510(k) Data Aggregation

    K Number
    K220699

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-04-08

    (29 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Prestige Coil System is indicated for arterial and venous embolizations in the peripheral vasculature.

    Device Description

    The Prestige Coil System is a series of specialized coils that are inserted into the vasculature under angiographic visualization to embolize peripheral vascular anomalies. The system consists of an embolization coil implant comprised of platinum/tungsten, affixed to a delivery pusher to facilitate insertion into the hub of a microcatheter. The system is available in various shapes, lengths, and sizes. The devices are to be placed into anomalies to create blood stasis, reducing flow in the target vasculature, and inducing thrombosis. Upon positioning coils into the vasculature, the coils are thermally detached from the delivery pusher in serial manner until the target vasculature is occluded.

    AI/ML Overview

    The provided text is a 510(k) summary for the Prestige Coil System, a vascular embolization device, which explicitly states that it is not an AI/ML powered device. Therefore, it does not involve any of the AI/ML related concepts you asked about, such as training sets, test sets, ground truth establishment, MRMC comparative effectiveness studies, or AI performance metrics.

    However, I can still extract the acceptance criteria and information about the studies performed for this non-AI medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Performance
    BiocompatibilityThe samples shall be biocompatible for their intended use based on the requirements of ISO 10993-1.Pass
    Shelf LifeThe samples shall meet established acceptance criteria based on device specifications after simulated aging and transportation/distribution simulation.Pass
    PackagingThe samples shall meet established acceptance criteria for packaging performance.Pass
    Gamma SterilizationThe samples shall meet established acceptance criteria for sterility.Pass
    CorrosionThe samples shall meet established acceptance criteria for corrosion.Pass
    Detachment Zone TensileThe samples shall meet established acceptance criteria for tensile strength.Pass
    Visual and Dimensional InspectionThe samples shall meet established acceptance criteria for visual physical damage and secondary diameter and length.Pass
    Simulated UseThe samples shall be prepared in accordance with the instructions for use and meet established acceptance criteria for device performance in a clinically relevant model.Pass
    Stretch-Resistance Thread Tensile TestingThe samples shall meet established acceptance criteria for tensile strength.Pass
    Usability ValidationThe devices shall be prepared in accordance with their respective instructions for use and meet established acceptance criteria for device performance in a clinically relevant model. The labeling shall be clear and understandable by the intended user.Pass

    2. Sample size used for the test set and the data provenance

    As this is a non-AI/ML device, the concept of a "test set" in the context of AI does not directly apply. The studies mentioned are bench tests and non-clinical device assessments. Specific sample sizes for each test are not provided in this summary. The data provenance is from non-clinical bench testing performed by the manufacturer (Balt USA, LLC) to demonstrate substantial equivalence to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as this is not an AI/ML device and does not involve establishing ground truth through expert review for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is not an AI/ML device and does not involve adjudication of expert assessments for a test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is not an AI/ML device and does not involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable, as this is not an AI/ML device and does not have an algorithm to be evaluated in standalone mode.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable, as this is not an AI/ML device. The "ground truth" for the device's performance in these non-clinical tests is based on pre-defined acceptance criteria derived from device specifications and recognized standards (e.g., ISO 10993-1).

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device and therefore no training set was used.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI/ML device and therefore no training set was used or ground truth established for it.