Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221311

    Validate with FDA (Live)

    Date Cleared
    2023-06-21

    (412 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Antibacterial bandage is to be applied topically to the skin for management of minor cuts, minor scrapes and minor burns.

    Device Description

    The subject device. Antibacterial bandage, is to be applied topically to the skin for the management of minor cuts, minor scrapes and minor burns. The subject devices are available sterile in three models. All of devices contain three layers: (1) adhesive backing layer (fabric or plastic); (2) exactly same antibacterial non-stick absorbent pad layer (composed of Polyester fiber, viscose fiber, PE mesh, Benzalkonium chloride); (3) release liner. Each model is available in different sizes with identical materials, chemicals and physical properties, therefore, size of the device will not affect the function and performance of products. All models of subject device, Antibacterial bandage, are for single use only and sterilized by Ethylene Oxide using conditions validated following ISO 11135:2014.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for an "Antibacterial Bandage" (K221311) and compares it to a predicate device (K113583) and a reference device (K200821). The submission aims to demonstrate substantial equivalence, not to prove that the device meets specific acceptance criteria through a clinical study. Therefore, a direct answer to the request regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of device performance metrics often expected for AI or diagnostic devices (e.g., accuracy, sensitivity, specificity, MRMC studies) cannot be fully provided from the given document.

    However, I can extract the non-clinical test conclusions which serve as the basis for demonstrating substantial equivalence based on performance and biocompatibility. These tests implicitly define "acceptance criteria" through their methodologies and the successful comparison to the predicate device.

    Here's a breakdown of the information that can be extracted, framed as closely as possible to the request:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly list numerical "acceptance criteria" with corresponding "reported device performance" in a typical clinical study outcome format like sensitivity or specificity. Instead, the "acceptance criteria" are implied by the successful completion of specified performance and biocompatibility tests, demonstrating equivalence to the predicate device or compliance with standards.

    For Antibacterial Effectiveness, a specific quantitative measure is mentioned for both the subject and predicate devices:

    Acceptance Criteria (Implied by Predicate/Standard)Reported Device Performance (Subject Device K221311)
    ≥4 Log reduction (Antibacterial Effectiveness)≥4 Log reduction

    For other performance tests and biocompatibility, the reported "performance" is that the device "met the requirement" or "complied with the corresponding standards," or "was not different" from the control.

    Here's how the other "acceptance criteria" and "reported performance" can be inferred from the non-clinical tests:

    Test/CategoryImplicit Acceptance CriteriaReported Device Performance
    Performance Tests
    Liquid absorbency (EN 13726-1:2002)Met requirements of standard/equivalent to predicateMet the requirement
    Moisture vapor transmission rate (EN 13726-2:2002)Met requirements of standard/equivalent to predicateMet the requirement
    Peel strength (ASTMD3330/D3330M)Met requirements of standard/equivalent to predicateMet the requirement
    Sterile barrier package testing (ASTM F88/F88M-15 & ASTM F1929-15)Package maintains integrityMaintained its integrity
    Biocompatibility Tests
    Cytotoxicity (ISO 10993-5:2009)Complied with standardComplied with standard
    Skin sensitization (ISO 10993-10:2010)Complied with standardComplied with standard
    Systemic toxicity (ISO 10993-11:2017)Complied with standardComplied with standard
    Material mediated pyrogenicity (USP 151:2017 & USP-NF:2021 <85>)Complied with standardComplied with standard
    Ethylene Oxide Sterilization Residuals (ISO 10993-7:2008)Complied with standardComplied with standard
    Animal Study (Wound Healing)No difference from predicate in wound healing and local toleranceNo difference observed between subject device and predicate device

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Non-Clinical Performance Tests & Biocompatibility Tests: The document does not specify sample sizes (e.g., number of bandages tested for peel strength) or the data provenance for these lab-based tests. These are typically performed in a controlled laboratory setting.
    • Animal Study: A "porcine wound healing study" was carried out, indicating an animal model. The sample size for this study is not specified, nor is the country of origin of the animal data. It would be considered prospective for the animal subjects involved.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable to the provided document. The device is an antibacterial bandage, not an AI or diagnostic device that requires expert review for ground truth establishment for a diagnostic test set. The ground truth for the non-clinical and animal studies is based on objective laboratory measurements and biological observations, not expert consensus on interpretations of images or signals.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods like 2+1 are typically used for establishing ground truth in clinical studies involving human interpretation (e.g., radiologists reviewing images). The studies described here are non-clinical and animal studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. This type of study is relevant for AI-powered diagnostic devices, which is not the case for an antibacterial bandage.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Non-Clinical Performance Tests: Ground truth is established by the specified test methods and standards (e.g., EN 13726-1:2002, ASTM D3330/D3330M). The "truth" is the measured physical or chemical property of the bandage.
    • Biocompatibility Tests: Ground truth is defined by the biological response observed according to the specific ISO standards (e.g., presence/absence of cytotoxicity, sensitization).
    • Animal Study: Ground truth is based on observed wound healing performance characteristics and local tolerance in the porcine model.

    8. The sample size for the training set

    This question is not applicable. The device (antibacterial bandage) does not involve machine learning or AI, and therefore does not have a "training set" in that context.

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1