Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K220530

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-08-17

    (174 days)

    Product Code
    Regulation Number
    868.2775
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered.

    Device Description

    The TetraGraph Neuromuscular Transmission (NMT) Monitor (TetraGraph) is a portable, batteryoperated EMG-based neuromuscular transmission monitor for use perioperative and in recovery and critical care environments following or during the application of Neuromuscular block. TetraGraph undertakes this function by electrical stimulation of the peripheral nerve and directly measuring the evoked response of the muscles (Muscle Action Potential (MAP)), thus providing a quantitative and automatic measurement of muscle response to a stimulus using electromyography (EMG). The TetraGraph is a prescription-only medical device and is indicated for use in hospitals. TetraGraph consists of the following main components: TetraGraph Monitor, TetraSens Electrode, TetraSens Pediatric Electrode (new accessory), Philips Interface (new and optional accessory), and Pole clamp kit (optional accessory).

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Senzime AB TetraGraph Neuromuscular Transmission Monitor, specifically for the addition of new accessories: the TetraSens Pediatric electrode and the Philips Interface. The submission aims to demonstrate substantial equivalence to a legally marketed predicate device (TetraGraph Neuromuscular Transmission Monitor, K190795).

    Based on the provided document, here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The document does not explicitly present a table of acceptance criteria with corresponding performance results in a clear, summarized format. Instead, it refers to generalized performance testing.

    However, it implicitly states that performance testing was conducted for the TetraGraph with the new TetraSens Pediatric electrodes and that "all of the acceptance criteria for this device" were met. These criteria are noted to be "the same or very similar to the predicate device's specifications."

    The types of performance testing mentioned are:

    Acceptance Criteria TypeReported Device Performance
    BiocompatibilityMet compliance in accordance with ISO 10993-1, including cytotoxicity, sensitization, and irritation.
    Electrode Performance (General)Met compliance with ANSI/AAMI EC12: 2000 for Disposable ECG Electrodes.
    EMG Evoked Response DetectionCompleted testing, and acceptance criteria were met.
    Electrode Tensile StrengthCompleted testing, and acceptance criteria were met.
    Other Performance TestingCompleted and met all acceptance criteria (e.g., shelf life, software, electrical and EMC).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    The document does not specify the sample sizes used for any of the performance tests mentioned (biocompatibility, electrode testing, EMG detection, tensile strength, shelf life, software, electrical, EMC).

    The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. It implies internal testing conducted by the manufacturer or their designated testing facilities.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This information is not provided in the document. The tests described are largely bench-top and engineering performance tests, which typically do not involve experts to establish ground truth in the same way clinical studies or image-based diagnostics would.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable/not provided as the tests described are technical performance tests rather than clinical evaluations requiring adjudication of subjective outcomes.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This type of study is not applicable and was not conducted. The device is a neuromuscular transmission monitor, not an AI-assisted diagnostic tool that aids human readers. The document explicitly states: "There was no clinical testing required to demonstrate that TetraGraph with the new TetraSens Pediatric electrodes is substantial equivalent to the predicate device as the predicate has similar technological characteristics."

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The device itself is a measurement and monitoring device, an "algorithm only" or "standalone" performance would refer to its accuracy and reliability in measuring neuromuscular transmission. The performance testing (EMG evoked response detection, compliance with standards) serves this purpose by demonstrating the technical capabilities of the device in isolation, even though it's used by humans. The document confirms that "performance testing made for the device" was completed and met acceptance criteria.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the described performance tests, the "ground truth" would be established by:

    • Scientific Standards: For biocompatibility (ISO 10993-1) and electrode testing (ANSI/AAMI EC12: 2000), the ground truth is defined by the requirements and test methods outlined in these international and national standards.
    • Engineering Specifications: For EMG evoked response detection, electrode tensile strength, and other performance tests, the ground truth is based on the device's design specifications and expected technical performance against which the measured results are compared.

    8. The sample size for the training set:

    This information is not applicable/not provided. The TetraGraph is a measurement device and the data provided does not suggest it uses machine learning or AI models that require a "training set" in the conventional sense. The submission focuses on hardware accessories and their technical performance.

    9. How the ground truth for the training set was established:

    This information is not applicable/not provided for the same reasons as in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1