510(k) Data Aggregation

    K Number
    K221352

    Validate with FDA (Live)

    Date Cleared
    2023-03-13

    (307 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surgical Face Mask is intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

    Device Description

    The Surgical Face Mask is 3-layer, and it is two model is HSD-TYP2R (Ear loops) and HSD-TYP2RT (Tie-On). Flat-folded masks constructed of nonwoven polypropylene materials. The mask is provided with ear loops (spandex and polyester) or tie tapes (spun-bond polypropylene). A malleable nose clip is placed in the layers of facemask for comfort and individualized fit. The surgical face mask color is blue. Product is Level 3 based on ASTM F2100-19. The surgical face mask is single-use, disposable devices, provided non-sterile.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from Wuhan Huashida Protective Products Co., Ltd. for their Surgical Face Mask. It is a submission to the FDA demonstrating that their device is substantially equivalent to a legally marketed predicate device.

    The premise of a 510(k) submission is to demonstrate "substantial equivalence' to a legally marketed device, not necessarily to prove the device's independent safety and effectiveness through extensive clinical trials as would be required for a PMA (Premarket Approval) application. For a device like a surgical face mask, the FDA generally relies on performance testing against recognized standards rather than clinical studies, especially if the device is similar in design and materials to existing cleared devices.

    Therefore, the information provided below is based on the provided document and tailored to how a 510(k) submission addresses "acceptance criteria" through non-clinical testing.

    Here's an analysis of the provided text in relation to your request about acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document provides a table of acceptance criteria and results from non-clinical performance and biocompatibility testing.

    ItemsPurposeAcceptance CriteriaResult
    For Surgical Face Mask Performance
    Fluid Resistance PerformanceThe purpose of the performance testing is to demonstrate the functionality of the subject device29 out of 32 pass at 160 mmHgPass
    Particulate Filtration Efficiency(Implied: To demonstrate the mask's ability to filter particulate material)$\geq$ 98%Pass
    Bacterial Filtration Efficiency(Implied: To demonstrate the mask's ability to filter microorganisms)$\geq$ 98%Pass
    Differential Pressure(Implied: To measure the breathability of the mask)<6.0 mmH2O/cm²Pass
    Flammability(Implied: To assess the flammability risk of the mask material)Class 1Pass
    For Surgical Face Mask Biocompatibility
    CytotoxicityThe purpose of the testing is to demonstrate safety of the subject deviceNon-CytotoxicUnder the conditions of the study, the device is non-cytotoxic.
    Irritation(Implied: To assess potential skin irritation from mask materials)Non-IrritatingUnder the conditions of the study, the device is non-irritating.
    Sensitization(Implied: To assess potential allergic reactions or sensitization from mask materials)Non-SensitizingUnder the conditions of the study, the device is non-sensitizing.

    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size:
      • For "Fluid Resistance Performance," the sample size is explicitly stated as 32 (29 out of 32 pass).
      • For other performance and biocompatibility tests, specific sample sizes are not explicitly stated in this summary document. They would typically be defined within the testing protocols for each standard (e.g., ASTM F2299, ASTM F2101, ISO 10993-5, etc.).
    • Data Provenance: The tests were conducted by the manufacturer, Wuhan Huashida Protective Products Co., Ltd., in China. The data would be considered prospective as it involves new testing performed specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information is not applicable to this submission. This document addresses a physical medical device (surgical face mask) evaluated through non-clinical performance testing against recognized standards. The "ground truth" for such devices is established by the specifications in the standards themselves (e.g., a BFE of >=98% is the target). There are no human "experts" establishing a subjective ground truth like in a diagnostic AI study. The results are quantitative measurements.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable for non-clinical performance and biocompatibility testing. Adjudication methods like 2+1 or 3+1 are used in studies where human readers (e.g., radiologists) are involved in interpreting medical images or data, and their disagreements need to be resolved to establish a "ground truth" or reference standard. This document pertains to physical device testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. An MRMC study is relevant for evaluating the performance of AI-assisted diagnostic tools where human interpretation plays a role. This submission pertains to a surgical face mask, which is a physical barrier device, not an AI diagnostic tool. The document explicitly states: "No clinical study is included in this submission."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This question refers to the performance of an AI algorithm in isolation. The product is a surgical face mask, not an algorithm. The testing performed is standalone in the sense that it's the mask's performance being measured directly against established standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is based on pre-defined, quantitative performance specifications established by relevant industry standards and regulatory guidance. For example:

    • Fluid Resistance: Resistance to penetration by synthetic blood at a specified pressure (ASTM F1862).
    • Filtration Efficiency (Particulate & Bacterial): Percentage of particles/bacteria filtered out (ASTM F2299, ASTM F2101).
    • Differential Pressure: How easily air can pass through (EN 14683).
    • Flammability: A classification based on burn rate (16 CFR 1610).
    • Biocompatibility: Absence of cytotoxic, irritating, or sensitizing effects as per ISO 10993 standards.

    These standards define the 'truth' or expected performance for a surgical mask of a certain level.

    8. The sample size for the training set

    This is not applicable. This device is a manufactured medical product, not an AI algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    This is not applicable, as there is no "training set" for this type of device submission.