510(k) Data Aggregation

    K Number
    K220749

    Validate with FDA (Live)

    Date Cleared
    2022-05-13

    (60 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

    Device Description

    Surgical Face Mask is composed of Nylon and Spandex ear loops, Iron-plastic nose clip, Inner layer Spunbond polypropylene, Middle Melt-blown polypropylene and outer Spun-bond polypropylene. The meltblown polypropylene material acts as the filter that stops microbes from entering or exiting the mask. Surgical Face Masks feature pleats or folds. Three pleats are used to allow the user to expand the mask such that it covers the area from the nose to the chin. The type is ear-loop, where a string-like material is attached to the mask and placed behind the ears. The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.

    AI/ML Overview

    The provided document describes the non-clinical testing performed for a Surgical Face Mask (Model: 17.5X9.5CM EAR-LOOP) to demonstrate its substantial equivalence to a legally marketed predicate device. This is not an AI/ML medical device, and therefore, the requested information regarding AI/ML-specific study aspects (like expert ground truth establishment, MRMC studies, training/test set details, etc.) is not applicable to this document.

    The document focuses on the physical performance and biocompatibility of the surgical face mask, comparing it to established industry standards for surgical apparel.

    Here's the relevant information extracted and presented based on the document:


    Device: Surgical Face Mask, Model: 17.5X9.5CM EAR-LOOP
    Regulatory Class: Class II
    Product Code: FXX
    Regulation Number: 21 CFR 878.4040 (Surgical Apparel)
    Predicate Device: K203426, Surgical Face Mask (non-sterile)

    Study Type: Non-clinical (performance and biocompatibility)
    Clinical Test Conclusion: No clinical study was included in this submission.


    1. Table of Acceptance Criteria and Reported Device Performance

    The device is evaluated against ASTM F2100-19 Level 2 performance standards and ISO 10993 for biocompatibility.

    ItemPurposeAcceptance Criteria (ASTM F2100-19 Level 2)Reported Device Performance (Result; average performance if provided)
    Fluid Resistance (ASTM F1862)Assess the performance of a mask to resistance to a synthetic blood preparation targeted toward the mask at a set pressure29 out of 32 pass at 120 mmHgPASS (Results for individual lots not detailed beyond "PASS" matching the criteria)
    Particulate Filtration Efficiency (ASTM F2299)Assess the performance of a mask to penetration by sub-micron polystyrene latex particles of 0.1 micron≥ 98%PASS (Lot1: 99.17%, Lot2: 99.49%, Lot3: 98.90%)
    Bacterial Filtration Efficiency (ASTM F2101)Assess the performance of a mask to penetration by a prepared solution with known concentration of an indicator bacterial organism≥ 98%PASS (Lot1: 99.87%, Lot2: 99.9%, Lot3: 99.84%)
    Differential Pressure (Delta P) (EN 14683 Annex C)Assess the performance of a mask for resistance to air movement through the materials of the face of the mask< 6.0 mmH2O/cm²PASS (Lot1: 2.08 mmH2O/cm², Lot2: 2.08 mmH2O/cm², Lot3: 2.07 mmH2O/cm²)
    Flammability (16 CFR 1610)Assess the resistance of a mask to ignitionClass 1PASS (Class 1)
    Biocompatibility Testing (ISO 10993)
    CytotoxicityAssess the potential risk of Cytotoxicity of mask materialNon-CytotoxicPASS (Under the conditions of the study, the device is non-cytotoxic.)
    IrritationAssess the potential risk of Irritation of mask materialNon-IrritatingPASS (Under the conditions of the study, the device is non-irritating.)
    SensitizationAssess the potential risk of Sensitization of mask materialNon-SensitizingPASS (Under the conditions of the study, the device is non-sensitizing.)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Size: For each performance test (Fluid Resistance, Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability), 3 non-consecutive lots were tested, using a sample size of 32 per lot.
    • Data Provenance: The document does not explicitly state the country of origin for the non-clinical test data. Given the applicant and correspondent are based in China, it is highly likely the testing was conducted in China. The data is retrospective in the sense that it was collected and compiled for this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    N/A. This is a physical device, and its performance is evaluated against established engineering standards (ASTM, EN, ISO, CFR) through laboratory testing, not through expert consensus on qualitative data or medical imaging.

    4. Adjudication Method for the Test Set

    N/A. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in clinical evaluations, especially for AI/ML devices where human experts may disagree on interpretations. For physical device testing against quantitative standards, the test results are objective measurements evaluated directly against the predefined acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    N/A. MRMC studies are used to evaluate the diagnostic accuracy of AI/ML systems in conjunction with human readers. This document pertains to a physical medical device (surgical face mask) with performance validated via non-clinical laboratory testing.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    N/A. This question is also specific to AI/ML devices. The "standalone" performance here would refer to the mask's inherent physical properties and functionality, which are indeed tested without human intervention in their function, only human operation of the test equipment.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is defined by established engineering and material science standards (e.g., ASTM F2100, ASTM F1862, EN 14683, ISO 10993, 16 CFR 1610). The device's measured properties are directly compared to the quantitative or qualitative requirements specified in these standards.

    8. The Sample Size for the Training Set

    N/A. This device does not use an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    N/A. (See point 8)