510(k) Data Aggregation

    K Number
    K221157

    Validate with FDA (Live)

    Device Name
    JR Medic
    Date Cleared
    2022-08-30

    (131 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JR Medic Blue Latex Examination Powder Free Gloves Tested for use with Chemotherapy drugs are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

    The tested chemotherapy drugs and their breakthrough detection times are as follows:

    Tested Chemotherapy Drug Name & ConcentrationMinimum Breakthrough Detection Time
    Carmustine (BCNU) (3.3 mg/ml)23.8 Minutes
    Carboplatin (10 mg/ml)>240 Minutes
    Cisplatin (1 mg/ml)>240 Minutes
    Cyclophosphamide (Cytoxan) (20 mg/ml)>240 Minutes
    Dacarbazine (10.0 mg/ml)>240 Minutes
    Doxorubicin HCl (2 mg/ml)>240 Minutes
    Etoposide (20 mg/ml)>240 Minutes
    Fluorouracil (50 mg/ml)>240 Minutes
    Ifosfamide (50 mg/ml)>240 Minutes
    Methotrexate (25 mg/ml)>240 Minutes
    Mitomycin C (0.5 mg/ml)>240 Minutes
    Mitoxantrone (2 mg/ml)>240 Minutes
    Paclitaxel (6 mg/ml)>240 Minutes
    Thiotepa (10 mg/ml)24.1 Minutes
    Vincristine Sulfate (1 mg/ml)>240 Minutes

    Please note that the following drugs have low permeation times: Carmustine (BCNU) (3.3 mg/ml) 23.8 Minutes Thiotepa (10 mg/ml) 24.1 Minutes

    Warning: Do not use with Carmustine or Thiotepa.

    Device Description

    The subject device in 510(K) notification K221157 is JR Medic Blue Latex Examination Powder Free Gloves tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from Natural Rubber Latex, blue color, powder free and non-sterile (as per 21CFR 880.6250, class I).

    The subject device meets all the current specifications listed under the ASTM Specification D3578-2019, Standard Specification for Rubber Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019).

    AI/ML Overview

    The provided document describes the safety and performance testing of "JR Medic Blue Latex Examination Powder Free Gloves Tested for use with Chemotherapy drugs" to demonstrate substantial equivalence to a predicate device, rather than a study proving the device meets general acceptance criteria for a new medical device technology like an AI algorithm.

    Therefore, the requested information elements related to AI/algorithm performance, such as sample size for test set with provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, and how ground truth was established for the training set are not applicable to this document.

    The document focuses on the physical, chemical, and biological properties of examining gloves and their resistance to chemotherapy drugs.

    However, I can extract the acceptance criteria and reported device performance from the provided "G. SUMMARY OF NON-CLINICAL PERFORMANCE DATA" section, focusing on the glove's performance criteria.


    1. Table of Acceptance Criteria and the Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo determine the length of the glovesMin 230 mm for all sizesSmall: 304 mm, Medium: 304 mm, Large: 305 mm, X-Large: 305 mm
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo determine the width of the glovesSmall: 80+/-10mm, Medium: 95+/-10 mm, Large: 111+/-10 mm, X-Large: 115+/-10 mmSmall: 84 mm, Medium: 94 mm, Large: 105 mm, X-Large: 114 mm
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo determine the thickness of the glovesPalm: 0.08 mm min, Finger: 0.08 mm min for all sizesPalm: 0.31mm (all sizes), Finger: 0.38mm (all sizes)
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Tensile strength (Before Ageing)18Mpa Min for all sizesSmall: 33.16 Mpa, Medium: 32.93 Mpa, Large: 32.43 Mpa, X-Large: 32.36 Mpa
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Tensile strength (After Ageing)14Mpa Min for all sizesSmall: 30.78 Mpa, Medium: 30.73 Mpa, Large: 30.23 Mpa, X-Large: 30.16 Mpa
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Ultimate Elongation (Before Ageing)650% Min for all sizesSmall: 1320%, Medium: 1288%, Large: 1376%, X-Large: 1200%
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Ultimate Elongation (After Ageing)500% Min for all sizesSmall: 1066%, Medium: 1105%, Large: 1225%, X-Large: 1041%
    ASTM D3578-2019 Standard Specification for Rubber Examination GlovesTo Determine the physical properties - Stress at 500% Elongation (Before Ageing)5.5 Mpa Max for all sizesSmall: 5.1 Mpa, Medium: 5.2 Mpa, Large: 5.2 Mpa, X-Large: 5.2 Mpa
    ASTM D5151-2019 Standard Test Method for Detection of Holes in Medical GlovesTo determine the holes in the glovesAQL 2.5Gloves Passes AQL 1.5
    ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual PowderTo determine the residual powder in the gloves2 Mg/Glove MaxSmall: 0.21 mg/glove, Medium: 0.22 mg/glove, Large: 0.22 mg/glove, X-Large: 0.22 mg/glove
    ASTM D 5712-95 (Reapproved 2010) Standard Test Method for the Analysis of ProteinTo determine the extractable protein in the gloves.200 µg/ dm² Max for all sizesSmall: 43.12 µg/ dm², Medium: 41.16 µg/ dm², Large: 42.06 µg/ dm², X-Large: 46.24 µg/ dm²
    ISO 10993-10 Biological Evaluation of Medical Devices: Irritation TestTo determine the potential of the material under test to produce dermal irritation.Under the condition of study not an irritantUnder the condition of study not an irritant
    ISO 10993-10 Biological Evaluation of Medical Devices: Skin Sensitization TestTo determine the skin sensitization potential of the material.Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    ISO 10993-5:2009 Biological Evaluation of Medical Devices: Cytotoxicity TestTo evaluate the in vitro cytotoxic potential of the test item.Under the conditions of study non cytotoxicUnder the conditions of the study cytotoxic.
    Acute systemic toxicityTo determine the acute systemic toxicity potential of the test item.No specific acceptance criteria stated in table, but implication is to show no systemic toxicityUnder the conditions of the study, no systemic toxicity
    Material Mediated Pyrogenicity ISO 10993-11:2017(E) / USP 41<151>To determine the pyrogenic potential of the test item extract.Under the conditions of the study non pyrogenicUnder the conditions of the study non pyrogenic
    ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy DrugsTo assess breakthrough time for various chemotherapy drugs.Varies per drug, generally aiming for longer breakthrough times. Specific minimums were not explicitly stated as "acceptance criteria" but rather as "minimum breakthrough detection time" in the Indications for Use.Carmustine (BCNU) (3.3 mg/ml): 23.8 MinutesCarboplatin (10 mg/ml): >240 MinutesCisplatin (1 mg/ml): >240 MinutesCyclophosphamide (Cytoxan) (20 mg/ml): >240 MinutesDacarbazine (10.0 mg/ml): >240 MinutesDoxorubicin HCl (2 mg/ml): >240 MinutesEtoposide (20 mg/ml): >240 MinutesFluorouracil (50 mg/ml): >240 MinutesIfosfamide (50 mg/ml): >240 MinutesMethotrexate (25 mg/ml): >240 MinutesMitomycin C (0.5 mg/ml): >240 MinutesMitoxantrone (2 mg/ml): >240 MinutesPaclitaxel (6 mg/ml): >240 MinutesThiotepa (10 mg/ml): 24.1 MinutesVincristine Sulfate (1 mg/ml): >240 Minutes

    Non-Applicable Information for this Document:

    The following requested information is not applicable to the provided document as it describes the FDA 510(k) clearance for examination gloves and their resistance to chemotherapy drugs, which is a physical device, not an AI/algorithm-based diagnostic or therapeutic tool.

    1. Sample size used for the test set and the data provenance: Not applicable - tests are on gloves, not a data set for an algorithm.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable - "ground truth" concept does not apply in this context; instead, it's about objective physical/chemical measurements and biological responses.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.