Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221885

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-10-26

    (119 days)

    Product Code
    Regulation Number
    862.3364
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 23andMe Personal Genome Service (PGS) is a qualitative genotyping assessment system applied to genomic DNA isolated from human saliva collected using the Oragene Dx OGD-500.001 to simultaneously detect, report, and interpret genetic variants in a broad multigene test. The assessment system is intended to enable users to access information about their genetics that could aid discussions with a healthcare professional. The 23andMe Personal Genome Service Pharmacogenetic Reports are indicated for reporting of the following variants:

    Gene: CYP2C19 Variant(s): *2, *3, *17
    Gene: CYP2C9 Variant(s): *2, *3, *5, *6, rs7089580
    Gene: CYP3A5 Variant(s): *3
    Gene: UGT1A1 Variant(s): *6, *28
    Gene: DPYD Variant(s):*2A, rs67376798
    Gene: TPMT Variant(s): *2, *3C
    Gene: SLCO1B1 Variant(s): c.521T>C (rs4149056)
    Gene: CYP2D6 Variant(s): *2, *3, *4, *5, *6, *7, *8, *9, *10, *11, *15, *17, *20, *29, *35, *40, *41

    This report is for over-the-counter use by adults over the age of 18 and provides genetic inform discussions with a healthcare professional about metabolism of therapeutics.

    The 23andMe Personal Genome Service pharmacogenetic reports for CYP2C9, CYP3A5, UGT1A1, DPYD, TPMT and CYP2D6 describe if a person has variants associated with metabolism of some therapeutics but does not describe if a person will or will not respond to a particular therapeutic and does not describe the association between detected variants and any specific therapeutic.

    23andMe Personal Genome Service pharmacogenetics report for CYP2C19 describes if a person has variants associated with metabolism of some therapeutics and provides interpretive drug information regarding the potential effect of the identified metabolizer phenotype on citalopram and clopidogrel therapy.

    23andMe Personal Genome Service pharmacogenetics report for SLCO1B1 describes if a person has variants associated with the processing of some therapeutics and provides interpretive drug information regarding the potential effect of the identified transport function phenotype on simvastatin therapy.

    The PGS Pharmacogenetic Reports are not a substitute for visits to a healthcare professional. The information provided by this report should not be used to start, stop, or change any course of treatment.

    Device Description

    The 23andMe Personal Genome Service (PGS) is a direct-to-consumer/over-the-counter, DNA testing service that provides information and tools for consumers to learn about and explore their DNA.

    The PGS is a currently marketed, non-invasive genetic information service that combines qualitative genotyping data covering genetic ancestry, traits, and certain heritable health conditions from a single multiplex assay with descriptive information derived from peer reviewed, published genetic research studies.

    Customer saliva specimens are self-collected using the Oragene-Dx® Device manufactured by DNA Genotek, Inc. cleared by FDA for use with the PGS device (K141410, DEN140044, DEN160026. DEN170046. DEN180028. K182784. K193492. and K211499), which consists of a sealable collection tube containing a stabilizing buffer solution. Once the sample is collected, it is shipped to a Clinical Laboratory Improvement Amendments (CLIA) certified laboratory for testing.

    DNA is isolated from the saliva and tested in a multiplex assay using a customized genotyping beadchip, reagents and instrumentation manufactured by Illumina. The device simultaneously tests for more than 600,000 variants, including those reported under the previously authorized PGS test indications.

    The raw data is generated using Illumina GenomeStudio software, and then sent to 23andMe for analysis and interpretation. The raw data received is analyzed using 23andMe's proprietary Coregen software, where a genotype is determined for each tested SNP. The results for certain of these SNPs are used to generate personalized reports for the customer that provide information about the detected genotype.

    Personalized reports are generated for each user that provide results of the testing performed. These reports tell the user which variant(s) has/have been detected in their sample and provide information on the metabolizer or transporter profile associated with the variant was detected, that information is also provided. The personalized reports are designed to present scientific concepts to users in an easy-to-understand format. The reports provide scientifically valid information about the metabolizer or transport function phenotype associated with the presence of a particular variant, or a combination of variants.

    In the pharmacogenetic report for SLCO1B1. information regarding interpretive drug information to certain medications will be provided to the user in a medication "mini report", which is accessed via a link in the pharmacogenetic report for SLCO1B1. The reports are designed to help users understand the meaning of their results and any appropriate actions that may be taken based on their results.

    As noted in Table 5.2, the PGS assay components for the SLCO1B1 Drug Transport report such as the custom beadchip, reagents, and instrumentation are the same as the predicate devices. No new reagents were needed and the beadchip was unchanged to test for the c.521T>C (rs4149056) variant. The probes to detect c.521T>C (rs4149056) already existed on the beadchip.

    The novel components in this Traditional 510(k) submission are to provide interpretive drug information to one specific medication (simvastatin), and to remove the limitation language requiring confirmatory testing in the 23andMe pharmacogenetics report for SLCO1B1. Pharmacogenetic reports for other genes authorized in DEN180028 will not be modified to remove the confirmatory testing limitation, include interpretive drug information, or add a prescription indication.

    Engineering drawings, schematics, etc. of the 23andMe Personal Genome Service (PGS) Pharmacogenetic Reports are not applicable to this device.

    AI/ML Overview

    The provided document, a 510(k) Summary for the 23andMe Personal Genome Service (PGS) Pharmacogenetic Reports (K221885), focuses on the modifications to the SLCO1B1 pharmacogenetic report. The primary changes are the addition of interpretive drug information for simvastatin and the removal of the confirmatory testing requirement for SLCO1B1. The summary details analytical performance studies conducted to support these changes.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of acceptance criteria and the reported device performance

    The document specifies acceptance criteria for analytical performance studies. These are primarily related to accuracy (method comparison) and precision (reproducibility) for the c.521T>C (rs4149056) variant in the SLCO1B1 gene.

    Study TypeAcceptance CriteriaReported Device Performance
    Method Comparison (Accuracy)>99% overall agreement (PPA and NPA both >99%) compared to bidirectional Sanger sequencing.100% overall agreement for all genotypes of the c.521T>C (rs4149056) variant. PPA and NPA both >99% (specifically reported as 100% concordance).
    Precision (Reproducibility)Minimum of 99% correct genotype calls at each of two laboratory sites.100% correct genotype calls for all samples across multiple days, operator teams, instruments, and reagent lots at 2 independent laboratory sites. The study also had >99% reproducibility and >99% repeatability.
    Minimum DNA Input (MDI)Passed acceptance criteria at a sample DNA concentration of 5 ng/µL.100% concordant test results and correct genotype calls for all samples and all reagent lots tested at sample DNA concentrations of 5, 15, and 50 ng/μL. Therefore, the study passed the acceptance criteria at a sample DNA concentration of 5 ng/μL.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: The document does not explicitly state a total numerical sample size for the method comparison and precision studies. It mentions that for the method comparison, samples were selected from 23andMe's database to enrich for prevalent variants in specific ethnicities, and allele and diplotype frequencies were used to inform the number of samples selected. For the precision study and MDI, "intended use (saliva) samples were selected from the 23andMe customer biobank based on their putative genotype" and "obtained for each of the c.521C>T genotype combinations." While specific numbers aren't given, the language suggests a controlled selection of samples representing relevant genotype combinations.
    • Data Provenance: The samples for performance testing were obtained from the "23andMe customer biobank." This suggests that the data is from real-world customer samples. The document does not specify the country of origin of these customers, but 23andMe is a US-based company, implying a significant US customer base. The data would be considered retrospective as it's from pre-existing biobank samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • The ground truth for the analytical performance studies (method comparison, precision, MDI) was established using bidirectional Sanger sequencing, which is a widely accepted laboratory gold standard for genetic variant confirmation.
    • The document does not mention the use of human experts (like radiologists or genetic counselors) to establish the ground truth for these analytical performance tests, as the ground truth is a direct genetic sequencing result, not an interpretation of an image or a clinical diagnosis.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Since the ground truth for the analytical performance studies was established by bidirectional Sanger sequencing, there was no human adjudication method employed for these specific tests. The comparison was directly between the device's genotyping result and the sequencing result.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was performed or is relevant for this type of device. The 23andMe Personal Genome Service is a direct-to-consumer genetic testing product that provides raw genetic data and interpretive reports; it is not an AI-assisted diagnostic imaging tool or a system where human readers interpret data with or without AI assistance. The "human readers" in this context are the end-users (consumers) who read the reports, and their "improvement" is not measured in the traditional MRMC sense.
    • The document does mention "robust user comprehension testing, previously reviewed and authorized under DEN180028 and K193492," indicating that the readability and understanding of the reports by users were assessed. However, this is distinct from an MRMC study and the specific results of this comprehension testing (e.g., effect size) are not detailed in this 510(k) summary, only referenced as having been sufficient for prior clearances.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, standalone performance was evaluated. The method comparison, precision, and minimum DNA input studies directly assess the accuracy and reliability of the 23andMe BeadChip assay and its associated software (Genome Studio, Coregen) in determining genotypes, without direct human intervention or interpretation during the genotyping process itself. The reported performance (e.g., 100% agreement with Sanger sequencing) is the standalone performance of the genotyping system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The ground truth used for the analytical performance studies (method comparison, precision, MDI) was bidirectional Sanger sequencing, which is a highly accurate and widely accepted method for determining the true genetic sequence/variant status.

    8. The sample size for the training set

    • The document does not explicitly state the sample size for the training set for the device's algorithms. As the device involves a genotyping array and software for data analysis (Coregen), it's likely that a substantial amount of genetic data was used in the development and initial training/optimization of these systems. However, this 510(k) summary focuses on the validation of modifications to a previously cleared device rather than the initial development of the core genotyping platform.

    9. How the ground truth for the training set was established

    • The document does not detail how the ground truth for the training set (if any specific to training was used for this modification) was established. For a genotyping platform like this, the "training" (or more accurately, the development and verification) would typically involve using samples with known genetic variations confirmed by gold-standard methods like Sanger sequencing or whole-genome sequencing. The 510(k) focuses on the validation of the current device's performance against a gold standard (Sanger sequencing).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1