510(k) Data Aggregation

    K Number
    K221020

    Validate with FDA (Live)

    Date Cleared
    2023-01-13

    (282 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Accuryn Monitoring System is intended for use in the drainage and/or collection of urine, and in the monitoring of urine output and core body temperature, in degrees Fahrenheit and degrees Celsius. The Accuryn Monitoring System with the Accuryn Sensing Urinary Catheter (SmartFoley®) - IAP UO Temp is also intended for use in the monitoring of intra-abdominal pressure. The measured pressures can be used as an aid in the diagnosis of intra-abdominal hypertension (IAH) and the associated clinical syndrome of abdominal compartment syndrome (ACS). The Accuryn Sensing Urinary Catheter is a single use device intended for short term use (less than 30 days).

    Device Description

    The Accuryn Monitoring System consists of three components, the Accuryn Sensing Urinary Catheter, Accuryn Urine Collection Set or Accuryn SmartFoley Adapter, and the Accuryn Monitor.

    The Accuryn Sensing Urinary Catheter is a single use, sterile, MR conditional, disposable (12F, 14F, and 16 F), two-way silicone urinary bladder catheter with two opposing drainage eyes and four lumens. The first lumen is for urine drainage, the second lumen for the retention balloon, the third lumen embedded with a thermistor, and the fourth lumen to measure intra-abdominal pressure (IAP).

    The Accuryn SmartFoley® comes in three configurations:

    • IAP, U/O, Temp (intra-abdominal pressure, urine output, and core temperature)
    • U/O. Temp (urine output and core temperature) ●
    • Adapter (that is compatible with commercially available Foley catheters between 6-24Fr)

    The Accuryn Urine Collection Set is provided permanently pre-connected to the Accuryn Sensing Urinary Catheter's drainage lumen conveys urine to a urine measurement cassette. which connects to the Accuryn Monitor. The cassette feeds the urine collection bag. Potrero Medical also provides a configuration of the Accuryn Urine Collection Set that is not provided pre-connected to the catheter/foley. It is known as Accuryn SmartFoley Adapter and enables the user to connect a Foley catheter of their choice to the Accuryn Urine Collection Set. The Accuryn Urine Collection Set and Accuryn SmartFoley Adapter function in conjunction with the Accuryn Monitor to measure and display urine output is considered a 510(k) exempt device per 21 CFR 876.1800 (urine flow or volume measuring system, product code EXY).

    The Accuryn Monitor is a portable electronic device which maintains urine flow from the foley through the Accuryn Urine Collection Set or Accuryn SmartFoley Adapter, measures the urine output volume in the measurement cassette, and displays the urine output. The Accuryn Monitor also measures and displays the temperature from the thermistor and intra-abdominal pressure (IAP) from the pressure lumen. The monitor contains software.

    AI/ML Overview

    The provided text is a 510(k) summary for the Accuryn Monitoring System. This document focuses on demonstrating substantial equivalence to a predicate device rather than outright proving device performance against pre-defined acceptance criteria in the manner of a novel device. Therefore, the information typically found in a study proving acceptance criteria for a new AI/ML medical device (like MRMC studies, detailed ground truth establishment, training set provenance, etc.) is largely not present in this type of submission.

    The "acceptance criteria" here are essentially that the new device configurations (smaller catheter sizes) perform comparably to the predicate device and don't introduce new safety or effectiveness concerns. The "study" proving this is a series of non-clinical performance tests designed to validate and verify the device's adherence to relevant standards and its functional characteristics.

    Here's an attempt to answer your questions based on the provided text, highlighting where information is absent due to the nature of a 510(k) submission for substantial equivalence:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of quantitative acceptance criteria with corresponding performance values in the way one might expect for a novel AI/ML device. Instead, it summarily states that the device "met all acceptance criteria" and "Performance testing data has demonstrated that the Accuryn Monitoring System is safe and effective for its intended use."

    The closest a comparison table comes to "acceptance criteria" is the "Comments" column in the Comparison of Technological Characteristics and Device Use with Predicate Device section, where "Same" indicates meeting the implicit criterion of being equivalent to the predicate.

    Comparison FeatureSubject DevicePredicate DeviceComments (Implicit Acceptance Criteria for Substantial Equivalence)
    Product ClassIIIISame
    Product CodeEZL, PHU, EXYEZL, PHU, EXYSame
    Indications for UseAs describedAs describedSubject device specifies the Accuryn Sensing Urinary Catheter (SmartFoley®) – IAP UO Temp is the catheter with intra-abdominal pressure measurement ability. The Accuryn SmartFoley UO Temp is an alternative option for users who do not need the IAP functionality. This is a clarification, not a difference in fundamental intended use.
    Use environmentHospitalHospitalSame
    UsersMedical StaffMedical StaffSame
    IAP UO Temp Catheter Size12Fr, 14Fr, 16Fr16FrTwo smaller French sizes are being introduced. Performance of these new sizes against relevant standards would be the implicit acceptance criteria.
    Lumens44Same
    Performance StandardASTM F623ASTM F623Same
    Use DurationFoley: <30 days; Monitor: Re-usableFoley: <30 days; Monitor: Re-usableSame
    SterileDisposable: Yes; Monitor: NoDisposable: Yes; Monitor: NoSame

    Reported Device Performance:
    The document broadly states: "Performance testing data has demonstrated that the Accuryn Monitoring System is safe and effective for its intended use" and "Performance testing and analysis of results provide assurance that the device meets its specifications, is acceptable for device functionality, and is safe and effective for its intended use without raising any new issues of safety and effectiveness as compared to the predicate."
    Specific numerical performance results (e.g., accuracy of temperature measurement with a specific tolerance) are not detailed in this summary. Instead, it lists the types of tests performed:

    • Urinary Catheter Tests per ASTM F623-19 and ISO 20696
    • Performance Testing (compliance with ASTM D3078-02, ASTM F88/F88M-15, ASTM D4332-14, ASTM D4169-16, ASTM F1980-16)
    • MR Compatibility
    • IAP Functionality Testing
    • Usability per ANSI AMMI IEC 62366-1
    • System and Software Verification and Validation per IEC 62304
    • Electromagnetic Compatibility and Electrical Safety
    • Cybersecurity
    • Sterilization
    • Biocompatibility

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document lists various engineering and bench tests (e.g., ASTM standards, electrical safety). For these types of tests, sample sizes are typically determined by the particular standard or internal quality protocols. This information is not detailed in the 510(k) summary. Given these are non-clinical hardware tests, "country of origin for data" or "retrospective/prospective" are not applicable in the typical sense of a clinical study. These are lab-based tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the ground truth for this medical device (a monitoring system and catheter) is established through physical standards, calibrated equipment, and engineering specifications, not expert consensus on medical images or diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as the evaluation is based on objective measurements against engineering specifications and standards, not subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was performed or required. This device is a monitoring system and catheter, not an AI-assisted diagnostic imaging tool. The document explicitly states: "No clinical data was required to demonstrate safe use of the Accuryn Monitoring System with the Accuryn SmartFolev IAP UO Temp Trav System 12 F and 14 F."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable in the context of an AI algorithm. The device itself (monitor and catheter) has standalone performance as a physical measurement device, which was evaluated through the non-clinical tests listed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for this device is based on engineering specifications, physical measurements from calibrated instruments, and compliance with recognized industry standards (e.g., ASTM, ISO, IEC). For instance, temperature measurements would be validated against a known, accurate temperature source, and pressure measurements against a calibrated pressure standard.

    8. The sample size for the training set

    This is not applicable. This is a hardware device with embedded software, not a machine learning model that requires a training set of data in the AI/ML sense.

    9. How the ground truth for the training set was established

    This is not applicable for the reasons stated above.