510(k) Data Aggregation

    K Number
    K220653

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-06

    (60 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Astroglide® Glycerin & Paraben Free is a personal lubricant for penile, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    Astroglide® Glycerin & Paraben Free is a non-sterile, water-based liquid personal lubricant for penile, anal, and/or vaginal application. This product is not a spermicide or contraceptive. It is compatible with natural rubber latex and polyisoprene condoms. It is not compatible with polyurethane condoms. The subject device is sold as an overthe-counter (OTC) product in a 2.5 oz. clear PETE bottle with a screw on polypropylene flip-top cap. The bottle is labeled front and back and then packaged in a cardboard carton which constitutes the final packaging. This device is composed of water, butylene glycol, propylene glycol, xylitol, methyl gluceth-20, polyquarternium-7, hydroxyethyl cellulose, potassium sorbate, sodium benzoate, citric acid.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a personal lubricant called Astroglide® Glycerin & Paraben Free. It primarily focuses on demonstrating substantial equivalence to a previously marketed predicate device, rather than providing details on a study of a device's performance against specific clinical acceptance criteria for diagnostic or AI-driven devices.

    Therefore, much of the requested information (sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone AI performance, training set details) is not applicable to this type of regulatory submission for a personal lubricant.

    However, I can extract information related to the acceptance criteria for the product's physical and chemical properties, as well as its biocompatibility and condom compatibility.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Test Method)Reported Device Performance
    ColorClear to slight goldenNot explicitly stated as "reported performance," but the device is described as having this specification.
    ClarityClearNot explicitly stated as "reported performance," but the device is described as having this specification.
    OdorSlight sweet odorNot explicitly stated as "reported performance," but the device is described as having this specification.
    Absence of particulate matterAbsentNot explicitly stated as "reported performance," but the device is described as having this specification.
    Viscosity1000-2800 cPNot explicitly stated as "reported performance," but the device is described as having this specification.
    pH3.5-5.5Not explicitly stated as "reported performance," but the device is described as having this specification.
    Osmolality230-300 mOsm/kg, 1:5 dilution factorNot explicitly stated as "reported performance," but the device is described as having this specification.
    Total yeast/mold count (TYMC)<10 cfu/mL (USP <61>)Not explicitly stated as "reported performance," but the device is described as having met this specification.
    Total aerobic microbial count (TAMC)<100 cfu/mL (USP <61>)Not explicitly stated as "reported performance," but the device is described as having met this specification.
    Presence of Pathogenic OrganismsAbsent (USP <62>) (Staphylococcus aureus, Pseudomonas aeruginosa, and Candida albicans)Not explicitly stated as "reported performance," but the device is described as having met this specification.
    Antimicrobial effectivenessMeets USP <51> acceptance criteria for Category 2 products.Not explicitly stated as "reported performance," but the device is described as having met this specification.
    Biocompatibility (Cytotoxicity)Non-cytotoxic (per ISO 10993-5:2009/(R)2014)Demonstrated non-cytotoxic
    Biocompatibility (Sensitization)Non-sensitizing (per Human Repeat Insult Patch Testing (HRIPT) per ASTM D6355)Demonstrated non-sensitizing
    Biocompatibility (Irritation)Non-irritating (per Human Repeat Insult Patch Testing (HRIPT) per ASTM D6355)Demonstrated non-irritating
    Biocompatibility (Systemic Toxicity)Not systemically toxic (per ISO 10993-11:2017)Demonstrated not systemically toxic
    Shelf Life (Device Specifications)Met specifications in Table 1 across shelf-lifeMet specifications at all time points over 22 months
    Condom Compatibility (Natural rubber latex)Compatible (per ASTM D7661-18)Compatible
    Condom Compatibility (Polyisoprene)Compatible (per ASTM D7661-18)Compatible
    Condom Compatibility (Polyurethane)Not Compatible (per ASTM D7661-18)Not Compatible

    2. Sample size used for the test set and the data provenance

    • For physical/chemical specifications: The document refers to standard test methods (e.g., USP <61>, USP <62>) but does not specify the sample size for these tests. The data provenance is implied to be from BioFilm, Inc.'s internal testing, likely from the U.S.
    • For biocompatibility: No specific sample sizes for human subjects or animal models are provided for the HRIPT or systemic toxicity tests. For cytotoxicity, it's generally in vitro testing. Data provenance is implied to be from BioFilm, Inc.'s testing.
    • For condom compatibility: No specific sample size (i.e., number of condoms tested) is explicitly mentioned, but the testing was conducted according to ASTM D7661-18. Data provenance is implied to be from BioFilm, Inc.'s testing.
    • For shelf life: The study was an accelerated aging study conducted for 8 months at 40°C, per ASTM F1980-16. No sample sizes (number of devices) are given.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not a study requiring expert clinical review or ground truth establishment for diagnostic accuracy. The ground truth for the physical/chemical properties, biocompatibility, and condom compatibility are defined by the standardized test methods themselves.

    4. Adjudication method for the test set

    Not applicable. There is no subjective assessment requiring adjudication in this type of testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a personal lubricant device, not an AI-driven diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a personal lubricant device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the various acceptance criteria were established by:

    • Standardized laboratory methods/specifications: For color, clarity, odor, particulate matter, viscosity, pH, osmolality, microbial counts (TYMC, TAMC, pathogens), and antimicrobial effectiveness, the ground truth is defined by specified ranges, limits, or absence/presence criteria according to recognized pharmacopeial (USP) or internal standards.
    • International Standards (ISO) and ASTM Standards:
      • Biocompatibility: ISO 10993-5 (cytotoxicity), ASTM D6355 (sensitization/irritation), ISO 10993-11 (systemic toxicity).
      • Shelf Life: ASTM F1980-16 (accelerated aging for medical devices).
      • Condom Compatibility: ASTM D7661-18.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI-driven device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.