510(k) Data Aggregation

    K Number
    K220463

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-18

    (90 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Osteotomy Truss System (OTS) is intended to be used for internal bone fractures or osteotomies in the foot, such as:

    • · Opening wedge osteotomies of the foot including osteotomies for Hallux Valgus
    • · Opening wedge of Medial Cuneiform or Cotton osteotomies
    • · Lateral Column Lengthening (Evans Lengthening Osteotomy or Calcaneal Z Osteotomy)
    • · Metatarsal/Cuneiform osteotomies
    • · Nonunion of arthrodesis of the Midfoot including Metatarsal/Cuneiform osteotomies (TMT or Lapidus)
    • · Hindfoot osteotomies
      The device is intended for use with supplemental fixation. The Osteotomy Truss System is not intended for use in the spine.
    Device Description

    The Osteotomy Truss System (OTS) consists of three implant designs in a variety of footprints and opening wedge height options to accommodate the patient's anatomy. It is intended to be used with supplemental fixations.
    The device is an open architecture truss design mathematically formulated to provide structural support with open space throughout the implant for bone growth and fusion. The 4WEB additive manufacturing process provides a hierarchical surface roughness. The implant is made from Ti6Al4V alloy.

    AI/ML Overview

    This 510(k) premarket notification describes a design update to an existing device, the Osteotomy Truss System (OTS). The submission focuses on demonstrating substantial equivalence to the predicate devices through non-clinical performance testing. Therefore, it does not involve clinical studies or in-depth performance metrics typically associated with AI/ML devices.

    Here's an analysis based on the provided text, focusing on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard Reference)Reported Device Performance (Summary)
    Static axial compression (ASTM F2077)Sufficient for intended use
    Dynamic axial compression fatigue (ASTM F2077)Sufficient for intended use
    Expulsion testing (ASTM F-04.25.02.02)Sufficient for intended use
    MR image artifact (ASTM F2119)Sufficient for intended use
    MR induced heating (ASTM F2182)Sufficient for intended use
    MR induced torque (ASTM F2213)Sufficient for intended use
    MR induced displacement force (ASTM F2052)Sufficient for intended use
    Finite element analysisDoes not introduce a new worst-case compared to previously cleared devices for mechanical properties.

    2. Sample Size Used for the Test Set and Data Provenance

    This submission relies on non-clinical performance testing (in vitro/bench testing and computational modeling). The document does not specify a "test set" in the context of patient data or clinical imaging. The tests were performed on the device itself and its components. Therefore, information regarding human data provenance (country of origin, retrospective/prospective) is not applicable.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. The ground truth for the non-clinical performance testing is established by the specifications defined in the referenced ASTM standards and engineering principles for finite element analysis.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study involving human interpretation of data where adjudication would be necessary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This is a submission for a medical device (implant) and not an AI/ML-driven diagnostic or assistive tool.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone performance study in the context of an algorithm or AI was not conducted. The device itself is the subject of the testing.

    7. Type of Ground Truth Used

    The ground truth used for this submission is based on:

    • Established engineering standards: ASTM standards for mechanical testing and MR compatibility.
    • Computational modeling results: Finite element analysis, which provides simulated performance data.
    • Predicate device characteristics: The substantially equivalent performance to previously cleared predicate devices serves as the benchmark.

    8. Sample Size for the Training Set

    Not applicable. There is no mention of a "training set" as this is not an AI/ML device or a study involving machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set involved.