Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221248

    Validate with FDA (Live)

    Device Name
    Rapid LVO
    Manufacturer
    Date Cleared
    2022-05-31

    (29 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rapid LVO is a radiological computer aided triage and notification software indicated for use in the analysis of CTA head images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive ICA or MCA-M1 Large Vessel Occlusion (LVO) findings in head CTA images.

    Rapid LVO uses a software algorithm to analyze images and highlight cases with suspected LVO on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO findings. Notifications include compressed preview images. These are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of Rapid LVO are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage /prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    Rapid LVO is a radiological computer-assisted triage and notification software device. The Rapid LVO module is a contrast enhanced CTA module which operates within the integrated Rapid Platform to provide triage and notification of suspected ICA and MCA-M1 Large Vessel Occlusion (LVO) based on the following definitions:

    ICA Occlusion: A high-grade stenosis or occlusion of the intracranial portion of the ICA.

    MCA-M1 Occlusion: A high-grade stenosis or occlusion of the horizontal segment of the MCA-M1, defined as the segment which extends from the ICA terminus until the vessel has turned upward into the Sylvian fissure. This includes post-bifurcation M1 segments in some patients.

    The LVO module uses traditional programming algorithms. The output of the module is a priority notification to clinicians indicating the suspicion of LVO based on positive findings. The Rapid LVO module uses the basic services supplied by the Rapid Platform including DICOM processing, job management, imaging module execution and imaging output including the notification and compressed image.

    AI/ML Overview

    The Rapid LVO device, a radiological computer-aided triage and notification software for detecting Large Vessel Occlusions (LVO) in CTA head images, was evaluated against specific acceptance criteria.

    1. Table of Acceptance Criteria and Reported Device Performance
    Acceptance CriterionReported Device Performance
    Sensitivity (Se): Lower bound of 95% Confidence Interval (CI) ≥ 80%0.96 (95% CI: 0.91 - 0.97)
    Specificity (Sp): Lower bound of 95% Confidence Interval (CI) ≥ 80%0.98 (95% CI: 0.93 - 0.99)
    Time to Notification: ≤ 3.5 minutes3.18 minutes (95% CI: 3.11 - 3.25)

    Additionally, the following performance metrics were reported:

    • Positive Predictive Value (PPV): 0.98
    • Negative Predictive Value (NPV): 0.96
    • Receiver Operating Characteristic (ROC) AUC: 0.99
    1. Sample Size and Data Provenance for the Test Set

      • Sample Size: 217 scans (135 positive LVO cases, 82 negative LVO cases).
      • Data Provenance: The data was collected from 8 sites/studies, with locations in both the US and OUS (Outside the US). The document does not explicitly state if the data was retrospective or prospective, but clinical validation testing typically uses retrospective data for ground truth establishment.
    2. Number of Experts and Qualifications for Ground Truth

      • Number of Experts: Three expert neuroradiologists.
      • Qualifications: They are described as "expert neuroradiologists." Specific years of experience are not provided.
    3. Adjudication Method for the Test Set

      • The ground truth was established using a "2:3 concurrence" method. This implies that at least two out of the three expert neuroradiologists had to agree on the presence or absence of an LVO for a case to be assigned its ground truth label.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

      • No MRMC comparative effectiveness study was mentioned. The study focused on the standalone performance of the algorithm.
    5. Standalone Performance Study

      • Yes, a standalone performance study was conducted. The reported sensitivity, specificity, PPV, NPV, and ROC AUC are all measures of the algorithm's performance without human-in-the-loop assistance.
    6. Type of Ground Truth Used

      • The ground truth was established by "expert neuroradiologists" using a "2:3 concurrence" method. This indicates expert consensus was the basis for the ground truth.
    7. Sample Size for the Training Set

      • The document does not explicitly state the sample size used for the training set. It only details the validation set used for performance testing.
    8. How Ground Truth for the Training Set Was Established

      • The document does not provide details on how the ground truth for the training set was established. It only describes the ground truth establishment for the validation/test set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1