    Device Name
    Hexanium ACIF
    Date Cleared
    2022-11-14

    (166 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hexanium® ACIF (Anterior Cervical Interbody Fusion) system is an intervertebral body fusion device indicated for use with autogenous bone graft in skeletally mature patients with Degenerative Disc Disease (DDD) at one level from C3-T1. DDD is defined as discogenic neck pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received at least 6 weeks of non-operative treatment prior to treatment with Hexanium® ACIF system. This device has to be filled with autogenous bone graft material. This device is implanted via an anterior approach.

    Device Description

    The Hexanium® ACIF is a titanium alloy (Ti6Al4V ELI) interbody cage manufactured via an Additive Manufacturing method. The honeycomb structure allows for bone through-growth through the structure of the device as well as providing lateral and vertical bone graft windows in the body of the cage. Hexanium® ACIF is available in 2 sagittal profiles (lordotic 6º and convex 6°), heights of 5-12 mm, and 3 footprints (15x12mm, 17x14mm, and 19x15mm). The Hexanium® ACIF screws are self-drilling and available in lengths of 10, 12, 14, and 16mm and diameters of 3.50 and 3.80mm. Hexanium® ACIF is provided sterile. The Hexanium® ACIF system includes a set of reusable surgical instruments. The purpose of this Special 510(k) application is to propose minor modifications to the currently cleared surgical instruments. There are no modifications being proposed for the implantable components of the system

    AI/ML Overview

    This is not an AI/ML medical device. This is a medical device for intervertebral body fusion. Thus the acceptance criteria for AI devices do not apply here.