Improvement); 40% responder rate (≥0.5-point average improvement). - Week 3: -0.40 ± 0.37 change; 44% responder rate.BF-ADL Tasks (average across 8 tasks): - Week 2: -0.44 ± 0.43 change (Improvement); 43% responder rate (≥0.5-point average improvement). - Week 3: -0.32 ± 0.39 change; 25% responder rate.CGI-I (Clinical Global Impression – Improvement): - Week 2: 78% (31/40 subjects). - Week 3: 83% (30/36 subjects).PGI-I (Patient Global Impression – Improvement): - Week 2: 75% (30/40 subjects). - Week 3: 81% (29/36 subjects).
    Safety ProfileLow incidence of Adverse Events: No device-related serious adverse events reported in either study. All device-related adverse events were resolved with minimal intervention.
    Compliance with Performance StandardsConfirmed to meet physical, mechanical, shelf life, system performance, EMC, Electrical Safety (IEC 60601), and Biocompatibility (ISO 129993-1:2018) standards.

    2. Sample Size Used for the Test Set and Data Provenance

    Study 1 (Parkinson's Disease - standalone performance):

    • Sample Size: 40 patients were enrolled.
    • Data Provenance: Not explicitly stated, but the submission is for a US FDA clearance, implying results relevant to a US or general population. The study was described as a "single-arm, non-significant risk study," suggesting it was prospective.

    Study 2 (Comparative Effectiveness - Cala kIQ vs. Cala Trio):

    • Sample Size: 19 subjects completed the study, 6 of whom had Parkinson's Disease.
    • Data Provenance: Prospective, single-center study. Country of origin not explicitly stated, but again, implied to be relevant to a US FDA clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The ground truth was established through clinician-rated and subject-rated assessments:

    • Clinician-rated assessments: MDS-UPDRS (Movement Disorder Society-Unified Parkinson's Disease Rating Scale) and TETRAS (Tremor Essential Tremor Rating Assessment Scale) were used. The document does not specify the number or qualifications of the clinicians who performed these assessments. However, these are standardized scales typically administered by neurologists or movement disorder specialists.
    • Subject-rated assessments: BF-ADLs (Bain and Findley Activities of Daily Living) were used, representing patient self-reported outcomes.
    • Global Impression: CGI-I (Clinical Global Impression – Improvement) and PGI-I (Patient Global Impression – Improvement) were also used, which involve clinician judgment and patient self-perception, respectively.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method for reconciling disagreements between experts. The assessments are presumably taken as direct measurements by the administering clinicians or self-reported by patients. This indicates that assessments were likely unadjudicated for the purposes described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done in the traditional sense of evaluating multiple readers' interpretations of diagnostic images or data. The studies involved:

    • Study 1: A single-arm physiological intervention study assessing the device's direct effect on tremor and ADLs.
    • Study 2: A crossover study comparing the Cala kIQ directly against the Cala Trio device in the same subjects, with assessments made by clinicians and subjects. This is a head-to-head device comparison, not an MRMC study related to interpretation.

    Therefore, there is no discussion of the "effect size of how much human readers improve with AI vs. without AI assistance."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, Study 1 indirectly represents a standalone performance assessment of the Cala kIQ device (though not an "algorithm-only" assessment in the context of interpretation, rather a direct clinical effect). In this study, the device was used to deliver therapy, and its effect on patients' tremors and ADLs was measured. The device itself (Cala kIQ) functions autonomously in delivering calibrated TAPS therapy based on the patient's tremor frequency measured by onboard sensors. The reported results are the direct outcome of the device's therapy application.

    7. The Type of Ground Truth Used

    The ground truth for evaluating the device's effectiveness was a combination of:

    • Expert Clinical Assessment: Clinician-administered standardized scales (MDS-UPDRS, TETRAS, CGI-I) which reflect expert observation and rating of tremor severity and overall improvement.
    • Patient-Reported Outcomes (PROs): Subject-rated scales (BF-ADLs, PGI-I) which reflect the patient's direct experience of symptom relief and functional improvement.

    There was no pathology or specific outcomes data (like long-term mortality or disease progression) used as ground truth for this temporary relief indication.

    8. The Sample Size for the Training Set

    The document does not explicitly state the sample size of a "training set" for the Cala kIQ device or its underlying algorithm. The device's TAPS output pattern is "calibrated to the individual patient's tremor frequency" during device setup using "onboard sensors." This suggests a patient-specific calibration rather than a broad, pre-trained algorithm in the AI/ML sense that would require a large training dataset. The studies described are for clinical validation, not algorithm training. Previous predicate devices (DEN170028, K182706, K203288 for Cala Trio) likely contributed to the iterative development and refinement of the TAPS technology, but specific training set information is not provided.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, a distinct "training set" with ground truth in the traditional AI/ML sense is not described. The device calibrates to each individual patient's tremor frequency. The underlying science and engineering of Transcutaneous Afferent Patterned Stimulation (TAPS) and the ability to measure tremor frequency would have been developed and validated through initial research and development phases, but the details of this are not provided in this 510(k) summary. The clinical studies (Study 1 and Study 2) served as validation, not training.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1