510(k) Data Aggregation

    K Number
    K220669

    Validate with FDA (Live)

    Date Cleared
    2022-05-16

    (70 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IPL HAIR REMOVAL HANDSET is an over-the-counter device intended for removal of unwanted body and/or facial hair.

    Device Description

    The IPL HAIR REMOVAL HANDSET is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon Quartz Lamp Tube to emit light and a skin sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot trigger a pulse emitting.

    AI/ML Overview

    The provided document is a 510(k) summary for the IPL HAIR REMOVAL HANDSET, Model: IPL-666. It focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria or extensive clinical study outcomes.

    Therefore, much of the requested information regarding specific acceptance criteria, device performance, sample sizes, expert involvement, efficacy studies (MRMC or standalone), and training set details for AI/algorithm performance cannot be extracted from this document. This type of document typically focuses on technical and safety equivalence to a legally marketed predicate.

    However, I can provide the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not specify quantitative acceptance criteria for device performance related to hair removal efficacy. Instead, it relies on demonstrating substantial equivalence to a predicate device based on technical characteristics and safety testing.

    Acceptance Criteria (Not Explicitly Stated for Performance)Reported Device Performance (Summary of Technical Characteristics)
    Safety and Functionality Standards:
    Electrical Safety (IEC 60601-1, -1-2, -1-11)Device complies with these standards.
    Particular Requirements for Home Light Therapy (IEC 60601-2-83)Device complies with this standard.
    Photobiological Safety (IEC 62471)Device complies with this standard.
    Biocompatibility (ISO 10993-5, -10)Skin-contacting materials comply with these standards.
    Software Verification & ValidationPerformed according to FDA guidance.
    Technical Design (Compared to Predicate):
    Light source: Intense Pulsed LightSame as predicate.
    Energy medium: Xenon Quartz Tube / Xenon lampSimilar to predicate (Xenon Arc Flashlamp / Xenon lamp).
    Wavelength range (nm): 470nm ~1100nmSimilar to predicate (475nm1200nm for predicate, 5101100nm for reference).
    Energy density (J/cm²): Max 2.49J/cm²Similar to predicate (Max 5J/cm² for predicate, 2.0-4.5 J/cm² for reference).
    Spot size (cm²): 3cm²Same as predicate (1.72 cm² or 3.02 cm² for predicate, various for reference).
    Pulse duration: 11.5-15msSimilar to predicate (11-12ms for predicate, 7.5-14ms for reference).
    Output intensity levels: 5 levelsSame as predicate.
    Skin sensor: PresentSame as predicate.
    Intended use: Removal of unwanted body and/or facial hairSame as predicate.
    Applicable skin phototypes: Fitzpatrick I-VSame as reference device. Unknown for predicate.

    2. Sample size used for the test set and the data provenance
    The document does not detail a clinical test set with human subjects for performance evaluation. The "tests" mentioned are non-clinical (electrical safety, biocompatibility, software V&V) and do not involve a "test set" in the context of clinical performance data. Therefore, data provenance (country of origin, retrospective/prospective) is not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not applicable as the document does not describe a clinical test set requiring expert-established ground truth for efficacy.

    4. Adjudication method for the test set
    Not applicable, as no clinical test set for efficacy is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done
    No MRMC comparative effectiveness study is mentioned. The document focuses on demonstrating substantial equivalence based on technical specifications and safety testing, not direct clinical performance comparison with human readers or AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    This device is an IPL hair removal handset, which is a physical medical device, not an AI/algorithm. Therefore, a standalone algorithm performance study is not applicable.

    7. The type of ground truth used
    Not applicable, as no clinical performance study requiring ground truth for efficacy is described. The "ground truth" in this context would generally relate to measured hair reduction, which is not presented here.

    8. The sample size for the training set
    Not applicable, as this is a physical device and not an AI/algorithm-based product requiring a training set.

    9. How the ground truth for the training set was established
    Not applicable, as this is a physical device and not an AI/algorithm-based product requiring a training set.