510(k) Data Aggregation

    K Number
    K221330

    Validate with FDA (Live)

    Device Name
    BriefCase
    Date Cleared
    2022-11-18

    (196 days)

    Product Code
    Regulation Number
    892.2080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase is a radiological computer-aided triage and notification software indicated for use in the analysis of frontal chest X-ray (CXR) images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of vertically malpositioned endotracheal tube (ETT) in relation to the carina.

    Findings are flagged when the ETT distal tip is assessed as being more than 5 cm above the carina, less than 2 cm above the carina, or when it is below the carina (i.e., in the right or left mainstem bronchus).

    The device assesses solely the vertical position of the ETT distal tip relative to the carina, does not factor patient positioning, and cannot detect esophageal intubation. The device does not provide results when the carina is not wellvisualized on the x-ray image. The device does not discriminate types of ETTs, as such a properly positioned double lumen ETT may trigger a false prioritization alert.

    BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of BriefCase are intended to be used in conjunction with other patient information and based on the user's professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase is a radiological computer-assisted triage and notification software device. The software system is based on an algorithm programmed component and consists of a standard off-the-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL, DICOM module and the BriefCase Image Processing Application. The device consists of the following three modules: (1) Aidoc Hospital Server (AHS/Orchestrator) for image acquisition: (2) Aidoc Cloud Server (ACS) for image processing; and (3) Aidoc Desktop Application for workflow integration.

    DICOM images are received, saved, filtered and de-identified before processing. Filtration matches metadata fields with keywords. Series are processed chronologically by running the algorithms on each series to detect suspected cases. The software then flags suspect cases by sending notifications to the desktop application, thereby facilitating triage and prioritization by the user. As the BriefCase software platform harbors several triage algorithms, the user may opt to filter out notifications by pathology, e.g., a chest radiologist may choose to filter out alerts on VO cases, and a neuro-radiologist would opt to divert vertically malpositioned ETT in relation to the carina alerts. Where several medical centers are linked to a shared PACS, a user may read cases for a certain center but not for another, and thus may opt to filter out alerts by center. Activating the filter does not impact the order in which notifications are presented in the Aidoc desktop application.

    The desktop application feed displays all incoming suspect cases, each notified case in a line. Hovering over a line in the feed pops up a compressed, low-quality, grayscale, unannotated image that is captioned "not for diagnostic use" and is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the BriefCase device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionPerformance GoalReported Performance (95% CI)
    Sensitivity80%86.81% (81.02% - 91.36%)
    Specificity80%92.02% (89.20% - 93.87%)
    Time-to-Notification(Implicitly, faster than standard of care)4.93 minutes (4.64 - 5.22)

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: 921 cases
    • Data Provenance: Retrospective, multicenter study with cases from 5 US-based clinical sites. The cases were distinct in time or center from the training data.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: 3 senior board-certified radiologists.
    • Qualifications of Experts: Senior board-certified radiologists. (Specific experience in years is not provided).

    4. Adjudication Method for the Test Set

    • Adjudication Method: 2 out of 3 majority voting of senior board-certified radiologists.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No. The study describes a standalone algorithm performance evaluation and a comparison of the algorithm's notification time to the standard of care time-to-exam-open, not a direct human reader improvement study.
    • Effect Size of Human Readers Improvement with AI vs. without AI: Not applicable, as an MRMC comparative effectiveness study was not performed. The study shows the device provided notifications significantly faster than standard of care exam opening.

    6. Standalone (Algorithm Only) Performance

    • Was standalone performance done? Yes. The primary endpoint evaluation of sensitivity and specificity was performed for the BriefCase software in identifying malpositioned ETTs compared to the expert-established ground truth.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus, determined by 2 out of 3 majority voting of senior board-certified radiologists.

    8. Sample Size for the Training Set

    • The document states, "The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm," but does not explicitly provide the sample size of the training set.

    9. How the Ground Truth for the Training Set Was Established

    • The document mentions "training of the algorithm on vertically malpositioned ETT in relation to the images" but does not specify how the ground truth for the training set was established.