510(k) Data Aggregation

    K Number
    K220355

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-26

    (107 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Astroglide Ultra Gentle Gel is a personal lubricant for penile, vaginal, and/or anal application intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    Astroglide Ultra Gentle Gel is a non-sterile, clear, odorless, and water based personal lubricant. This product is not a spermicide or contraceptive. It is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. The device is composed of water, propylene glycol, hydroxyethylcellulose, polyquaternium 7, potassium sorbate, sodium benzoate and citric acid. The product primary packaging is a white LDPE 3-ounce tube with a screw on cap. The tube and cap constitute the device final packaging.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the supporting study, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Specification)Reported Device Performance (Met/Not Met/Not Applicable)
    ColorClear to goldenMet (Implied, as testing "met specifications")
    ClarityClearMet (Implied, as testing "met specifications")
    OdorOdorlessMet (Implied, as testing "met specifications")
    Absence of particulate matterNo particlesMet (Implied, as testing "met specifications")
    Viscosity10,000-25,000 cPMet (Implied, as testing "met specifications")
    pH3.5-5.5Met (Implied, as testing "met specifications")
    Osmolality200-300 mOsm/kg, 1:5 dilution factorMet (Implied, as testing "met specifications")
    Total yeast/mold count (TYMC)<10 cfu/mL (USP <61>)Met (Implied, as testing "met specifications")
    Total aerobic microbial count (TAMC)<100 cfu/mL (USP <61>)Met (Implied, as testing "met specifications")
    Presence of Pathogenic Organisms (Staphylococcus aureus, Pseudomonas aeruginosa, Candida albicans)Absent (USP <62>)Met (Implied, as testing "met specifications")
    Antimicrobial effectivenessMeets USP <51> acceptance criteria for Category 2 productsMet (Implied, as testing "met specifications")
    CytotoxicityNon-cytotoxic (per ISO 10993-5:2009/(R)2014)Met
    Sensitization/IrritationNon-sensitizing, Non-irritating (per ASTM D6355)Met
    Systemic ToxicityNot systemically toxic (per ISO 10993-11:2017)Met
    Shelf Life (24 months)Maintain all specifications listed in Table 1Met (Confirmed by accelerated aging study)
    Compatibility with Natural Rubber Latex CondomsCompatible (per ASTM D7661-18)Compatible
    Compatibility with Polyisoprene CondomsCompatible (per ASTM D7661-18)Compatible
    Compatibility with Polyurethane CondomsNot Compatible (per ASTM D7661-18)Not Compatible (This is a reported finding, not a failure of a positive criterion)

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each test in the "Summary of Non-Clinical Performance Testing" section. For the Shelf Life study, it mentions an "accelerated aging study, conducted for 8 months at 40°C," but doesn't detail the number of samples or batches tested. The provenance of the data is internal to BioFilm, Inc., as these are non-clinical performance tests conducted for device clearance. There is no information regarding country of origin or whether it's retrospective or prospective beyond the description of the tests themselves.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this type of device and study. The testing described (biocompatibility, shelf life, condom compatibility) are laboratory-based, objective performance tests conducted against established international or national standards (e.g., ISO, ASTM, USP). Ground truth is established by the test method's specified criteria, not by expert consensus in the way a clinical diagnostic device would require.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as the tests are objective laboratory performance tests, not subjective interpretations requiring adjudication among experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for AI-driven diagnostic or screening devices involving human interpretation, not for a personal lubricant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a personal lubricant, not an AI algorithm. The performance tests are directly on the physical product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the acceptance criteria for most of the parameters (Color, Clarity, Odor, Absence of particulate matter, Viscosity, pH, Osmolality, Microbial Limits, Antimicrobial effectiveness) is based on pre-defined quantifiable specifications and standard laboratory test methods (USP <61>, USP <62>, USP <51>).

    For Biocompatibility, the ground truth is established by the pass/fail criteria defined in international standards like ISO 10993-1, ISO 10993-5, ISO 10993-11, and ASTM D6355.

    For Condom Compatibility, the ground truth is established by the pass/fail criteria defined in ASTM D7661-18.

    For Shelf Life, the ground truth is that the device must continue to meet all the physical, chemical, and microbiological specifications after accelerated aging, following ASTM F1980-16.

    8. The sample size for the training set

    This is not applicable as the device is not an AI/ML algorithm. There is no "training set" for a physical personal lubricant.

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set for this device.