Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221860

    Validate with FDA (Live)

    Date Cleared
    2023-09-15

    (445 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PLPT LDV (Low Dead Volume) LC(Luer-Cone) Sterile Syringe / PLPT LDV (Low Dead Volume) LL(Luer-Lock) Sterile Syringe is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the surface of the skin.

    Device Description

    PLPT LDV LC Sterile Syringe / PLPT LDV LL Sterile Syringe is a device intended to inject fluids into, or withdraw fluids from, parts of the body the surface of the skin. The device consists of a calibrated barrel (cylinder) with plunger, and metal tube that is sharpened at one end and at the other end joined to a female connector (hub). This device is intended for various medical applications and is not dedication administration. At the distal end of the barrel has a Luer-Cone or Luer-lock connector for the attachment to a hypodermic needle or an administration set. Attached needle has covered by the protective cap which is intended to provide physical protection to the needle tube. Attached needle may have safety guard that protect the user after use. This product is packed by sterile paper and sterilized by EO gas, and single-use device.

    AI/ML Overview

    This document describes the regulatory approval for a medical device, specifically a Low Dead Volume (LDV) Sterile Syringe (PLPT LDV LC/LL Sterile Syringe). The provided text is an FDA 510(k) Premarket Notification, which focuses on demonstrating substantial equivalence to a predicate device, rather than proving novel effectiveness or clinical superiority through extensive comparative clinical trials.

    Therefore, the information about acceptance criteria and study proving the device meets those criteria will be focused on bench testing and biocompatibility assessments demonstrating equivalence to established standards and predicate devices, rather than complex multi-reader, multi-case studies typically seen for AI/ML devices.

    Here's the breakdown based on the provided text:

    Device: PLPT LDV (Low Dead Volume) LC (Luer-Cone) Sterile Syringe and PLPT LDV (Low Dead Volume) LL (Luer-Lock) Sterile Syringe.

    Indications for Use: Intended for use by healthcare professionals for general purpose aspiration of fluid from vials, ampoules, and liquid injection below the surface of the skin.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily based on established international standards for medical devices and comparison to predicate devices. The device's performance is demonstrated through adherence to these standards and the fulfillment of specific regulatory requirements.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Standard / Test Method / Limit)Reported Device Performance (Compliance / Result)
    Mechanical PerformanceISO 7886-1:2017 Sterile hypodermic syringes for single use (Syringe performance)Complies with ISO 7886-1:2017
    ISO 7864:2016 Sterile hypodermic needles for single use (Needle performance)Complies with ISO 7864:2016
    ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devicesComplies with ISO 9626:2016
    ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications (Connector performance)Complies with ISO 80369-7:2021
    BiocompatibilityISO 10993-1:2018 (Body contact: External communication device – Blood path indirect; Duration: Limited (< 24 hours))All evaluation acceptance criteria were met.
    Cytotoxicity (ISO 10993-5)Satisfied requirements
    Hemolysis (ISO 10993-4)Satisfied requirements
    Intracutaneous Reactivity (ISO 10993-10)Satisfied requirements
    Skin Sensitization (ISO 10993-10, GPMT)Satisfied requirements
    Acute Systemic Toxicity (ISO 10993-11)Satisfied requirements
    Pyrogenicity (ISO 10993-11)Satisfied requirements
    LAL Test (USP39 <85>, Bacterial Endotoxins Test)Satisfied requirements
    Particulate Matter Injection (USP <788>)Satisfied requirements
    SterilityISO 11135:2014 Ethylene Oxide (EO) Sterilization Validation (Parameters: EO 30%, CO2 70%; Temp: 50 ±7°C; Exposure: 5 hours)Validated and deemed acceptable.
    Sterility Test (ISO 11737-2)Complied
    EO Residual Test (ISO 10993-7:2008)Met test requirements
    Sterility Assurance Level (SAL)10⁻⁶
    Endotoxin Limit20 EU per device
    Low Dead VolumeSyringe Dead Space: ≤ 0.023 mL≤ 0.023 mL (Same as predicate)
    Needle Dead Space: < 0.0054 mL< 0.0054 mL (Same as predicate)
    Sharps Injury ProtectionISO 23908:2011 Sharps injury protectionComplies with ISO 23908:2011
    Shelf-LifeISO 11607-1:2006 and ISO 11607-2:2006 (for 3 years)Validated for 3 years
    Labeling21 CFR part 801Complied

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the numerical sample sizes for each specific test (e.g., how many syringes were tested for each mechanical property, or how many biological samples were used for biocompatibility tests). However, for medical devices cleared via 510(k), it is expected that sufficient samples were used to meet the statistical requirements of the referenced ISO standards for design verification and validation.

    • Data Provenance: The studies were bench tests and biocompatibility evaluations. The data would be prospective, generated specifically for this 510(k) submission, likely conducted at the manufacturer's facilities or accredited testing labs. The country of origin for the data generation is not explicitly stated, but the manufacturer is based in Republic of Korea.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This type of information is generally not applicable for a 510(k) submission for a physical medical device like a syringe. The "ground truth" for these tests comes from adherence to established international engineering and biological standards (ISO, USP). There isn't a subjective "ground truth" established by human experts in the way there would be for an AI-based diagnostic tool. The "experts" are the engineers and scientists conducting the tests according to the specified protocols.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing consensus "ground truth" in human-in-the-loop diagnostic studies, typically involving image interpretation. For physical device performance, the results are quantitative measurements against predefined pass/fail criteria established by engineering standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a physical syringe, not an AI-assisted diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical syringe, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device's performance is based on objective measurements and results from standardized bench tests and validated biological assays as defined by:

    • International Organization for Standardization (ISO) standards (e.g., ISO 7886-1, ISO 80369-7, ISO 10993 series, ISO 11135, ISO 11607, ISO 23908).
    • United States Pharmacopeia (USP) guidelines (e.g., USP <85>, USP <788>).

    These standards define the acceptable performance characteristics for syringes and needles.

    8. The sample size for the training set:

    Not applicable. This device is a physical product and does not involve AI/ML requiring a "training set" in the computational sense.

    9. How the ground truth for the training set was established:

    Not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1