Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222221

    Validate with FDA (Live)

    Device Name
    DEKA HELIX
    Date Cleared
    2023-01-04

    (163 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Helix Laser System with its accessories is intended for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery).

    CO2 Handpieces
    The Helix CO2 handpieces with wavelength of 10600 nm are indicated for use for the particular indications as follows:
    Dermatology & Plastic Surgery
    The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:

    • laser skin resurfacing
    • laser derm-abrasion
    • laser burn debridement.
      Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of: warts, acne scars, nevi epidermal, syringoma. Vaporization/coagulation of warts.

    Scanning unit
    The Helix Scanning unit, with wavelength of 10600 nm is indicated for:
    ·Laser skin resurfacing (ablation and/or vaporization) of soft tissue
    The Helix Scanning unit, with wavelength of 1570 nm, is indicated for:
    · Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue
    The Helix Scanning unit, with wavelengths of 10600 nm & 1570mm is indicated for laser skin resurfacing (ablation and/or vaporization) of soft tissue.

    Device Description

    The Helix system consists of:
    • Laser system console (containing the optical bench assembly and laser, the microcontroller control electronics and system software, the high voltage power supply, the laser cooling system, the compressed air-purge system, and the service panel)
    • LCD control panel with touch-screen technology: the LCD display provides information on the status and settings of the Helix system with touch-screen technology provided to input commands into the system.

    • Two wavelengths CO2 and 1570 nm
    • Articulated arm
    • Footswitch
    • Delivery devices (CO₂ non-fractional applicators, Scanner CO₂/1570nm fractional applicator)
      Electrical specifications are:
      100-230V ~ single phase, 50/60 Hz, Absorbed electric power 1500 VA (max)
    AI/ML Overview

    The DEKA HELIX is a laser surgical instrument for use in general, plastic surgery, and dermatology. The acceptance criteria and the study that proves the device meets the acceptance criteria are detailed below, based on the provided FDA 510(k) summary (K222221).

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) submission primarily demonstrates substantial equivalence to a predicate device (Alma Hybrid Laser System, K203441) rather than explicitly defining and meeting specific analytical or clinical performance acceptance criteria in the format of a separate study. Instead, the "acceptance criteria" are implied through the comparison of the proposed device's characteristics and performance to those of the legally marketed predicate device, and through compliance with recognized electrical safety and laser safety standards.

    The table below summarizes the comparison between the DEKA HELIX and its predicate device, highlighting where the proposed device is considered to be "as safe, as effective, and performs as well." Differences are noted as not affecting safety and effectiveness, implying performance within acceptable limits.

    Acceptance Criteria (Implied by Predicate Equivalence & Standards Compliance)Reported Device Performance (DEKA HELIX)Comment
    Indications for Use (General)Intended for surgical applications requiring ablation, vaporization, excision, incision, and coagulation of soft tissue in aesthetic surgery (dermatology and plastic surgery). Specific indications for CO2 handpieces and scanning unit are listed, including laser skin resurfacing for warts, acne scars, nevi epidermal, syringoma, and fractional skin resurfacing/coagulation with 1570 nm.Identical to predicate device in overall scope.
    Regulation Number21 CFR 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatologyIdentical
    Product CodeGEX, ONGIdentical
    System ControlFully computerizedIdentical
    Emission ControlFootswitch, InterlockIdentical
    User InterfaceTouch-screen display, Emergency button, Key switchIdentical
    Cooling SystemClosed loop, heat exchangerIdentical
    Emission IndicatorVisual, AudibleIdentical
    Electrical Requirements100-230 V, 16A 50/60 Hz, single phaseDifferences do not affect safety and effectiveness.
    Physical Dimensions and Weight62 cm x 63 cm x 138/202 cm (erected), ~ 70 KgDifferences do not affect safety and effectiveness.
    Laser TypeRF-excited CO2 laser, Class 4; Fiber Laser, Class 4Identical
    CO2 Handpieces Wavelength10600 nmIdentical
    CO2 Handpieces Pilot Beam635nm, 4mWDifferences do not affect safety and effectiveness.
    CO2 Handpieces Spot Size (mm)0.125 – 2.0 mmDifferences do not affect safety and effectiveness.
    CO2 Handpieces Pulse Duration10 - 900msDifferences do not affect safety and effectiveness.
    CO2 Handpieces Pulse Repetition Frequency0.2 — 200 HzDifferences do not affect safety and effectiveness.
    CO2 Handpieces Min Output Energy5 mJIdentical
    CO2 Handpieces Output Power70WDifferences do not affect safety and effectiveness (Predicate has 30W, 70W; Proposed has 70W).
    Scanning Unit Wavelength10600 nm and/or 1570 nmIdentical
    Scanning Unit Pilot Beam635nm, 4mWDifferences do not affect safety and effectiveness.
    Scanning Unit ScannerDual axis scannerIdentical
    Scanning Unit TipCooledIdentical
    Scanning Unit Spot Size @ 10600 nm0.374mmDifferences do not affect safety and effectiveness.
    Scanning Unit Scan Size @ 1570 nm20x20 mmDifferences do not affect safety and effectiveness.
    Scanning Unit Spot Size @ 1570 nm0.75mmIdentical
    Scanning Unit Output Energy @ 10600 nmMax 251mJDifferences do not affect safety and effectiveness.
    Scanning Unit Output Energy @ 1570 nmMax 144mJIdentical
    Scanning Unit Beam Density @ 10600 nm61% - 98% untreated tissue between spotsDifferences do not affect safety and effectiveness.
    Scanning Unit Beam Density @ 1570 nmUp to 400 pixels/cm2Differences do not affect safety and effectiveness.
    Electrical Safety & EMCCompliance with ANSI AAMI ES60601-1 and IEC 60601-1-2.Testing conducted and documented.
    Software Validation & VerificationConducted and documented as recommended by FDA guidance.Testing conducted and documented.
    Laser SafetyCompliance with IEC 60601-2-22 and IEC 60825-1.Testing conducted and documented.

    2. Sample Size Used for the Test Set and Data Provenance

    No clinical performance data (human subject studies) were conducted for this 510(k) submission. Therefore, there is no "test set" in the context of clinical data for which a sample size or data provenance would be specified. The submission relies solely on non-clinical performance data and comparison to a predicate device.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. Since no clinical performance data was used, there was no need for experts to establish ground truth for a test set.

    4. Adjudication Method

    Not applicable. No test set requiring adjudication was used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The submission explicitly states "Clinical Performance Data: None."

    6. Standalone Algorithm Performance Study

    Not applicable. This device is a hardware laser system, not an AI algorithm, so there is no standalone algorithm performance to evaluate.

    7. Type of Ground Truth Used

    Not applicable. No clinical performance data was presented, so no ground truth from pathology, outcomes data, or expert consensus was required for direct device evaluation in this submission. The ground truth for proving safety and effectiveness relies on established engineering principles, international safety standards compliance, and the equivalence to a legally marketed predicate device.

    8. Sample Size for the Training Set

    Not applicable. As a hardware device with no AI algorithm, there is no "training set" in the machine learning sense. The development of the device likely involved engineering design, prototyping, and iterative testing, but this is distinct from an AI training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There was no training set in the context of machine learning.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1