    K Number
    K220453

    Validate with FDA (Live)

    Date Cleared
    2022-03-24

    (35 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FlareHawk Interbody Fusion System is indicated for spinal intervertebral body fusion with autogenous bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone in skeletally mature individuals with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, following discectomy. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have at least six (6) months of non-operative treatment. Additionally, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). FlareHawk system spacers are intended to be used with supplemental fixation instrumentation, which has been cleared for use in the lumbar spine.

    Device Description

    The Integrity Implants FlareHawk Interbody Fusion System is an expandable lumbar intervertebral body fusion device intended for use in the lumbosacral spine from L2 to S1. The FlareHawk interbody fusion device consists of a shell and a shim component that are offered in various lengths, heights, and lordotic angles to accommodate variations in patient anatomy. When the FlareHawk device is deployed within the intervertebral disc space, the shell and shim components lock together to create a complete implant construct to provide structural stability for interbody fusion. The final dimensions of the deployed device construct are determined by the dimensions of the selected shell and shim. Once implanted, the FlareHawk interbody fusion device is designed to restore intervertebral disc height, provide anterior column support, and maintain structural stability of the motion segment to facilitate intervertebral body fusion. The FlareHawk interbody fusion device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone, and with supplemental fixation instrumentation that has been cleared for use in the lumbar spine. The FlareHawk Interbody Fusion System includes manual surgical instruments for delivery of the implant device and for disc preparation.

    AI/ML Overview

    This looks like a 510(k) summary for a medical device called the FlareHawk Interbody Fusion System, which is an expandable lumbar intervertebral body fusion device. The document primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices rather than providing detailed acceptance criteria and a study proving those criteria are met for a new AI/ML-driven device.

    Therefore, many of the requested categories (e.g., sample sizes for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth for training data, how ground truth for training data was established) are not applicable to this document because it is not about an AI/ML device that generates diagnostic or prognostic outputs requiring such evaluations.

    However, I can extract the closest analogous information regarding acceptance criteria and the study performed, focusing on the bench performance testing mentioned.

    Here's a breakdown of the information that can be extracted or noted as not applicable:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in precise numerical terms with corresponding reported performance for comparison in a table format as might be expected for AI/ML device metrics (e.g., sensitivity, specificity, AUC). Instead, it refers to industry standards for mechanical testing.

    Acceptance Criteria (Implied)Reported Device Performance
    Compliance with ASTM F2077 for Static Axial CompressionConfirmatory bench performance testing conducted to demonstrate compliance.
    Compliance with ASTM F2077 for Dynamic Axial CompressionConfirmatory bench performance testing conducted to demonstrate compliance.
    Compliance with ASTM F2077 for Static Compression ShearConfirmatory bench performance testing conducted to demonstrate compliance.
    Compliance with ASTM F2077 for Dynamic Compression ShearConfirmatory bench performance testing conducted to demonstrate compliance.
    No new risks introduced by device modificationsRisk analysis completed; engineering rationales demonstrated no new risks.
    Device does not create a new worst case for the overall systemRisk analysis completed; engineering rationales demonstrated no new worst case.
    Substantial Equivalence to predicate devicesDemonstrated through engineering rationales, confirmatory bench testing, and supporting documentation.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated as this is mechanical testing of physical implants, not a data-driven AI/based test. The "test set" would be the number of physical devices subjected to the bench tests.
    • Data Provenance: Not applicable in the context of digital data; this refers to results from physical bench testing performed by Integrity Implants Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. Ground truth for mechanical testing is typically established by physical measurement against engineering specifications and industry standards, not expert consensus on interpretations.

    4. Adjudication method for the test set

    • Not applicable. This is mechanical testing against standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an interbody fusion system, not an AI/ML diagnostic or assistive device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical implant.

    7. The type of ground truth used

    • Ground Truth: Engineering specifications and compliance with established ASTM standards (specifically ASTM F2077) for mechanical properties (Static Axial Compression, Dynamic Axial Compression, Static Compression Shear, Dynamic Compression Shear).

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/ML device that requires a training set.

    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The document describes confirmatory bench performance testing as the study method. This testing was conducted in support of the premarket notification submission to demonstrate that the modified FlareHawk Interbody Fusion System maintains performance characteristics comparable to its predicate devices and adheres to relevant industry standards.

    The specific tests performed were:

    • Static Axial Compression in accordance with ASTM F2077
    • Dynamic Axial Compression in accordance with ASTM F2077
    • Static Compression Shear in accordance with ASTM F2077
    • Dynamic Compression Shear in accordance with ASTM F2077

    Additionally, a risk analysis was completed, demonstrating that no new risks were introduced and that the device modifications did not create a new worst-case scenario.

    The conclusion stated is that, based on engineering rationales, confirmatory bench performance testing, and other supporting documentation, the subject FlareHawk® Interbody Fusion System demonstrates substantial equivalence to legally marketed predicate devices. This indicates that the device's performance met the implicit acceptance criteria of conforming to the established mechanical standards and not introducing new risks compared to its predicates.