Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K220529

    Validate with FDA (Live)

    Date Cleared
    2022-03-18

    (22 days)

    Product Code
    Regulation Number
    890.3860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Merits P335 Powered Wheelchair is intended to provide indoor mobility to persons limited to a seating position that are capable of operating a powered wheelchair.

    The Merits P335 Powered Wheelchair is to provide indoor and outdoor mobility to persons limited to a seating position that are capable of operating a powered wheelchair.

    Device Description

    The Merits P335 Powered Wheelchair is battery powered, front wheel motor driven and is controlled by the PG power wheelchair R-net 120 amp controller. The user interface is a joystick. P335 is powered by two 12 VDC 60ah batteries. The batteries are charged by 6A off-board charger connect with 3-pin Microphone Connector to charging socket on joystick. The approximate driving range on fully charged batteries is up to 30km (19mi). The chair frame is a welded steel construction and includes two front drive wheels with drive units (including motor, gear, and brake), batteries and rear pivoting casters. Depending on users' needs, the joystick motor control is mounted to the left or right armrest. When the user activates the joystick, the controller receives a signal to release the brakes. With the brakes released, the wheelchair is allowed to move in the direction the joystick is actuated. When the user releases the joystick, the chair slows to a stop and the brakes are automatically re-engaged. The solenoid electromechanical brakes allow the user stop by letting go of the joystick.

    The device is equipped with a Power Seating System consists of a power tilt unit, a power recline module and an optional power elevating seat module. The tilt, reclining and elevating systems are actuated by 24V DC motorized linear actuator. The tilt system includes one motorized linear actuator caused the seat frame to shift forward. This enhances stability since the center of gravity is kept substantially in place while the user is tilting. The recline system include one motorized linear actuator change the position of the backrest with respect to the seat pan. The optional elevating seat module use one motorized linear actuator which allows the user to elevate the seat pan for easy accessing things stored in higher places.

    The upholstery of the device complies with ISO 7176-16:2012: Wheelchairs -- Part 16: Resistance to ignition of postural support devices.

    The device can be operated on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a powered wheelchair (Merits P335). It does not describe an AI/ML powered medical device, an imaging device, or diagnostic software. Therefore, the requested information about acceptance criteria for an AI/ML device, sample sizes for test sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, training set details, or data provenance is not applicable to this document.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Permobil F3 FWD Powered Wheelchair). The acceptance criteria described are in the form of compliance with various ISO and IEC standards related to powered wheelchairs, and the "study" is the non-clinical testing performed to confirm adherence to these standards.

    Here's the relevant information based on the document:

    1. A table of acceptance criteria and the reported device performance:

    The acceptance criteria are implicitly meeting the requirements of the listed international standards (ISO and IEC). The reported device performance is that the "results of the testing confirm that the device meets specifications and is substantially equivalent to the predicate device."

    STD No.Title of StandardFDA Rec. No.Reported Device Performance
    ISO 7176-1:2014Determination of Static Stability16-195Passed
    ISO 7176-2:2017Determination of Dynamic Stability of electricwheelchairs16-202Passed
    ISO 7176-3:2012Determination of effectiveness of brakes16-192Passed
    ISO 7176-4:2008Energy consumption of electric wheelchairs andscooters for determination of theoretical distancerange16-162Met
    ISO 7176-5:2008Determination of overall dimensions, mass andmaneuvering space16-163Met
    ISO 7176-6:2018Determination of maximum speed, acceleration anddeceleration of electric wheelchairs16-204Met
    ISO 7176-7:1998Method of Measurement of Seating and WheelDimensions16-196Met
    ISO 7176-8:2014Requirements and test methods for static, impact andfatigue strengths16-197Passed
    ISO 7176-9:2009Climatic tests for wheelchairs16-167Met
    ISO 7176-10:2008Determination of obstacle-climbing ability ofelectrically power wheelchairs16-164Met
    ISO 7176-11:2012Test dummies16-190Used as specified
    ISO 7176-13:1989Determination of coefficient of friction of test surfaces16-25Met
    ISO 7176-14:2008Power and control systems for electrically poweredwheelchairs and scooter- Requirements and testmethods.16-165Met
    ISO 7176-15:1996Requirements for Information Disclosure,Documentation and Labeling16-27Met
    ISO 7176-16:2012Resistance to ignition of postural support devices16-191Met
    ISO 7176-21:2009Requirements and test methods for electromagneticcompatibility of electrically powered wheelchairs andscooters, and battery chargers.16-166Met
    ISO 7176-25:2013Batteries and charger for powered wheelchairs16-194Passed
    ISO 10993-1:2018Biological evaluation of medical devices - Part 1:Evaluation and testing within a risk managementprocess2-258Compliant
    IEC60601-1-2:2020Medical electrical equipment - Part 1-2: Generalrequirements for basic safety and essentialperformance - Collateral Standard: Electromagneticdisturbances - Requirements and tests19-36Met

    2. Sample size used for the test set and the data provenance: Not applicable. This is a physical device, and testing is performed according to the specified standards (e.g., durability cycles, static/dynamic stability tests), not on a "test set" of data in the AI/ML sense. Data provenance is not specified beyond "non-clinical testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical medical device meeting performance standards is established by adherence to the objective criteria and methodologies outlined in the international standards, not by expert consensus on a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-powered device or a diagnostic imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" here is compliance with established international performance and safety standards for powered wheelchairs (e.g., ISO 7176 series, IEC 60601-1-2) which involve objective measurements and tests.

    8. The sample size for the training set: Not applicable. There is no AI/ML component; therefore, no training set.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1