510(k) Data Aggregation

    K Number
    K220526

    Validate with FDA (Live)

    Device Name
    HD+12, HD+15
    Manufacturer
    Date Cleared
    2023-05-15

    (445 days)

    Product Code
    Regulation Number
    870.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The function of the device is the acquisition and transmission of the ECG signal in order to support diagnosis of the patient's conditions.

    HD+ is a wireless or USB (with appropriate option) ECG acquisition device, to be used primarily as common front-end for standard PC/tablet platforms (Windows/MAC OS/others), for resting ECG applications.

    The device implements wireless communication vireless technology or wired with USB communication. With both connection modes, HD+ sends the data to the receiver device without performing any analysis or filtering. HD+ is not intended for monitoring of the cardiac function or to diagnose the patient's health condition. Display, print and analysis applications on the receiver device are separate products. HD+ is not able to permanently store the acquired data, therefore it does not work unless a connection has been established with a receiver application. Furthermore, HD+ does not collect any of the patient's name, age, previous health conditions etc.).

    HD+ detects the QRS complexes and transmit the receiving device. The QRS detection function is intended for patients aged 12 years or older.

    *HD+ is indicated for the acquisition of ECG signals with the patient in resting conditions, for example for diagnostic ECG's and rhythm strips

    *HD+ is suitable for working at high altitudes, with restrictions.

    *HD+ is intended for use on patients, with no limits of age or gender, except for the HR detector function which is intended for patients 12 years or older

    • HD+ is intended for use in medical facilities (hospitals, clinics), at home or emergency settings (ambulances).

    *HD+ is intended for use by a physician, nurse or other healthcare professional (e.g., ECG technician) who acts following orders by a physician or authorized nurse, including when operated in home environments

    *HD+ is not intended for real-time monitoring of vital physiological parameters.

    Device Description

    HD+12 and HD+15 are the new revisions of Cardioline's HD+ wireless acquisition module. The HD+ is a digital portable acquisition device which can acquire the electrocardiographic signal of 12 and 15 standard leads. Connected with a receiver via Bluetooth or USB (with the optional USB connector), the HD+ performs no analysis or filtering but sends acquired data to a host application where the User Interface is implemented. By default, the device includes a demo compatible software "HD+ Display" whose purpose is providing a UI for viewing the traces acquired by HD+. HD+ Display provides a simple UI, sufficient to configure the application to connect to an HD+ device via Bluetooth (the application automatically detects if an HD+ device is connected via USB and does not require additional configuration) and sufficient to display the ECG traces as received by HD+. HD+ Display does not provide any clinical functionality; therefore, it does not provide any capability to store, print or analyze the acquired ECG.

    lt is the host (PC or Tablet separated by HD+) which performs the analysis. HD+ is not intended to control or analysis heart function and/or to diagnose the patient's health status. The analysis program on the host is a separate product not marketed with the HD+.

    HD+ is a wireless acquisition device, to be primarily used as ECG front-end acquisition device for PC/tablet (Windows/MAC OS/Android/iOS/others) standard platforms for Resting ECG applications. Depending on performance/price ratio, HD+ could be also used with selected embedded electrocardiographs. HD+ allows the patient to be ambulatory.

    HD+ uses a standard Bluetooth data transmission technology to transmit 12-lead and 15-lead ECG data over a proximity range, providing electrical insulation and freedom of movement for the patient. The device implements the wireless communication via Bluetooth wireless technology. The Bluetooth radio protocol is implemented by a dedicated module, FCC compliant. In order to support the data transmission speed of the application, the device implements the BLE 5 protocol with DLE (Data Length Extension) and 2M PHY (bandwidth up to 2 Mbit/s). The minimum specifications of the device connected to HD+ is BLE 4.2 to support the band required by acquisitions at 500 s/s. In order to operate at 1000 s/s, the connected device must be BLE 5.0 (or higher) and have a compatible 2M PHY radio (2 Mbit/s). Alternatively, in addition to Bluetooth connectivity described above, the HD+ has an optional USB interface that can be used to transmit data. The USB interface provides an electrical insulation offering two means of patient protection (2MPP), allowing HD+ to be connected to any IT equipment conforming to IEC 62368-1.

    The HD+ function consists of acquiring and transmitting ECG signals for display processing ECG signals for the purpose of supporting the diagnosis of patient conditions. The device does not store nor does it associate patient identification data to the acquired signal, nor does it perform analysis on such signal. The HD+ is used solely for transmission of ECG signals from patient to a host analysis platform.

    The HD+ transmits a continuous stream of ECG samples at a rate of 500 s/s or 1000 s/s, with a resolution of 0.817 uV/LSB or 2.495 uV/LSB, selectable by the calling application of the host analysis platform. The average required transmission throughput for sending 15 leads at 1000 s/s is approximately 155kbit/s, while to send 12 leads at 500 s/s the required throughput is approximately 54 kbit/s. BLE 5 2M PHY provides the bandwidth needed to support the maximum required throughput.

    The BLE communication link ensures that data is either received correctly or not received at all. It is up to the host application to detect packet losses, handling the data gaps appropriately (e.g. by filling the stream with invalid dummy samples, signaling transmission errors etc...). This approach has been preferred over enabling the data retransmission (supported by the module) to reduce the data iitter and transmission delay.

    HD+ uses standard 12 lead or 15 lead ECG cables to acquire the physiological signal from the patient. HD+ is light and compact, comfortable to wear, minimizing motion artifacts caused by traditional electrodes and patient cables.

    HD+ offers full ECG acquisition - meeting the standards used in clinical and diagnostic applications (AAMI, ANSI, AHA, ACC). HD+ uses a LED indicator to comfortably monitor the link status (off when unit is powered down, blinking when unit is attempting to connect with the receiver, steady when unit is connected with the receiver). HD+ uses a programmable key to send macro commands to the receiving system (i.e. acquire and print an ECG). Low-power technology allows continuous usage of the device for more than 10 hours (from full battery charge).

    The HD+ continuously transmits the acquired data to a computer platform where compatible software, a Host Application authorized by Cardioline, is installed that acts as its User Interface. As an example of such Host Application, the device includes a compatible demo Windows application "HD+ display" from Cardioline to visualize the ECG traces and check the functionality of the results of the analysis must always be validated by qualified, trained medical personnel and the HD+ is intended for use in a medical environment. HD+ is intended to be used on adult and all pediatric patients. The device must be handled with care by taking all the necessary precautions in order to prevent and avoid shocks, vibrations, heat sources, liquids and anything else that may damage it.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Cardioline HD+12, HD+15), which is a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device. This type of submission focuses on demonstrating equivalence rather than proving a device meets specific acceptance criteria through a clinical study in the way an AI/ML algorithm might.

    Therefore, the document does not contain details about acceptance criteria or a study proving the device meets those criteria, especially in the context of an AI/ML algorithm's performance. The device described in the document is an ECG signal acquisition and transmission device, explicitly stating that it "sends the data to the receiver device without performing any analysis or filtering" and "is not intended for monitoring or analysis of the cardiac function or to diagnose the patient's health condition." It also mentions that "clinical evaluation is not required" because it's an acquisition device without analysis, diagnosis, or monitoring features.

    However, I can extract information related to performance testing conducted to support the substantial equivalence claim, which serves a similar purpose to meeting acceptance criteria for a non-AI/ML device.

    Here's an analysis based on the provided text, while acknowledging that it doesn't align with the detailed AI/ML study requirements in the prompt:


    Acceptance Criteria and Device Performance (for a non-AI/ML acquisition device)

    The document primarily focuses on demonstrating that the new devices (HD+12, HD+15) are substantially equivalent to a predicate device (HD+) by showing adherence to relevant standards and similar performance characteristics. The "acceptance criteria" here are essentially compliance with safety and performance standards for a physiological signal acquisition device.

    1. Table of Acceptance Criteria (based on standards compliance and comparative features):

    Category / "Acceptance Criteria" (Standard/Feature)Reported Device Performance (HD+12, HD+15)Notes on "Acceptance" / Compliance
    Safety Standards
    IEC 60601-1 (General medical electrical equipment)Full compliance demonstratedMet
    IEC 60601-2-25 (ECG Specific)Full compliance demonstratedMet
    IEC 60601-1-11 (Home healthcare environment)Full compliance demonstratedMet (Additional for home use compared to predicate)
    Electromagnetic Compatibility (EMC)
    IEC 60601-1-2Full compliance demonstratedMet
    Radio Equipment Standards
    ETSI EN 300 328Full compliance demonstratedMet
    ETSI EN 301 489-1Full compliance demonstratedMet
    ETSI EN 301 489-17Full compliance demonstratedMet
    ETSI EN 300 440-2Full compliance demonstratedMet
    FCC CFR 47 Part 15 (US)Full compliance demonstratedMet
    Specific ECG Performance Parameters (based on IEC 60601-2-25)
    Patient Derivation PolarityTestedImplicitly meeting expected performance for substantial equivalence.
    Minimum System Switch LeadsTestedImplicitly meeting expected performance for substantial equivalence.
    Goldberger and Wilson DerivationsTestedImplicitly meeting expected performance for substantial equivalence.
    Input Impedance and Circuit LeadTestedImplicitly meeting expected performance for substantial equivalence.
    Common Mode Rejection (CMRR)≥100dB (Predicate: 115 dB)Equivalent performance (considered acceptable).
    Noise LevelTestedImplicitly meeting expected performance for substantial equivalence.
    Writing Speed and Trace WidthTestedImplicitly meeting expected performance for substantial equivalence.
    Channels InteractionTestedImplicitly meeting expected performance for substantial equivalence.
    High Frequency ResponsePerformances equivalent to 0.05-150 Hz (@ 500 c/s); 0.05-300 Hz (@ 1000 c/s) (Predicate: 0.05-300 Hz)Same performance where applicable.
    Low Frequency ResponseTestedImplicitly meeting expected performance for substantial equivalence.
    Linearity and Dynamic RangeHiRES mode: +/- 330mV; StdRES mode: +/- 500 mV (Predicate: +/- 400mV @ < 1uV/LSB)HIRES equivalent; StdRES offers higher dynamic range with reduced resolution. Considered acceptable.
    Minimal Signal ResponseTestedImplicitly meeting expected performance for substantial equivalence.
    Sampling and Resolution500 and 1000 samples/second/channel (analysis); 24 bit A/D; HiRES: <1 µV/LSB; StdRES: ~2.5 µV/LSB (Predicate: 1000 samples/s; 24 bit; 20 bit, <1uV/LSB)Allows for downgrading if bandwidth is insufficient; HIRES equivalent; StdRES offers option. Considered acceptable.
    Recording SpeedTestedImplicitly meeting expected performance for substantial equivalence.
    ECG DistortionTestedImplicitly meeting expected performance for substantial equivalence.
    Impulse Visibility of Pace-MakerSoftware on 128K c/s simultaneous on lead pairs; Impulse duration: 0.2-2ms; Impulse width: 2-250mV (Predicate: Hardware detection + digital filtering)Increased detection sensitivity, IEC 60601-2-25 requirements met. Considered acceptable.
    Internal Electrical Source RequirementsTestedImplicitly meeting expected performance for substantial equivalence.
    Functional Features
    ECG Leads12 and 15 Leads (Predicate: 12 Leads)Additional 3 optional leads.
    Wireless SystemBLE 5 or higher with DLE and 2M PHY support (Predicate: Bluetooth 2.1 + EDR)Updated to most current Bluetooth version.
    USB CommunicationUSB Communication Device Class (RS-232 port emulation) (Predicate: None)Addition of USB option.
    Patient Cable10 wire replaceable wire patient lead; 13 wire replaceable patient cable (Predicate: 10 wire single connector)Additional cable for HD+15.
    Target PopulationAdults and pediatric patients (same as predicate)Same.
    Use EnvironmentHospitals, Clinics, Home, Ambulance (Predicate: Hospitals, Clinics)Expanded.
    UserPhysician, nurse, other healthcare professional (e.g., ECG technician) (Predicate: Nurse, Physician and trained medical personnel)Expanded.

    Regarding the AI/ML-specific criteria from your prompt:

    The provided document does not include information for most of these categories because the device is a data acquisition and transmission device, not an AI/ML diagnostic or analysis tool. The key statement is: "HD+ is only an acquisition device without an analysis, diagnosis and monitoring features, so clinical evaluation is not required."

    Here's an assessment based on the document's content:

    1. Sample sized used for the test set and the data provenance: Not applicable. The "test set" here refers to specific performance tests on the device's hardware and software (e.g., electrical signal accuracy, EMC), not a clinical dataset for an AI algorithm.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an ECG acquisition device relates to physical and electrical properties, not expert interpretation of ECGs (as the device itself performs no analysis).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For the "performance tests" mentioned (e.g., noise level, frequency response), the ground truth is established by engineering standards, calibrated measurement equipment, and known electrical signals, not clinical outcomes or expert consensus on patient data.
    7. The sample size for the training set: Not applicable (not an AI/ML device).
    8. How the ground truth for the training set was established: Not applicable (not an AI/ML device).

    In summary, the provided FDA 510(k) summary is for a traditional medical device (ECG signal acquisition system) and thus does not contain the specific AI/ML-related study and acceptance criteria details requested in your prompt.