    Intended Use

    The ADVIA Centaur® Anti-Müllerian Hormone (AMH) assay is for in vitro diagnostic use in the quantitative determination of anti-Müllerian hormone (AMH) in human serum and plasma (lithium heparin) using the ADVIA Centaur® XP system.

    The measurement of AMH is used as an aid in the assessment of the ovarian reserve in women presenting to fertility clinics. This assay is intended to distinguish between women with AFC (antral follicle count) values > 15 (high ovarian reserve) and women with AFC values ≤ 15 (normal or diminished ovarian reserve).

    This assay is intended to be used in conjunction with other clinical and laboratory findings, such as AFC, before starting fertility therapy. This assay is not intended to be used for monitoring women undergoing controlled ovarian stimulation in an Assisted Reproduction Technology program.

    Device Description

    The ADVIA Centaur AMH assay is a sandwich immunoassay using direct acridinium ester-based chemiluminometric technology. Two monoclonal anti-AMH antibodies are employed in the assay. The first antibody in the Lite Reagent is a mouse monoclonal anti-AMH antibody labeled with acridinium ester. The second antibody is a biotinylated mouse monoclonal anti-AMH antibody coupled to streptavidin-coated magnetic particles in the Solid Phase.

    A direct relationship exists between the amount of AMH present in the patient sample and the amount of relative light units detected by the system. Dose concentration results (ng/mL) are calculated based on a 2-point calibration from a pre-defined master curve.

    Materials include:
    ADVIA Centaur AMH ReadyPack® primary reagent pack: Solid Phase (Streptavidin-coated paramagnetic microparticles with biotinylated mouse monoclonal anti-human AMH antibody in buffer; sodium azide (< 0.1%); blocker (bovine); surfactant; preservatives)
    ADVIA Centaur AMH ReadyPack® ancillary reagent pack: Ancillary Reagent (Mouse monoclonal anti-human AMH antibody labeled with acridinium ester in buffer (~0.6 µg/mL); sodium azide (<0.1%); blocker (bovine, murine); stabilizers; surfactant; preservatives)
    AMH CAL: After reconstitution, low and high levels of AMH antigen (bovine) in defibrinated human plasma; sodium azide (< 0.1%); preservatives

    AI/ML Overview

    Here's the detailed breakdown of the acceptance criteria and study information for the ADVIA Centaur® Anti-Müllerian Hormone (AMH) assay, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Design Goal)Reported Device Performance
    Detection Capability
    Limit of Blank (LoB)(Implicitly compared to predicate LoB of ≤ 0.01 ng/mL)0.010 ng/mL (0.071 pmol/L)
    Limit of Detection (LoD)(Implicitly compared to predicate LoD of ≤ 0.02 ng/mL)0.020 ng/mL (0.143 pmol/L)
    Limit of Quantitation (LoQ)(Implicitly compared to predicate LoQ of ≤ 0.08 ng/mL)0.043 ng/mL (0.307 pmol/L)
    Precision (Total CV)≤ 10% CV for concentration ≥ 0.100 ng/mLRanges from 2.5% to 4.4% for samples at various AMH concentrations (0.112 ng/mL to 16.4 ng/mL). For controls, ranges from 2.9% to 3.3% (0.955 ng/mL to 14.1 ng/mL). All reported total CVs are ≤ 4.4% at or above 0.112 ng/mL, meeting the criterion.
    Reproducibility (Total CV)(Implicitly compared to predicate Total CV of ≤ 10% for concentration ≥ 0.16 ng/mL)Ranges from 2.1% to 3.1% for samples at various AMH concentrations (0.199 ng/mL to 17.0 ng/mL). For controls, ranges from 2.6% to 2.9% (1.01 ng/mL to 14.4 ng/mL). All reported reproducibility CVs are ≤ 3.1% at or above 0.199 ng/mL, meeting the likely implied criterion.
    LinearityLinear for the measuring interval of 0.043-24.0 ng/mLLinear for the measuring interval of 0.043-24.0 ng/mL (0.307-171 pmol/L).
    Assay ComparisonCorrelation coefficient ≥ 0.950, slope of 1.00 ± 0.10, intercept of ± 0.035 ng/mL (vs. commercial AMH assay)Serum: Correlation coefficient (r) = 0.994 Regression Equation: y = 1.04x - 0.032 ng/mL (Slope: 1.04, Intercept: -0.032 ng/mL). This meets the criteria for correlation, slope (within 1.00 ± 0.10), and intercept (within ± 0.035 ng/mL).
    Specimen EquivalenceCorrelation coefficient ≥ 0.950, slope of 0.90-1.10, intercept of ± 0.035 ng/mL (vs. serum)Gel-barrier tube (serum) vs. Serum: Correlation coefficient (r) = 0.997 Regression Equation: y = 1.00x + 0.003 ng/mL (Slope: 1.00, Intercept: +0.003 ng/mL). Meets criteria. Plasma, lithium heparin vs. Serum: Correlation coefficient (r) = 0.997 Regression Equation: y = 1.08x - 0.004 ng/mL (Slope: 1.08, Intercept: -0.004 ng/mL). Meets criteria.
    Interferences (HIL)Bias due to substances not to exceed 10% at specified AMH concentrationsHemoglobin: No interference (1000 mg/dL). Bilirubin, conjugated: No interference (66.0 mg/dL). Bilirubin, unconjugated: No interference up to 39.0 mg/dL; however, >10% bias observed at ≥ 40 mg/dL (10.6% bias at 6.79 ng/mL AMH, 11.4% bias at 0.936 ng/mL AMH). Lipemia (Intralipid): No interference (2000 mg/dL).
    Interferences (Other Substances)Bias due to substances not to exceed 10% at specified AMH concentrationsAcetaminophen, Acetylcysteine, Acetylsalicylic Acid, Ampicillin sodium, L-Ascorbic acid, Biotin, Cefoxitin sodium salt, Cholesterol, Cyclosporine, Doxycycline hyclate, Folic acid, Gonapeptyl, Heparin, Human IgA, Human IgG, Human IgM, Ibuprofen, Levodopa, Levothyroxine, Metformin hydrochloride, Methyldopa, Metronidazole, Phenylbutazone, Rheumatoid Factor, Rifampicin, Theophylline, Total Protein, Uric acid: All showed no interference (bias ≤ 10%) at tested concentrations.
    Cross-ReactivityBias does not exceed 10%Activin A, Activin B, Activin AB, Inhibin A, Inhibin B, TGF b-1: ≤ 0.1% cross-reactivity. Follicle stimulating hormone (FSH) at 500 mIU/mL: Not Detectable, 0.2% bias. Luteinizing hormone (LH) at 500 mIU/mL: Not Detectable, 2.9% bias. All considered insignificant.
    Stability (On-board Reagents)Reagents stable for 70 daysDetermined to be 70 days.
    Stability (Calibrators)Calibrators stable at 2-8°C and ≤ -20°C for 90 days after reconstitutionDetermined to be stable at 2-8°C and ≤ -20°C for 90 days after reconstitution.
    High Dose Hook(No explicit criterion given, but predicate states no hook effect up to 1000 ng/mL)No hook effect observed up to 1151 ng/mL (8218 pmo/L). This exceeds the predicate.
    Clinical Performance (Overall)(Aid in distinguishing AFC > 15 vs ≤ 15 in fertility clinics)Sensitivity: 90.5% (256/283) (95% CI: 86.47, 93.36) Specificity: 52.0% (130/250) (95% CI: 45.82, 58.12) PPV: 68.1% (256/376) (95% CI: 63.21, 72.59) NPV: 82.8% (130/157) (95% CI: 76.13, 87.90)
    Clinical Performance (Age < 35)(Aid in distinguishing AFC > 15 vs ≤ 15 in fertility clinics for this age group)Prevalence (AFC > 15): 67.4% PPV: 73.6% (67.47, 78.88) NPV: 65.1% (50.17, 77.58)
    Clinical Performance (Age ≥ 35)(Aid in distinguishing AFC > 15 vs ≤ 15 in fertility clinics for this age group)Prevalence (AFC > 15): 38.4% PPV: 59.7% (51.71, 67.27) NPV: 89.5% (82.50, 93.88)

    2. Sample Size Used for the Test Set and the Data Provenance

    • Detection Capability (LoB, LoD, LoQ): Not explicitly stated, but determined as described in CLSI protocol EP17-A2. These are typically derived from analytical studies involving numerous replicates of blank, low-concentration, and relevant samples.
    • Precision and Reproducibility:
      • Precision: 480 measurements per sample/control (replicates of 2, 2 runs/day, 20-day protocol). This was across 2 instruments and 3 reagent lots.
      • Reproducibility: 90 measurements per sample/control (triplicate, 2 runs/day, 5 days) across 3 sites and 1 reagent lot.
    • Assay Comparison: 120 samples (serum) vs. a commercial AMH assay.
    • Specimen Equivalence: 88 samples for Gel-barrier tube (serum) vs. Serum, and 88 samples for Plasma (lithium heparin) vs. Serum.
    • Interferences: The number of samples tested for each substance is not specified, but the testing was performed in accordance with CLSI Document EP07-ed3 and EP37-ed1.
    • Cross-Reactivity: Number of samples not specified, performed in accordance with CLSI Document EP07-ed3.
    • Stability: Not sample-based but rather experimental conditions and time points.
    • Expected Values (Reference Intervals):
      • Females (18–25 years): 209 samples
      • Females (26–30 years): 122 samples
      • Females (31–35 years): 123 samples
      • Females (36–40 years): 126 samples
      • Females (41–45 years): 152 samples
      • Females (46–50 years): 121 samples
      • Females (51 years and older): 139 samples
    • Clinical Sensitivity and Specificity: 533 women.
    • Data Provenance:
      • Clinical Study: Prospectively collected from women presenting to fertility clinics for evaluation.
      • Country of Origin: 11 sites across the United States.
      • Expected Values: Samples were collected retrospectively or prospectively from "apparently healthy subjects," but the exact nature (retrospective/prospective) and location of collection is not explicitly detailed beyond "apparently healthy subjects". Given the clinical study provenance, it's likely linked, but not explicitly stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    For the clinical performance study (Clinical Sensitivity and Specificity), the ground truth for ovarian reserve was established by Antral Follicle Count (AFC) values determined by transvaginal ultrasound.

    • The document does not specify the number of experts (e.g., sonographers, radiologists) who performed or interpreted these ultrasounds.
    • It also does not specify the individual qualifications (e.g., years of experience, board certification) of these experts. It only states that the AFC result was determined by transvaginal ultrasound.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (such as 2+1, 3+1) for establishing the ground truth (AFC values). The AFC results were simply stated as being "determined by transvaginal ultrasound." This suggests that the individual AFC measurements were taken as the singular truth, without a multi-reader review or adjudication process outlined to resolve discrepancies, if any.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, an MRMC comparative effectiveness study involving human readers and AI assistance was not conducted. This device is an in vitro diagnostic (IVD) assay that quantitatively measures a biomarker (AMH) in serum/plasma. It does not involve interpretation of medical images or other data by human readers, and thus, AI assistance in the context of human reading is not applicable here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies reported are essentially "standalone" performance evaluations of the ADVIA Centaur® Anti-Müllerian Hormone (AMH) assay itself. The clinical performance study evaluates the assay's ability to distinguish between high and normal/diminished ovarian reserve based on AMH measurements, against the ground truth of AFC. This is the performance of the device (the assay) as an algorithm or test method, independent of subsequent human interpretation enhancements.

    7. The Type of Ground Truth Used

    • Analytical Studies (Detection Capability, Precision, Linearity, Interference, Cross-reactivity, Stability, Hook Effect): The ground truth for these studies is typically derived from highly characterized reference materials, spiked samples with known concentrations, or established analytical methods.
    • Expected Values (Reference Intervals): Established by collecting samples from "apparently healthy subjects" and calculating statistical percentiles (90th and 95th reference intervals). The ground truth here is the statistical distribution of AMH levels in a healthy population defined by age.
    • Clinical Sensitivity and Specificity: The ground truth for ovarian reserve was Antral Follicle Count (AFC) values, as measured by transvaginal ultrasound. This is a clinical measure widely accepted in fertility assessment.

    8. The Sample Size for the Training Set

    The document does not specify a separate "training set" in the context of a machine learning or AI algorithm development. This is an in vitro diagnostic assay, where performance is typically established through analytical validation and clinical correlation studies, not through AI model training. The data used for most performance characteristics are considered validation data.

    • The "Expected Values" data set (totaling 209+122+123+126+152+121+139 = 992 samples) could be seen as reference data used to establish norms, but not a "training set" for an algorithm in the AI sense.
    • The "Clinical Sensitivity and Specificity" study of 533 women served as a clinical validation dataset.

    9. How the Ground Truth for the Training Set Was Established

    As no "training set" in the AI sense is explicitly mentioned for algorithm development, there's no described method for establishing ground truth for such a set. For the validation data described:

    • Clinical Study Ground Truth: The ground truth was Antral Follicle Count (AFC) values, determined by transvaginal ultrasound by unspecified qualified personnel at 11 sites across the US.
    • Expected Values Ground Truth: These are based on AMH measurements from "apparently healthy subjects," where their health status (absence of relevant pathologies) constitutes the ground truth for establishing normal ranges.